Comparative Analysis of Golay Code based Excitation and Coherent Averaging for Non-Invasive Glucose Monitoring System

被引:3
作者
Sarangi, Satyabrata [1 ]
Pai, Praful P. [1 ]
Sanki, Pradyut Kumar [1 ]
Banerjee, Swapna [1 ]
机构
[1] Indian Inst Technol, Dept Elect & Elect Commun Engn, Kharagpur 721302, W Bengal, India
来源
2014 IEEE 27TH INTERNATIONAL SYMPOSIUM ON COMPUTER-BASED MEDICAL SYSTEMS (CBMS) | 2014年
关键词
Averaging; Diabetes; Golay Code; Non-invasive measurement; Photo acoustic spectroscopy;
D O I
10.1109/CBMS.2014.102
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents a method for improvement in SNR and resolution of photo acoustic (PA) signal using Golay coded excitation for a non-invasive and continuous glucose monitoring. The simulation results with a Golay code of length 4 has been shown, which gives a coding gain of 9.03 dB in comparison to regular pulse triggering. Furthermore, Golay coded excitation results in better resolution of PA signal as compared to coherent averaging technique, which can help in improving calibration accuracy.
引用
收藏
页码:485 / 486
页数:2
相关论文
共 13 条
[1]  
[Anonymous], 2008, HDB OPTICAL SENSING
[2]  
[Anonymous], 2010, NUTR THER METAB, V28, P33
[3]  
[Anonymous], 2008, HDB OPTICAL SENSING
[4]  
Cunningham D.D., 2010, In vivo glucose sensing
[5]  
GOLAY MJE, 1949, P IEEE, V37, P637
[6]  
Khalil Omar S, 2004, Diabetes Technol Ther, V6, P660, DOI 10.1089/dia.2004.6.660
[7]  
Khalil OS, 1999, CLIN CHEM, V45, P165
[8]  
Kulkarni O.C., 2010, BIOINF BIOM ENG ICBB, P1
[9]   THEORY OF THE PULSED OPTOACOUSTIC TECHNIQUE [J].
LAI, HM ;
YOUNG, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1982, 72 (06) :2000-2007
[10]   Simulation Study of Photoacoustic Coded Excitation using Golay Codes [J].
Mienkina, Martin P. ;
Eder, Annika ;
Schmitz, Georg ;
Friedrich, Claus-Stefan ;
Gerhardt, Nils C. ;
Hofmann, Martin R. .
2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, :1242-+