Hawthorne effect: shortening of fluoroscopy times during radiation measurement studies

被引:55
作者
Vehmas, T
机构
[1] Department of Radiology, Helsinki University, Central Hospital, FIN-00290 Helsinki
关键词
D O I
10.1259/bjr.70.838.9404210
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Screening times were recorded before (n = 92, 13 radiologists) and after (n = 75, 6 radiologists) commencing a protocol with dose-area product (DAP) measurements and filling of structured questionnaires. Fluoroscopy times were significantly (p = 0.0001) longer before starting these measurements (median 4.3, mean 5.2 min) than during them (median 3.2, mean 3.6 min), which indicates a Hawthorne effect. Fluoroscopy times did not increase during the DAP measurement period up to 21 barium enemas and a study period of up to 45 days per radiologist. Previous fluoroscopic radiation measurements based on action during an analysis period may be biased towards too short fluoroscopy times and too low doses. Radiation measurement, even if not scientifically indicated, seems a practical way of reducing doses.
引用
收藏
页码:1053 / 1055
页数:3
相关论文
共 9 条
[1]   SINGLE-BLIND STUDIES OF THE EFFECTS OF IMPROVED PERIODONTAL HEALTH ON METABOLIC CONTROL IN TYPE-1 DIABETES-MELLITUS [J].
ALDRIDGE, JP ;
LESTER, V ;
WATTS, TLP ;
COLLINS, A ;
VIBERTI, G ;
WILSON, RF .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1995, 22 (04) :271-275
[2]   Nonspecific effects in longitudinal studies: Impact on quality of life measures [J].
Bouchet, C ;
Guillemin, F ;
Briancon, S .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 1996, 49 (01) :15-20
[3]  
GOODLEE F, 1992, BRIT MED J, V304, P299
[4]   AUDIBLE RADIATION MONITORS - THE VALUE IN REDUCING RADIATION EXPOSURE TO FLUOROSCOPY PERSONNEL [J].
HOUGH, DM ;
BRADY, A ;
STEVENSON, GW .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (02) :407-408
[5]  
National Radiological Protection Board, 1990, PAT DOS RED DIAGN RA, V1
[6]   WHAT HAPPENED AT HAWTHORNE [J].
PARSONS, HM .
SCIENCE, 1974, 183 (4128) :922-932
[7]   EXTREMITY DOSES DURING INTERVENTIONAL RADIOLOGY [J].
RAMSDALE, ML ;
WALKER, WJ ;
HORTON, PW .
CLINICAL RADIOLOGY, 1990, 41 (01) :34-36
[8]   AN INVESTIGATION INTO THE LEVELS OF RADIATION EXPOSURE IN DIAGNOSTIC EXAMINATIONS INVOLVING FLUOROSCOPY [J].
ROWLEY, KA ;
HILL, SJ ;
WATKINS, RA ;
MOORES, BM .
BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (710) :167-173
[9]   WHAT FACTORS INFLUENCE RADIOLOGISTS FINGER DOSES DURING PERCUTANEOUS DRAINAGES UNDER FLUOROSCOPIC GUIDANCE [J].
VEHMAS, T .
HEALTH PHYSICS, 1993, 65 (02) :161-163