Application of dose kernel calculation using a simplified Monte Carlo method to treatment plan for scanned proton beams

被引:1
|
作者
Mizutani, Shohei [1 ]
Takada, Yoshihisa [1 ]
Kohno, Ryosuke [2 ]
Hotta, Kenji [2 ]
Tansho, Ryohei [1 ]
Akimoto, Tetsuo [2 ]
机构
[1] Univ Tsukuba, Grad Sch Pure & Appl Sci, Tsukuba, Ibaraki, Japan
[2] Natl Canc Ctr Hosp East, Res Ctr Innovat Oncol, Particle Therapy Div, Chiba, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2016年 / 17卷 / 02期
关键词
dose kernel calculation; simplified Monte Carlo method; treatment planning system; nasopharyngeal tumor case; proton pencil beam scanning; PARTICLE THERAPY; CLINICAL IMPLEMENTATION; SYSTEM; IRRADIATION; ALGORITHM; ACCURACY;
D O I
10.1120/jacmp.v17i2.5747
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Full Monte Carlo (FMC) calculation of dose distribution has been recognized to have superior accuracy, compared with the pencil beam algorithm (PBA). However, since the FMC methods require long calculation time, it is difficult to apply them to routine treatment planning at present. In order to improve the situation, a simplified Monte Carlo (SMC) method has been introduced to the dose kernel calculation applicable to dose optimization procedure for the proton pencil beam scanning. We have evaluated accuracy of the SMC calculation by comparing a result of the dose kernel calculation using the SMC method with that using the FMC method in an inhomogeneous phantom. The dose distribution obtained by the SMC method was in good agreement with that obtained by the FMC method. To assess the usefulness of SMC calculation in clinical situations, we have compared results of the dose calculation using the SMC with those using the PBA method for three clinical cases of tumor treatment. The dose distributions calculated with the PBA dose kernels appear to be homogeneous in the planning target volumes (PTVs). In practice, the dose distributions calculated with the SMC dose kernels with the spot weights optimized with the PBA method show largely inhomogeneous dose distributions in the PTVs, while those with the spot weights optimized with the SMC method have moderately homogeneous distributions in the PTVs. Calculation using the SMC method is faster than that using the GEANT4 by three orders of magnitude. In addition, the graphic processing unit (GPU) boosts the calculation speed by 13 times for the treatment planning using the SMC method. Thence, the SMC method will be applicable to routine clinical treatment planning for reproduction of the complex dose distribution more accurately than the PBA method in a reasonably short time by use of the GPU-based calculation engine.
引用
收藏
页码:315 / 327
页数:13
相关论文
共 50 条
  • [1] Macro Monte Carlo for dose calculation of proton beams
    Fix, Michael K.
    Frei, Daniel
    Volken, Werner
    Born, Ernst J.
    Aebersold, Daniel M.
    Manser, Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (07) : 2027 - 2044
  • [2] Monte Carlo investigation of the low-dose envelope from scanned proton pencil beams
    Sawakuchi, Gabriel O.
    Titt, Uwe
    Mirkovic, Dragan
    Ciangaru, George
    Zhu, X. Ronald
    Sahoo, Narayan
    Gillin, Michael T.
    Mohan, Radhe
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (03) : 711 - 721
  • [3] Improved dose-calculation accuracy in proton treatment planning using a simplified Monte Carlo method verified with three-dimensional measurements in an anthropomorphic phantom
    Hotta, Kenji
    Kohno, Ryosuke
    Takada, Yoshihisa
    Hara, Yousuke
    Tansho, Ryohei
    Himukai, Takeshi
    Kameoka, Satoru
    Matsuura, Taeko
    Nishio, Teiji
    Ogino, Takashi
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12) : 3545 - 3556
  • [4] Monte Carlo calculation of beam quality correction factors in proton beams using PENH
    Goma, Carles
    Sterpin, Edmond
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (18)
  • [5] Commissioning of pencil beam and Monte Carlo dose engines for non-isocentric treatments in scanned proton beam therapy
    Carlino, A.
    Boehlen, T.
    Vatnitsky, S.
    Grevillot, L.
    Osorio, J.
    Dreindl, R.
    Palmans, H.
    Stock, M.
    Kragl, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (17)
  • [6] Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams
    Dreindl, Ralf
    Bolsa-Ferruz, Marta
    Fayos-Sola, Rosa
    Cabal, Fatima Padilla
    Scheuchenpflug, Lukas
    Elia, Alessio
    Amico, Antonio
    Carlino, Antonio
    Stock, Markus
    Grevillot, Loic
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (05):
  • [7] Monte Carlo dose calculation improvements for low energy electron beams using eMC
    Fix, Michael K.
    Frei, Daniel
    Volken, Werner
    Neuenschwander, Hans
    Born, Ernst J.
    Manser, Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (16) : 4577 - 4588
  • [8] Experimental verification of dose calculation using the simplified Monte Carlo method with an improved initial beam model for a beam-wobbling system
    Tansho, Ryohei
    Takada, Yoshihisa
    Kohno, Ryosuke
    Hotta, Kenji
    Hara, Yousuke
    Mizutani, Shohei
    Akimoto, Tetsuo
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (17) : 6047 - 6064
  • [9] A simplified Monte Carlo algorithm considering large-angle scattering for fast and accurate calculation of proton dose
    Takayanagi, Taisuke
    Hirayama, Shusuke
    Fujitaka, Shinichiro
    Fujimoto, Rintaro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 60 - 72
  • [10] FLUKA particle therapy tool for Monte Carlo independent calculation of scanned proton and carbon ion beam therapy
    Kozlowska, Wioletta S.
    Bohlen, Till T.
    Cuccagna, Caterina
    Ferrari, Alfredo
    Fracchiolla, Francesco
    Magro, Giuseppe
    Mairani, Andrea
    Schwarz, Marco
    Vlachoudis, Vasilis
    Georg, Dietmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (07)