Method comparison of four clinically available assays for serum free light chain analysis

被引:24
作者
Fleming, Cherina K. A. [1 ]
Swarttouw, Tim [1 ]
Angelino, Corrie M. de Kat [2 ]
Jacobs, Joannes F. M. [2 ]
Russcher, Henk [1 ]
机构
[1] Erasmus MC, Univ Med Ctr Rotterdam, Dept Clin Chem, Dr Molewaterpl 40,Room Na407, NL-3015 GD Rotterdam, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Lab Med, Nijmegen, Netherlands
关键词
free light chains; M-protein; monoclonal gammopathy; multiple myeloma; serum protein electrophoresis; N LATEX FLC; KAPPA; LAMBDA; PERFORMANCE; DIAGNOSIS; MYELOMA;
D O I
10.1515/cclm-2019-0533
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Serum free light chain (sFLC) measurements are increasingly important in the context of screening for monoclonal gammopathies, prognostic stratification and monitoring of therapy responses. In this study we have performed a method comparison of four sFLC assays that are currently available for routine clinical use. Methods: In a retrospective study, sFLC analyses were performed on a cohort that included 139 patients with various monoclonal gammopathies and 54 control sera without an M-protein. Method comparisons of the following four FLC assays were performed: Freelite (Binding Site), N-Latex FLC (Siemens), Seralite (Abingdon Health) and Sebia FLC (Sebia). Results: Bland-Altman agreement analysis showed biases varying between -0.1 and 16.2 mg/L for kappa FLC, -6.0 and 6.8 mg/L for lambda FLC and -0.04 and 0.38 for the ratio of the involved to uninvolved FLC. Strong agreements were observed for FLC-concentrations below 100 mg/L. The clinical concordance of the kappa/lambda FLC-ratio of the four methods varied between 86% and 92%. Significant quantitative differences were observed between the different methods, mainly in sera with high FLC concentrations. Most assays consistently overestimated FLC concentrations compared to SPE. Conclusions: Good overall clinical concordances were observed between the four sFLC assays that were compared in this study. Although good agreements were observed between the FLC assays, significant absolute differences in FLC concentrations in individual patients can be seen, particularly at higher FLC concentrations. Because of inequivalent absolute sFLC values between the methods in individual patients, none of the four sFLC assays can be used interchangeably.
引用
收藏
页码:85 / 94
页数:10
相关论文
共 50 条
  • [41] Evaluation of the Impact of Renal Failure on Correlation and Concordance Between 2 Free Light Chain Assays
    Moreau, Caroline
    Autier, Brice
    Cavey, Thibault
    Rouger, Emmanuel
    Norwood, James
    Bendavid, Claude
    Escoffre, Martine
    Sebillot, Martine
    Decaux, Olivier
    CLINICAL LYMPHOMA MYELOMA & LEUKEMIA, 2016, 16 (12) : 693 - 704
  • [43] Performance Evaluation of Serum Free Light Chain Analysis Nephelometry vs Turbidimetry, Monoclonal vs Polyclonal Reagents
    Messiaen, Anne-Sophie
    De Sloovere, Maxime M. W.
    Claus, Paul-Emile
    Vercammen, Martine
    Van Hoovels, Lieve
    Heylen, Olivier
    Debrabandere, Johan
    Vanpoucke, Hilde
    De Smet, Dieter
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2017, 147 (06) : 611 - 622
  • [44] Effects of Long-Term Storage on Serum Free Light Chain Stability
    Hoerber, Sebastian
    Klein, Reinhild
    Peter, Andreas
    CLINICAL LABORATORY, 2019, 65 (05) : 845 - 851
  • [45] Serum free light chain reference values: A critical approach
    Altinier, Sara
    Seguso, Mara
    Zaninotto, Martina
    Varagnolo, Mariacristina
    Adami, Fausto
    Angeli, Paolo
    Plebani, Mario
    CLINICAL BIOCHEMISTRY, 2013, 46 (7-8) : 691 - 693
  • [46] A Study on Free Light Chain Assay and Serum Immunofixation Electrophoresis for the Diagnosis of Monoclonal Gammopathies
    Kuriakose, Eldho
    Cheppayil, Sumithra Narayanan Unni
    Narayanan, Subhakumari Kuzhikandathil
    Vasudevan, Anu
    INDIAN JOURNAL OF CLINICAL BIOCHEMISTRY, 2019, 34 (01) : 76 - 81
  • [47] Serum Free Light Chain Assay in Monoclonal Gammopathic Manifestations
    Lee, Won S.
    Singh, Gurmukh
    LABORATORY MEDICINE, 2019, 50 (04) : 381 - 389
  • [48] Comparison of the Freelite serum free light chain (SFLC) assay with serum and urine electrophoresis/immunofixation and the N Latex FLC assay
    Sasson, S. C.
    McGill, K.
    Wienholt, L.
    Carr, A.
    Brown, D. A.
    Kelleher, A. D.
    Breit, S. N.
    Sewell, W. A.
    PATHOLOGY, 2015, 47 (06) : 564 - 569
  • [49] Serum free light chain assay reduces the need for serum and urine immunofixation electrophoresis in the evaluation of monoclonal gammopathy
    Fulton, Richard B.
    Fernando, Suran L.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 : 407 - 412
  • [50] Potential pitfalls of serum free light chain analysis to assess treatment response for multiple myeloma
    Abbi, Kamal Kant Singh
    Silverman, Margarida
    Farooq, Umar
    Tricot, Annick
    Dozeman, Lindsey
    Nadiminti, Kalyan
    Krasowski, Matthew D.
    Tricot, Guido J.
    BRITISH JOURNAL OF HAEMATOLOGY, 2016, 174 (04) : 536 - 540