Applicability of Glass Dosimeters for In-vivo Dosimetry in Brachytherapy

被引:1
作者
Moon, Sun Young [1 ]
Son, Jaeman [1 ]
Yoon, Myonggeun [1 ]
Jeang, EunHee [2 ]
Lim, Young Kyung [2 ]
Chung, Weon Kyu [3 ]
Kim, Dong Wook [3 ]
机构
[1] Korea Univ, Dept Bioconvergence Engn, Seoul 02841, South Korea
[2] Natl Canc Ctr, Proton Therapy Ctr, Goyang 10408, South Korea
[3] Kyung Hee Univ Hosp Gangdong, Dept Radiat Oncol, Seoul 05278, South Korea
基金
新加坡国家研究基金会;
关键词
Brachytherapy; Glass dosimeter; In-vivo dosimetry; Patient dose; DOSE-RATE BRACHYTHERAPY; IR-192 HDR BRACHYTHERAPY; THERMOLUMINESCENCE DOSIMETRY; SCINTILLATION DETECTORS; PROSTATE-CANCER; ROD DETECTOR; VERIFICATION; PHANTOM; ARRAY;
D O I
10.3938/jkps.72.1320
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
During brachytherapy, confirming the dose delivered is very important in order to prevent radiation-associated side effects. Therefore, we aimed to confirm the accuracy of dose delivery near the source by inserting glass dosimeters within the applicator. We created an alternative pelvic phantom with the same shape and internal structures as the usual patient. In addition, we created a tandem for insertion of the glass dosimeters and measured the dose near the source by inserting the glass dosimeters into the tandem and evaluating the accuracy of the dwell position and time through the dose near the source. Errors between the values obtained from the five glass dosimeters and the values from the treatment planning system were -6.27, -2.1, -4.18, 6.31, and -0.39%, respectively. The mean error was 3.85%. This value was acceptable considering that the error of the glass dosimeter itself is approximately 3%. Even though a complement of the applicator and the error calibration is required in order to apply this technique clinically, we believe that radiation accidents and overdoses can be prevented through in-vivo dosimetry using a glass dosimeter for brachytherapy.
引用
收藏
页码:1320 / 1325
页数:6
相关论文
共 28 条
[1]   Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations [J].
Ahn, Woo Sang ;
Choi, Wonsik ;
Shin, Seong Soo ;
Jung, Jinhong .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (04) :557-564
[2]   In vivo dosimetry in the urethra using alanine/ESR during 192Ir HDR brachytherapy of prostate cancer-a phantom study [J].
Anton, Mathias ;
Wagner, Daniela ;
Selbach, Hans-Joachim ;
Hackel, Thomas ;
Hermann, Robert Michael ;
Hess, Clemens Friedrich ;
Vorwerk, Hilke .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :2915-2931
[3]   Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and Cyber-Knife [J].
Araki, F ;
Moribe, N ;
Shimonobou, T ;
Yamashita, Y .
MEDICAL PHYSICS, 2004, 31 (07) :1980-1986
[4]  
BALLESTER F, 2001, PHYS MED BIOL, V46
[5]   A method to estimate the transit dose on the beam axis for verification of dose delivery with portal images [J].
Bogaerts, R ;
Van Esch, A ;
Reymen, R ;
Huyskens, D .
RADIOTHERAPY AND ONCOLOGY, 2000, 54 (01) :39-46
[6]  
CHEKA JS, 1964, HEALTH PHYS, V10, P303
[7]   Characteristic study of a radio-photoluminescence glass rod detector for clinical usages: Skin and inner body in-vivo verification [J].
Chung, Weon Kuu ;
Kim, Dong Wook .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 62 (04) :670-676
[8]   Thermoluminescence dosimetry for in-vivo verification of high dose rate brachytherapy for prostate cancer [J].
Das, R. ;
Toye, W. ;
Kron, T. ;
Williams, S. ;
Duchesne, G. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (03) :178-184
[9]  
Devlin P.M., 2015, Brachytherapy: Applications and Techniques
[10]   The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance [J].
Espinoza, A. ;
Petasecca, M. ;
Fuduli, I. ;
Howie, A. ;
Bucci, J. ;
Corde, S. ;
Jackson, M. ;
Lerch, M. L. F. ;
Rosenfeld, A. B. .
MEDICAL PHYSICS, 2015, 42 (02) :663-673