Increasing the spatial resolution and sensitivity of magnetic resonance elastography by correcting for subject motion and susceptibility-induced image distortions

被引:30
作者
Fehlner, Andreas [1 ]
Hirsch, Sebastian [2 ]
Weygandt, Martin [3 ,4 ]
Christophel, Thomas [3 ,4 ]
Barnhill, Eric [1 ]
Kadobianskyi, Mykola [3 ,4 ]
Braun, Juergen [2 ]
Bernarding, Johannes [5 ]
Luetzkendorf, Ralf [5 ]
Sack, Ingolf [1 ]
Hetzer, Stefan [3 ,4 ]
机构
[1] Charite Univ Med Berlin, Dept Radiol, Berlin, Germany
[2] Charite Univ Med Berlin, Inst Med Informat, Berlin, Germany
[3] Charite Univ Med Berlin, Berlin Ctr Adv Neuroimaging, Berlin, Germany
[4] Bernstein Ctr Computat Neurosci, Berlin, Germany
[5] Otto Von Guericke Univ, Inst Biometry & Med Informat, Magdeburg, Germany
关键词
MR elastography; motion correction; distortion correction; phase-based contrast; multifrequency MRE; ECHO-PLANAR IMAGES; MR ELASTOGRAPHY; HUMAN BRAIN; CONTRAST; TRACKING; PHASE; FMRI; VISCOELASTICITY; ACQUISITION;
D O I
10.1002/jmri.25516
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo improve the resolution of elasticity maps by adapting motion and distortion correction methods for phase-based magnetic resonance imaging (MRI) contrasts such as magnetic resonance elastography (MRE), a technique for measuring mechanical tissue properties in vivo. Materials and MethodsMRE data of the brain were acquired with echo-planar imaging (EPI) at 3T (n=14) and 7T (n=18). Motion and distortion correction parameters were estimated using the magnitude images. The real and imaginary part of the complex MRE data were corrected separately and recombined. The width of the point-spread function (PSF) and the position variability were calculated. The images were normalized to the Montreal Neurological Institute (MNI) anatomical template. The gray-to-white matter separability of the elasticity maps was tested. ResultsMotion correction sharpened the |G*| maps as demonstrated by a narrowing of the PSF by 0.780.51mm at 7T and 0.520.63 mm at 3T. The amount of individual head motion during MRE acquisition correlated with the decrease in the width of the PSF at 7T (r=0.53, P=0.025) and at 3T (r=0.69, P=0.006) and with the increase of gray-to-white matter separability after motion correction at 7T (r=0.64, P=0.0039) and at 3T (r=0.57, P=0.0319). Improved spatial accuracy after distortion correction results in a significant increase in separability of gray and white matter stiffness (P=0.0067), especially in inferior parts of the brain suffering from strong B-0 inhomogeneities. ConclusionWe demonstrate that our method leads to sharper images and higher spatial accuracy, raising the prospect of the investigation of smaller brain areas with increased sensitivity in studies using MRE. Level of Evidence: 1 Technical Efficacy: Stage 1 J. MAGN. RESON. IMAGING 2017;46:134-141
引用
收藏
页码:134 / 141
页数:8
相关论文
共 43 条
[1]   How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging [J].
Andersson, JLR ;
Skare, S ;
Ashburner, J .
NEUROIMAGE, 2003, 20 (02) :870-888
[2]   Functional quantitative susceptibility mapping (fQSM) [J].
Balla, David Z. ;
Sanchez-Panchuelo, Rosa M. ;
Wharton, Samuel J. ;
Hagberg, Gisela E. ;
Scheffler, Klaus ;
Francis, Susan T. ;
Bowtell, Richard .
NEUROIMAGE, 2014, 100 :112-124
[3]   Phase-Based Arterial Input Function Measurements for Dynamic Susceptibility Contrast MRI [J].
Bleeker, Egbert J. W. ;
van Buchem, Mark A. ;
Webb, Andrew G. ;
van Osch, Matthias J. P. .
MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (02) :358-368
[4]   High-resolution mechanical imaging of the human brain by three-dimensional multifrequency magnetic resonance elastography at 7T [J].
Braun, Juergen ;
Guo, Jing ;
Luetzkendorf, Ralf ;
Stadler, Joerg ;
Papazoglou, Sebastian ;
Hirsch, Sebastian ;
Sack, Ingolf ;
Bernarding, Johannes .
NEUROIMAGE, 2014, 90 :308-314
[5]   A TECHNIQUE FOR ACCURATE MAGNETIC-RESONANCE-IMAGING IN THE PRESENCE OF FIELD INHOMOGENEITIES [J].
CHANG, H ;
FITZPATRICK, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (03) :319-329
[6]   ULTRA-FAST IMAGING [J].
COHEN, MS ;
WEISSKOFF, RM .
MAGNETIC RESONANCE IMAGING, 1991, 9 (01) :1-37
[7]   An evaluation of the use of magnetic field maps to undistort echo-planar images [J].
Cusack, R ;
Brett, M ;
Osswald, K .
NEUROIMAGE, 2003, 18 (01) :127-142
[8]   Flow quantitation with echo-planar phase-contrast velocity mapping: In vitro anti in vivo evaluation [J].
Debatin, JF ;
Leung, DA ;
Wildermuth, S ;
Botnar, R ;
Felblinger, J ;
McKinnon, GC .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (06) :656-662
[9]   Dynamic scan-plane tracking using MR position monitoring [J].
Derbyshire, JA ;
Wright, GA ;
Henkelman, RM ;
Hinks, RS .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (04) :924-932
[10]   A method for the dynamic correction of B0-related distortions in single-echo EPI at 7 T [J].
Dymerska, Barbara ;
Poser, Benedikt A. ;
Barth, Markus ;
Trattnig, Siegfried ;
Robinson, Simon D. .
NEUROIMAGE, 2018, 168 :321-331