Evaluation of a Dry Format Reagent Alternative for CD4 T-Cell Enumeration for the FACSCount System: A Report on a Moroccan-Canadian Study

被引:10
作者
Bergeron, Michele [1 ]
Ding, Tao [1 ]
Elharti, Elmir [2 ]
Oumzil, Hicham [2 ]
Soucy, Nadia [1 ]
Harmouche, Hicham [3 ]
Chaouch, Saad [4 ]
El Aouad, Rajae [2 ]
Chabot, Christian [1 ]
Mandy, Francis [1 ]
机构
[1] Publ Hlth Agcy Canada, Natl HIV Immunol Lab, Ottawa, ON K1A 0K9, Canada
[2] Moroccan Minist Hlth, Natl Reference Lab HIV, Natl Inst Hyg, Rabat 10000, Morocco
[3] CHU Ibn Sina Rabat, Moroccan Minist Hlth, Internal Med Unit, Rabat 10000, Morocco
[4] Hosp Duc Tovar, HIV Referral Unit, Moroccan Minist Hlth, Tangier 90000, Morocco
关键词
CD4+T-cells absolute count; FACSCount system; alternative method; dry reagents; LYMPHOCYTE; METHODOLOGIES;
D O I
10.1002/cyto.b.20501
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Efforts to improve alternative CD4 T-cell counting methods are critical to accelerate the implementation of HIV antiretroviral therapy in resources limited regions. Substituting liquid format reagents to eliminate cold-chain transportation and refrigerated storage with dry format reagents contributes to higher efficiency supply management solution especially for laboratories at remote locations. ReaMetrix has developed dry format reagent kits compatible with the FACSCount system, a dedicated flow cytometer for T-cell subset enumeration widely used in resource limited settings. A dual site collaborative study was designed to compare T-cell subsets using both the new dry format ReaMetrix reagent and the original BD Biosciences liquid reagents. Method: A total of 167 HIV positive samples prepared with Rea T Count (ReaMetrix) and FACSCount (BD Biosciences) reagents were analyzed using FACSCount Systems. To compare both methods, Bland-Altman, Pollock, Scott % similarity and correlation coefficient statistical analysis was applied. Immuno-Trol served as an assay processing control and quality indicator of interlaboratory and intralaboratory variation. Results: The mean bias and limits of agreement for CD4 T-cell measurements between Rea T Count and FACSCount reagents were 16 cells/mu l (-4.6%) and 74 to +43, respectively. The correlation obtained was 0.988 with a similarity of 97.9%. Between laboratory variation data was very good with %CV below 10%. Conclusion: The introduction of dry reagents permits the elimination of cold-chain transportation and the on-site refrigerated storage without compromise to assay quality. The substitution of dry reagents facilitates easier supply management practice that will assure wider access to quality HIV treatment. (C) 2009 Clinical Cytometry Society
引用
收藏
页码:188 / 193
页数:6
相关论文
共 16 条
[1]  
[Anonymous], 2009, PRIOR INT HIV AIDS P
[2]   Evaluation of a universal template for single-platform absolute T-lymphocyte subset enumeration [J].
Bergeron, M ;
Faucher, S ;
Ding, T ;
Phaneuf, S ;
Mandy, F .
CYTOMETRY, 2002, 50 (02) :62-68
[3]   QASI, an International Quality Management System for CD4 T-Cell Enumeration Focused to Make a Global Difference [J].
Bergeron, Michele ;
Ding, Tao ;
Houle, Guy ;
Ares, Linda ;
Chabot, Christian ;
Soucy, Nadia ;
Seely, Peggy ;
Sherring, Alice ;
Bogdanovic, Dragica ;
Faucher, Sylvie ;
Summers, Randy ;
Somorjai, Ray ;
Sandstrom, Paul .
CYTOMETRY PART B-CLINICAL CYTOMETRY, 2010, 78B (01) :41-48
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Absolute CD4 T-Cell counting in resource-poor settings - Direct volumetric measurements versus bead-based clinical flow cytometry instruments [J].
Dieye, TN ;
Vereecken, C ;
Diallo, AA ;
Ondoa, P ;
Diaw, PA ;
Camara, M ;
Karam, F ;
Mboup, S ;
Kestens, L .
JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2005, 39 (01) :32-37
[6]   EVALUATION OF 4 ALTERNATIVE METHODOLOGIES FOR DETERMINATION OF ABSOLUTE CD4(+) LYMPHOCYTE COUNTS [J].
JOHNSON, D ;
HIRSCHKORN, D ;
BUSCH, MP .
JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES AND HUMAN RETROVIROLOGY, 1995, 10 (05) :522-530
[7]  
Lopez A, 1999, CYTOMETRY, V38, P231, DOI 10.1002/(SICI)1097-0320(19991015)38:5<231::AID-CYTO5>3.0.CO
[8]  
2-Y
[9]   Evaluation of the FACScount, TRAx CD4 and Dynabeads methods for CD4 lymphocyte determination [J].
Lyamuya, EF ;
Kagoma, C ;
Mbena, EC ;
Urassa, WK ;
Pallangyo, K ;
Mhalu, FS ;
Biberfeld, G .
JOURNAL OF IMMUNOLOGICAL METHODS, 1996, 195 (1-2) :103-112
[10]   Affordable diagnostics - Changing the paradigm in India [J].
Manian, Bala S. .