Materials and methods for higher performance screen-printed flexible MRI receive coils

被引:37
作者
Corea, Joseph R. [1 ]
Lechene, P. Balthazar [1 ]
Lustig, Michael [1 ]
Arias, Ana C. [1 ]
机构
[1] Univ Calif Berkeley, Dept Elect Engn & Comp Sci, Berkeley, CA 94720 USA
基金
美国国家卫生研究院;
关键词
MRI coils; flexible MRI coils; phased arrays; NMR probes; screen printing; plastic substrates; TO-NOISE RATIO; PARALLEL MRI; ARRAYS; SENSE;
D O I
10.1002/mrm.26399
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo develop methods for characterizing materials used in screen-printed MRI coils and improve signal-to-noise ratio (SNR) with new lower-loss materials. MethodsAn experimental apparatus was created to characterize dielectric properties of plastic substrates used in receive coils. Coils were fabricated by screen printing conductive ink onto several plastic substrates. Unloaded and sample loaded quality factor (Q(Unloaded)/Q(Loaded)) measurements and scans on a 3T scanner were used to characterize coil performance. An experimental method was developed to describe the relationship between a coil's Q(Unloaded) and the SNR it provides in images of a phantom. In addition, 3T scans of a phantom and the head of a volunteer were obtained with a proof-of-concept printed eight-channel array, and the results were compared with a commercial 12-channel array. ResultsPrinted coils with optimized substrates exhibited up to 97% of the image SNR when compared with a traditional coil on a loading phantom. Q(Unloaded) and the SNR of coils were successfully correlated. The printed array resulted in images comparable to the quality given by the commercial array. ConclusionUsing the proposed methods and materials, the SNR of printed coils approached that of commercial coils while using a new fabrication technique that provided more flexibility and close contact with the patient's body. Magn Reson Med 78:775-783, 2017. (c) 2016 International Society for Magnetic Resonance in Medicine
引用
收藏
页码:775 / 783
页数:9
相关论文
共 29 条
  • [1] Screen-printed flexible MRI receive coils
    Corea, Joseph R.
    Flynn, Anita M.
    Lechene, Balthazar
    Scott, Greig
    Reed, Galen D.
    Shin, Peter J.
    Lustig, Michael
    Arias, Ana C.
    [J]. NATURE COMMUNICATIONS, 2016, 7
  • [2] Perspectives with cryogenic RF probes in biomedical MRI
    Darrasse, L
    Ginefri, JC
    [J]. BIOCHIMIE, 2003, 85 (09) : 915 - 937
  • [3] Doty F.D., 2007, Encyclopedia of Magnetic Resonance
  • [4] Maximizing signal-to-noise ratio in the presence of coil coupling
    Duensing, GR
    Brooker, HR
    Fitzsimmons, JR
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 111 (03): : 230 - 235
  • [5] THE INTRINSIC SIGNAL-TO-NOISE RATIO IN NMR IMAGING
    EDELSTEIN, WA
    GLOVER, GH
    HARDY, CJ
    REDINGTON, RW
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (04) : 604 - 618
  • [6] NOISE PERFORMANCE OF SURFACE COILS FOR MAGNETIC-RESONANCE IMAGING AT 1.5-T
    HAYES, CE
    AXEL, L
    [J]. MEDICAL PHYSICS, 1985, 12 (05) : 604 - 607
  • [7] Horch RA, 2015, P 23 ANN M ISMRM TOR
  • [8] Jin J., 1998, Electromagnetic Analysis and Design in Magnetic Resonance Image, V1st
  • [9] A 64-channel 3T array coil for accelerated brain MRI
    Keil, Boris
    Blau, James N.
    Biber, Stephan
    Hoecht, Philipp
    Tountcheva, Veneta
    Setsompop, Kawin
    Triantafyllou, Christina
    Wald, Lawrence L.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (01) : 248 - 258
  • [10] Massively parallel MRI detector arrays
    Keil, Boris
    Wald, Lawrence L.
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2013, 229 : 75 - 89