Accuracy of ultra-high-resolution computed tomography with a 0.3-mm detector for quantitative assessment of coronary artery stenosis grading in comparison with conventional computed tomography: A phantom study

被引:6
作者
Yamada, Minoru [1 ]
Yamada, Yoshitake [1 ]
Nakahara, Takehiro [1 ]
Okuda, Shigeo [1 ]
Abe, Takayuki [2 ,3 ]
Kuribayashi, Sachio [1 ,4 ]
Jinzaki, Masahiro [1 ]
机构
[1] Keio Univ, Dept Radiol, Sch Med, Shinjuku Ku, 35 Shinanomachi, Tokyo 1608582, Japan
[2] Keio Univ, Sch Med, Clin & Translat Res Ctr, Shinjuku Ku, 35 Shinanomachi, Tokyo 1608582, Japan
[3] Yokohama City Univ, Sch Data Sci, 22-2 Seto, Yokohama, Kanagawa 2360027, Japan
[4] HIMEDIC Imaging Ctr Lake Yamanaka, XIV Yamanakako B2F,562-12 Hirano, Minamitsuru Dist, Yamanashi 4010502, Japan
关键词
Multidetector-row computed tomography; Coronary stenosis; Ultra-high-resolution CT; Phantoms; Imaging; Dimensional measurement accuracy; LUMEN DIAMETER; CT; ANGIOGRAPHY; PERFORMANCE;
D O I
10.1016/j.jcct.2021.11.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The development of ultra-high-resolution CT (U-HRCT) is expected to improve the accuracy of coronary stenosis evaluation. This study aimed to evaluate the accuracy of the stenosis severities of coronary artery phantoms estimated using U-HRCT by comparing them to those estimated with conventional CT. Methods: Coronary artery phantoms with non-calcified and calcified lesions were scanned with conventional CT (64-row x 0.625 mm) and U-HRCT (32-row x 0.3125 mm). The coronary artery phantoms had lumen diameters of 2.0, 3.0, and 4.0 mm with non-calcified lesions representing 0%, 25%, 50%, and 75% stenosis and 3.0 and 4.0 mm with calcified lesions representing 0%, 25%, 50%, and 75% stenosis. The lumen diameters at the stenotic and non-stenotic regions were measured, and the stenosis severities were compared with the true values. Results: For non-calcified lesions, conventional CT significantly underestimated the stenosis severity in the phantom showing 75% stenosis with lumen diameters of 2.0 and 3.0 mm (p < 0.05), while the estimated stenosis severities were not significantly different from the true values at all settings with U-HRCT. For the calcified lesions, conventional CT overestimated the stenosis severities at all settings (p < 0.05), while U-HRCT yielded estimations closer to the true values, although still with some overestimation (p < 0.05). Conclusion: By using U-HRCT, the estimated stenosis severities of the coronary artery with non-calcified lesion become almost equal to the true value, while those with calcified lesion are still overestimated although they become closer to the true value.
引用
收藏
页码:239 / 244
页数:6
相关论文
共 16 条
[1]   64-multislice detector computed tomography coronary angiography as potential alternative to conventional coronary angiography: a systematic review and meta-analysis [J].
Abdulla, Jawdat ;
Abildstrom, Steen Z. ;
Gotzsche, Ole ;
Christensen, Erik ;
Kober, Lars ;
Torp-Pedersen, Christian .
EUROPEAN HEART JOURNAL, 2007, 28 (24) :3042-3050
[2]   Precise measurement of coronary stenosis diameter with CCTA using CT number calibration [J].
Chen, Zhennong ;
Contijoch, Francisco ;
Schluchter, Andrew ;
Grady, Leo ;
Schaap, Michiel ;
Stayman, Web ;
Pack, Jed ;
McVeigh, Elliot .
MEDICAL PHYSICS, 2019, 46 (12) :5514-5527
[3]   LUMEN DIAMETER OF NORMAL HUMAN CORONARY-ARTERIES - INFLUENCE OF AGE, SEX, ANATOMIC VARIATION, AND LEFT-VENTRICULAR HYPERTROPHY OR DILATION [J].
DODGE, JT ;
BROWN, BG ;
BOLSON, EL ;
DODGE, HT .
CIRCULATION, 1992, 86 (01) :232-246
[4]   Development and performance evaluation of an experimental fine pitch detector multislice CT scanner [J].
Imai, Yasuhiro ;
Nukui, Masatake ;
Ishihara, Yotaro ;
Fujishige, Takashi ;
Ogata, Kentaro ;
Moritake, Masahiro ;
Kurochi, Haruo ;
Ogata, Tsuyoshi ;
Yahata, Mitsuru ;
Tang, Xiangyang .
MEDICAL PHYSICS, 2009, 36 (04) :1120-1127
[5]   Body weight-tailored contrast material injection protocol for 64-detector row computed tomography coronary angiography [J].
Isogai, Tsuyoshi ;
Jinzaki, Masahiro ;
Tanami, Yutaka ;
Kusuzaki, Hiroyuki ;
Yamada, Minoru ;
Kuribayashi, Sachio .
JAPANESE JOURNAL OF RADIOLOGY, 2011, 29 (01) :33-38
[6]   Evaluation of In-Stent Restenosis by High Spatial Resolution CT [J].
Jinzaki M. ;
Yamada M. ;
Tanami Y. ;
Kuribayashi S. .
Current Cardiovascular Imaging Reports, 2011, 4 (6) :431-436
[7]   Detection of Attenuated Plaque in Stable Angina With 64-Multidetector Computed Tomography - A Comparison With Intravascular Ultrasound [J].
Jinzaki, Masahiro ;
Okabe, Teruo ;
Endo, Ayaka ;
Kawamura, Akio ;
Koga, Seiko ;
Yamada, Minoru ;
Fukuda, Keiichi ;
Kuribayashi, Sachio .
CIRCULATION JOURNAL, 2012, 76 (05) :1182-1189
[8]   Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT - A Pilot Study Comparison With Conventional Post-Processing Methods and Axial Images Alone [J].
Jinzaki, Masahiro ;
Sato, Kozo ;
Tanami, Yutaka ;
Yamada, Minoru ;
Anzai, Toshihisa ;
Kawamura, Akio ;
Ueno, Koji ;
Kuribayashi, Sachio .
CIRCULATION JOURNAL, 2009, 73 (04) :691-698
[9]  
Kapoor K, 2021, RADIOL CARDIOTHORAC, V3
[10]   Ultra-High-Resolution Computed Tomography Angiography for Assessment of Coronary Artery Stenosis [J].
Motoyama, Sadako ;
Ito, Hajime ;
Sarai, Masayoshi ;
Nagahara, Yasuomi ;
Miyajima, Keiichi ;
Matsumoto, Ryota ;
Doi, Yujiro ;
Kataoka, Yumi ;
Takahashi, Hiroshi ;
Ozaki, Yukio ;
Toyama, Hiroshi ;
Katada, Kazuhiro .
CIRCULATION JOURNAL, 2018, 82 (07) :1844-1851