Application of dose kernel calculation using a simplified Monte Carlo method to treatment plan for scanned proton beams

被引:1
|
作者
Mizutani, Shohei [1 ]
Takada, Yoshihisa [1 ]
Kohno, Ryosuke [2 ]
Hotta, Kenji [2 ]
Tansho, Ryohei [1 ]
Akimoto, Tetsuo [2 ]
机构
[1] Univ Tsukuba, Grad Sch Pure & Appl Sci, Tsukuba, Ibaraki, Japan
[2] Natl Canc Ctr Hosp East, Res Ctr Innovat Oncol, Particle Therapy Div, Chiba, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2016年 / 17卷 / 02期
关键词
dose kernel calculation; simplified Monte Carlo method; treatment planning system; nasopharyngeal tumor case; proton pencil beam scanning; PARTICLE THERAPY; CLINICAL IMPLEMENTATION; SYSTEM; IRRADIATION; ALGORITHM; ACCURACY;
D O I
10.1120/jacmp.v17i2.5747
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Full Monte Carlo (FMC) calculation of dose distribution has been recognized to have superior accuracy, compared with the pencil beam algorithm (PBA). However, since the FMC methods require long calculation time, it is difficult to apply them to routine treatment planning at present. In order to improve the situation, a simplified Monte Carlo (SMC) method has been introduced to the dose kernel calculation applicable to dose optimization procedure for the proton pencil beam scanning. We have evaluated accuracy of the SMC calculation by comparing a result of the dose kernel calculation using the SMC method with that using the FMC method in an inhomogeneous phantom. The dose distribution obtained by the SMC method was in good agreement with that obtained by the FMC method. To assess the usefulness of SMC calculation in clinical situations, we have compared results of the dose calculation using the SMC with those using the PBA method for three clinical cases of tumor treatment. The dose distributions calculated with the PBA dose kernels appear to be homogeneous in the planning target volumes (PTVs). In practice, the dose distributions calculated with the SMC dose kernels with the spot weights optimized with the PBA method show largely inhomogeneous dose distributions in the PTVs, while those with the spot weights optimized with the SMC method have moderately homogeneous distributions in the PTVs. Calculation using the SMC method is faster than that using the GEANT4 by three orders of magnitude. In addition, the graphic processing unit (GPU) boosts the calculation speed by 13 times for the treatment planning using the SMC method. Thence, the SMC method will be applicable to routine clinical treatment planning for reproduction of the complex dose distribution more accurately than the PBA method in a reasonably short time by use of the GPU-based calculation engine.
引用
收藏
页码:315 / 327
页数:13
相关论文
共 50 条
  • [21] Millisecond speed deep learning based proton dose calculation with Monte Carlo accuracy
    Pastor-Serrano, Oscar
    Perko, Zoltan
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (10)
  • [22] Measurement-based validation of a commercial Monte Carlo dose calculation algorithm for electron beams
    Pittomvils, Geert
    Bogaert, Evelien
    Traneus, Erik
    Thysebaert, Pieternel
    De Wagter, Carlos
    MEDICAL PHYSICS, 2022, 49 (07) : 4755 - 4767
  • [23] Moving GPU-OpenCL-based Monte Carlo dose calculation toward clinical use: Automatic beam commissioning and source sampling for treatment plan dose calculation
    Tian, Zhen
    Li, Yongbao
    Hassan-Rezaeian, Nima
    Jiang, Steve B.
    Jia, Xun
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02): : 69 - 84
  • [24] Evaluation of a commercial Monte Carlo dose calculation algorithm for electron treatment planning
    Huang, Jessie Y.
    Dunkerley, David
    Smilowitz, Jennifer B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (06): : 184 - 193
  • [25] Experimental verification of a commercial Monte Carlo-based dose calculation module for high-energy photon beams
    Kuenzler, Thomas
    Fotina, Irina
    Stock, Markus
    Georg, Dietmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (24) : 7363 - 7377
  • [26] A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4
    Grevillot, L.
    Bertrand, D.
    Dessy, F.
    Freud, N.
    Sarrut, D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) : 5203 - 5219
  • [27] Clinical dose assessment for scanned carbon-ion radiotherapy using linear energy transfer measurements and Monte Carlo simulations
    Nakaji, Taku
    Kanai, Tatsuaki
    Takashina, Masaaki
    Matsumura, Akihiko
    Osaki, Kohei
    Yagi, Masashi
    Tsubouchi, Toshiro
    Hamatani, Noriaki
    Ogawa, Kazuhiko
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (24)
  • [28] Calculation of dose point kernel values for monoenergetic electrons and beta emitting radionuclides: Intercomparison of Monte Carlo codes
    Mendes, Bruno Melo
    Guimaraes Antunes, Paula Cristina
    Lopes Branco, Isabela Soares
    do Nascimento, Eduardo
    Seniwal, Baljeet
    Ferreira Fonseca, Telma Cristina
    Yoriyaz, Helio
    RADIATION PHYSICS AND CHEMISTRY, 2021, 181
  • [29] MOQUI: an open-source GPU-based Monte Carlo code for proton dose calculation with efficient data structure
    Lee, Hoyeon
    Shin, Jungwook
    Verburg, Joost M.
    Bobic, Mislav
    Winey, Brian
    Schuemann, Jan
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (17)
  • [30] Treatment plan prescreening for patient-specific quality assurance measurements using independent Monte Carlo dose calculations
    Xu, Yuan
    Zhang, Ke
    Liu, Zhiqiang
    Liang, Bin
    Ma, Xiangyu
    Ren, Wenting
    Men, Kuo
    Dai, Jianrong
    FRONTIERS IN ONCOLOGY, 2022, 12