Thermal risks from LED- and high-intensity QTH-curing units during polymerization of dental resins

被引:59
作者
Bouillaguet, S
Caillot, G
Forchelet, J
Cattani-Lorente, M
Wataha, JC
Krejci, N
机构
[1] Univ Geneva, Sch Dent Med, Div Dent Mat, Sect Dent Med, CH-1205 Geneva, Switzerland
[2] Sch Engn, Dept Ind Syst & Microtechnol, Yverdon, Switzerland
[3] Med Coll Georgia, Sch Dent, Dept Oral Rehabil, Augusta, GA 30912 USA
[4] Univ Geneva, Sch Dent Med, Div Cariol & Endodontol, CH-1205 Geneva, Switzerland
关键词
photopolymerization; composites; infrared thermography; temperature; dentin;
D O I
10.1002/jbm.b.30143
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this study was to test the ability of an infrared (IR) camera to assess temperature changes and distributions in teeth below restorations when quartz-tungsten-halogen (QTH) and light-emitting diode (LED) curing lights were used to photopolymerize the restorative material. Our hypothesis was that the higher power density and broader spectral distribution of the QTH source would cause greater increases in tooth temperature than the LED source, and that these differences would be best demonstrated with the IR camera. Cavities were prepared on human third molars and restored with a resin composite restorative material. The material was light-cured using three light-curing sources using several exposure times. The external (outside the tooth) and internal (inside the pulp chamber) temperature changes during polymerization of the composite material were recorded over 360 s with thermocouples and an IR camera. Using thermocouples the maximum increase in external temperature (+17.7degreesC) was reported for the Swiss Master light after 20 s of curing time while the minimum temperature rise (+7.8degreesC) was reported for the Freelight 2. Whereas a 2.6degreesC increase in internal temperature was observed after curing 20 s with the Freelight 2, 7.1degreesC was reported after 60 s of light exposure to Astralis 10. Infrared images showed similar trends in external-internal rises in temperature as the thermocouples, although temperatures measured by the IR were generally higher. These results indicate that the higher power density QTH sources caused greater increases in tooth temperatures than the LED source and that thermocouples may underestimate the heat applied to the tooth. (C) 2004 Wiley Periodicals, Inc.
引用
收藏
页码:260 / 267
页数:8
相关论文
共 34 条
[1]  
Anic I, 1996, LASER SURG MED, V19, P438
[2]   Clinical and histological evaluation of thermal injury thresholds in human teeth: A preliminary study [J].
Baldissara, P ;
Catapano, S ;
Scotti, R .
JOURNAL OF ORAL REHABILITATION, 1997, 24 (11) :791-801
[3]   Thermal changes observed at implant sites during internal and external irrigation [J].
Benington, IC ;
Biagioni, PA ;
Briggs, J ;
Sheridan, S ;
Lamey, PJ .
CLINICAL ORAL IMPLANTS RESEARCH, 2002, 13 (03) :293-297
[4]   Performance of two blue light-emitting-diode dental light curing units with distance and irradiation-time [J].
Bennett, AW ;
Watts, DC .
DENTAL MATERIALS, 2004, 20 (01) :72-79
[5]  
Bornman L, 1999, J CELL PHYSIOL, V178, P1
[6]  
Chang JC, 1998, LASER SURG MED, V22, P46, DOI 10.1002/(SICI)1096-9101(1998)22:1<46::AID-LSM11>3.0.CO
[7]  
2-6
[8]   The use of thermovision camera to observe physiological and pathological conditions of oral cavity mucous membrane [J].
Dabrowski, M ;
Dulski, R ;
Zmuda, S ;
Zaborowski, P ;
Pogorzelski, C .
INFRARED PHYSICS & TECHNOLOGY, 2002, 43 (3-5) :265-269
[9]  
Goodis HE, 2002, SELTZER AND BENDER'S DENTAL PULP, P371
[10]   In-vitro pulp chamber temperature rise during composite resin polymerization with various light-curing sources [J].
Hannig, M ;
Bott, B .
DENTAL MATERIALS, 1999, 15 (04) :275-281