Validation of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors in a general adult population according to the European Society of Hypertension and the ANSI/AAMI/ISO 81060-2:2013 protocols

被引:6
作者
Bing, Sen [1 ,2 ]
Chen, Kang [2 ]
Hou, Hong [1 ]
Zhang, Weijuan [1 ]
Li, Linyi [2 ,3 ]
Wei, Jiao [3 ]
Shu, Chang [3 ]
Wan, Yi [3 ]
机构
[1] Second Hosp Xian City, Dept Cardiol, Xian, Shaanxi, Peoples R China
[2] Xijing Hosp, Dept Cardiol, Xian, Shaanxi, Peoples R China
[3] Fourth Mil Med Univ, Sch Publ Hlth, Dept Hlth Serv, 169 West Changle Rd, Xian 710032, Shaanxi, Peoples R China
关键词
blood pressure; oscillometry; self-measurement; validation; RELIABILITY; GUIDELINES; DEVICES;
D O I
10.1097/MBP.0000000000000169
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
ObjectiveThis study aimed to determine the accuracy of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors according to the European Society of Hypertension International Protocol revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols.MethodsThe devices were assessed on 33 participants according to the European Society of Hypertension requirements and were then tested on 85 participants according to the ANSI/AAMI/ISO 81060-2:2013 criteria. Procedures and data analysis were carried out following protocol guidelines precisely.ResultsThe Microlife BP A200 Comfort and W2 Slim devices passed the criteria of the European Society of Hypertension International Protocol revision 2010 for both systolic blood pressure and diastolic blood pressure. The devices also fulfilled the ANSI/AAMI/ISO 81060-2:2013 criteria, with mean differences of SBP and DPB between the devices and observers of 0.385.12 and 0.28 +/- 4.29mmHg for the BP A200 Comfort and 1.01 +/- 6.80 and 0.34 +/- 5.62mmHg for the W2 Slim, respectively.ConclusionThe Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors fulfilled the European Society of Hypertension revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols, and can be recommended for self-measurement in the general population.
引用
收藏
页码:118 / 123
页数:6
相关论文
共 10 条
  • [1] Association for the Advancement of Medical Instrumentation, 2009, 8106022009 ANSI AAMI
  • [2] The Japanese Society of Hypertension Guidelines for Self-monitoring of Blood Pressure at Home (Second Edition)
    Imai, Yutaka
    Kario, Kazuomi
    Shimada, Kazuyuki
    Kawano, Yuhei
    Hasebe, Naoyuki
    Matsuura, Hideo
    Tsuchihashi, Takuya
    Ohkubo, Takayoshi
    Kuwajima, Iwao
    Miyakawa, Masaaki
    [J]. HYPERTENSION RESEARCH, 2012, 35 (08) : 777 - 795
  • [3] International Organization for Standardization, 2013, 8106022013 ISO 2
  • [4] Reliability of home blood pressure monitoring: in the context of validation and accuracy
    Jung, Mi-Hyang
    Kim, Gee-Hee
    Kim, Ji-Hoon
    Moon, Keon-Woong
    Yoo, Ki-Dong
    Rho, Tae-Ho
    Kim, Chul-Min
    [J]. BLOOD PRESSURE MONITORING, 2015, 20 (04) : 215 - 220
  • [5] Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure
    Kikuya, M
    Chonan, K
    Imai, Y
    Goto, E
    Ishii, M
    [J]. JOURNAL OF HYPERTENSION, 2002, 20 (04) : 629 - 638
  • [6] 2013 ESH/ESC Guidelines for the management of arterial hypertension
    Taylor, Jennifer
    [J]. EUROPEAN HEART JOURNAL, 2013, 34 (28) : 2108 - 2109
  • [7] Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement
    O'Brien, E
    Asmar, R
    Beilin, L
    Imai, Y
    Mancia, G
    Mengden, T
    Myers, M
    Padfield, P
    Palatini, P
    Parati, G
    Pickering, T
    Redon, J
    Staessen, J
    Stergiou, G
    Verdecchia, P
    [J]. JOURNAL OF HYPERTENSION, 2005, 23 (04) : 697 - 701
  • [8] European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults
    O'Brien, Eoin
    Atkins, Neil
    Stergiou, George
    Karpettas, Nikos
    Parati, Gianfranco
    Asmar, Roland
    Imai, Yutaka
    Wang, Jiguang
    Mengden, Thomas
    Shennan, Andrew
    [J]. BLOOD PRESSURE MONITORING, 2010, 15 (01) : 23 - 38
  • [9] Noninvasive assessment of local pulse pressure - Importance of brachial-to-radial pressure amplification
    Verbeke, F
    Segers, P
    Heireman, S
    Vanholder, R
    Verdonck, P
    Van Bortel, LM
    [J]. HYPERTENSION, 2005, 46 (01) : 244 - 248
  • [10] Validation of the SAW-102 wrist home blood pressure monitor according to the protocols of the British Hypertension Society, the Association for the Advancement of Medical Instrumentation, and the European Society of Hypertension
    Zaetta, Vania
    Longo, Daniele
    Perkovic, Davor
    Perfetti, Paola
    Gabrieli, Alberto
    Prattico, Francesco
    Winnicki, Mikolaj
    [J]. BLOOD PRESSURE MONITORING, 2009, 14 (01) : 32 - 36