Analysis and Evaluation of a Robotic Trephination in Penetrating Keratoplasty

被引:8
作者
Su, Peng [1 ]
Deng, Shijing [2 ]
Huang, Long [1 ]
Song, Yanming [1 ]
Liu, Xiaoyu [3 ]
Yang, Yang [1 ]
机构
[1] Beihang Univ, Sch Mech Engn & Automat, Beijing 100191, Peoples R China
[2] Capital Med Univ, Beijing Tongren Eye Ctr, Beijing Inst Ophthalmol, Beijing 100005, Peoples R China
[3] Beihang Univ, Sch Biol Sci & Med Engn, Minist Educ, Key Lab Biomech & Mechanobiol, Beijing 100191, Peoples R China
来源
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME | 2016年 / 10卷 / 02期
基金
中国国家自然科学基金;
关键词
cornea; keratoplasty; trephination; trephine force; biomechanics; BIOMECHANICAL PROPERTIES; NEEDLE INSERTION; FORCE;
D O I
10.1115/1.4032869
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is difficult to achieve a stably delicate operation in manual microsurgery, and the aim of this paper is to evaluate the robotic trephination that can open a promising perspective for the development of robotic microsurgical system for keratoplasty. A robot for corneal trephination integrating a force/torque sensor is designed based on manual trephine action. The manual experiments and the robotic experiments about penetrating trephination are performed in porcine eyes. The expected values of operational parameters that are references to the robotic trephination are obtained from the manual experiments using probability density functions (PDFs), including linear velocity, angular velocity, and rotating angle. Considering the meanings of the forces/torques, the results of the manual and robotic experiments such as trephine forces/torques and photomicrographs are compared to evaluate the effectiveness of robotic trephination. The manual trephination shows some randomness and this leads to large fluctuations in the trephine forces/torques during the surgery, but the robot may improve overall outcome of the graft, as it is able to carry out the operation stably and produce a uniform cutting margin. There is potential to improve the biomechanical properties in the delicate microsurgery by using the trephine robot and such devices can assist the surgeon to achieve a consistently high-quality result.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] Influence of different keratoplasty techniques on the biomechanical properties of the cornea
    Abdelkader, Almamoun
    [J]. ACTA OPHTHALMOLOGICA, 2013, 91 (07) : E567 - E572
  • [2] Confocal Microscopy of Corneal Wound Healing After Deep Lamellar Keratoplasty in Rabbits
    Abdelkader, Almamoun
    Elewah, El-Sayed M.
    Kaufman, Herbert E.
    [J]. ARCHIVES OF OPHTHALMOLOGY, 2010, 128 (01) : 75 - 80
  • [3] Needle insertion into soft tissue: A survey
    Abolhassani, Niki
    Patel, Rajni
    Moallem, Mehrdad
    [J]. MEDICAL ENGINEERING & PHYSICS, 2007, 29 (04) : 413 - 431
  • [4] Use of near infrared femtosecond lasers as sub-micron radiation microbeam for cell DNA damage and repair studies
    Botchway, S. W.
    Reynolds, P.
    Parker, A. W.
    O'Neill, P.
    [J]. MUTATION RESEARCH-REVIEWS IN MUTATION RESEARCH, 2010, 704 (1-3) : 38 - 44
  • [5] Bourne W M, 1983, Trans Am Ophthalmol Soc, V81, P403
  • [6] Brett P, 2012, I C MECH MACH VIS PR, P447
  • [7] Craig J.J., 1989, INTRO ROBOTICS MECH, P125
  • [8] Biomechanical properties of human and porcine corneas
    Elsheikh, Ahmed
    Alhasso, Daad
    Rama, Paolo
    [J]. EXPERIMENTAL EYE RESEARCH, 2008, 86 (05) : 783 - 790
  • [9] Design of a Lightweight, Cost Effective Thimble-Like Sensor for Haptic Applications Based on Contact Force Sensors
    Ferre, Manuel
    Galiana, Ignacio
    Aracil, Rafael
    [J]. SENSORS, 2011, 11 (12) : 11495 - 11509
  • [10] Galiana I, 2010, LECT NOTES COMPUT SC, V6191, P65, DOI 10.1007/978-3-642-14064-8_10