Development of a Candidate Reference Measurement Procedure for the Analysis of Cortisol in Human Serum Samples by Isotope Dilution-Gas Chromatography-Mass Spectrometry

被引:5
作者
Kawaguchi, Migaku [1 ]
Takatsu, Akiko [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, NMIJ, Biomed Stand Sect, Tsukuba, Ibaraki 3058563, Japan
关键词
CLINICAL-CHEMISTRY; DEFINITIVE METHOD; ROUTINE METHODS; PLASMA;
D O I
10.2116/analsci.25.989
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A candidate reference measurement procedure involving isotope dilution coupled with gas chromatography-mass spectrometry (GC-MS) has been developed and critically evaluated. An isotopically labeled internal standard, cortisol-d(2), was added to a serum sample. After equilibration, solid-phase extractions (SPE) for sample preparation and derivatization with heptafluorobutyric anhydride (HFBA) were performed for GC-MS analysis. The limit of detection (LOD) and the limit of quantification (LOQ) were 5 and 20 ng g(-1), respectively. The recovery of the added cortisol ranged from 99.8 to 101.0%. Excellent precision was obtained with a within-day variation (RSD) of 0.7% for GC-MS analysis. The accuracy of the measurement was evaluated by comparing of results of this reference measurement procedure on lyophilized human serum reference materials for cortisol (European Reference Materials (ERM)-DA 192) as Certified Reference Materials (CRMs). The results of this method for total cortisol agreed with the certified values within some uncertainty. This method, which demonstrates simply, easy, good accuracy, high precision, and is free from interferences from structural analogues, qualifies as a reference measurement procedure.
引用
收藏
页码:989 / 992
页数:4
相关论文
共 13 条
[1]  
BJORKHEM I, 1981, CLIN CHEM, V27, P733
[2]  
DEBRABANDERE VI, 1995, CLIN CHEM, V41, P1781
[3]  
GASKELL SJ, 1983, CLIN CHEM, V29, P862
[4]   STANDARDIZATION OF HAPTEN IMMUNOPROCEDURES - TOTAL CORTISOL [J].
GOSLING, JP ;
MIDDLE, J ;
SIEKMANN, L ;
READ, G .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1993, 53 :3-41
[5]   CANDIDATE DEFINITIVE METHOD FOR THE DETERMINATION OF CORTISOL IN HUMAN-SERUM [J].
JONCKHEERE, JA ;
DELEENHEER, AP .
BIOMEDICAL MASS SPECTROMETRY, 1983, 10 (03) :197-202
[6]   Determination of serum cortisol using isotope dilution-liquid chromatography-mass spectrometry as a candidate reference method [J].
Jung, PG ;
Kim, B ;
Park, SR ;
So, HY ;
Shi, LH ;
Kim, Y .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 380 (5-6) :782-788
[7]  
LAWSON A, 1985, 9661 EUR EN
[8]  
PATTERSON DG, 1984, CLIN CHEM, V30, P619
[9]  
SIEKMANN L, 1982, J CLIN CHEM CLIN BIO, V20, P883
[10]   Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry [J].
Tai, SSC ;
Welch, MJ .
ANALYTICAL CHEMISTRY, 2004, 76 (04) :1008-1014