Method comparison of four clinically available assays for serum free light chain analysis

被引:24
|
作者
Fleming, Cherina K. A. [1 ]
Swarttouw, Tim [1 ]
Angelino, Corrie M. de Kat [2 ]
Jacobs, Joannes F. M. [2 ]
Russcher, Henk [1 ]
机构
[1] Erasmus MC, Univ Med Ctr Rotterdam, Dept Clin Chem, Dr Molewaterpl 40,Room Na407, NL-3015 GD Rotterdam, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Lab Med, Nijmegen, Netherlands
关键词
free light chains; M-protein; monoclonal gammopathy; multiple myeloma; serum protein electrophoresis; N LATEX FLC; KAPPA; LAMBDA; PERFORMANCE; DIAGNOSIS; MYELOMA;
D O I
10.1515/cclm-2019-0533
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Serum free light chain (sFLC) measurements are increasingly important in the context of screening for monoclonal gammopathies, prognostic stratification and monitoring of therapy responses. In this study we have performed a method comparison of four sFLC assays that are currently available for routine clinical use. Methods: In a retrospective study, sFLC analyses were performed on a cohort that included 139 patients with various monoclonal gammopathies and 54 control sera without an M-protein. Method comparisons of the following four FLC assays were performed: Freelite (Binding Site), N-Latex FLC (Siemens), Seralite (Abingdon Health) and Sebia FLC (Sebia). Results: Bland-Altman agreement analysis showed biases varying between -0.1 and 16.2 mg/L for kappa FLC, -6.0 and 6.8 mg/L for lambda FLC and -0.04 and 0.38 for the ratio of the involved to uninvolved FLC. Strong agreements were observed for FLC-concentrations below 100 mg/L. The clinical concordance of the kappa/lambda FLC-ratio of the four methods varied between 86% and 92%. Significant quantitative differences were observed between the different methods, mainly in sera with high FLC concentrations. Most assays consistently overestimated FLC concentrations compared to SPE. Conclusions: Good overall clinical concordances were observed between the four sFLC assays that were compared in this study. Although good agreements were observed between the FLC assays, significant absolute differences in FLC concentrations in individual patients can be seen, particularly at higher FLC concentrations. Because of inequivalent absolute sFLC values between the methods in individual patients, none of the four sFLC assays can be used interchangeably.
引用
收藏
页码:85 / 94
页数:10
相关论文
共 50 条
  • [21] Serum free light chain analysis
    Davids, Matthew S.
    Murali, Mandakolathur R.
    Kuter, David J.
    AMERICAN JOURNAL OF HEMATOLOGY, 2010, 85 (10) : 787 - 790
  • [22] Measurement uncertainty for serum free light chain assays: Estimation and implication on result interpretation
    Wang, Li
    Chan, Pak Cheung R.
    CLINICAL BIOCHEMISTRY, 2013, 46 (4-5) : 381 - 384
  • [23] Serum Free Light Chain Assay: Shift Toward a Higher κ/λ Ratio
    Rindlisbacher, Barbara
    Schild, Christof
    Egger, Florence
    Bacher, Vera U.
    Pabst, Thomas
    Leichtle, Alexander
    Andres, Martin
    Sedille-Mostafaie, Nazanin
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2020, 5 (01) : 114 - 125
  • [24] Serum free light chain analysis in multiple myeloma and plasma cell dyscrasias
    Ozsan, Guner Hayri
    Dispenzieri, Angela
    EXPERT REVIEW OF CLINICAL IMMUNOLOGY, 2011, 7 (01) : 65 - 73
  • [25] Analysis of patients with γ-heavy chain disease by the heavy/light chain and free light chain assays
    Kaleta, Erin
    Kyle, Robert
    Clark, Raynell
    Katzmann, Jerry
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (05) : 665 - 669
  • [26] The choice of serum-free light chain analysis method could potentially have clinical consequences for myeloma patients
    Veskovski, Ljupco
    Jakobsson, Ingvar
    Andersson, Per-Ola
    Gustafsson, Therese
    Sedigh, Annelie
    Knut-Bojanowska, Dorota
    Hansson, Markus
    Hveding Blimark, Cecilie
    Mellqvist, Ulf-Henrik
    EJHAEM, 2024, 5 (03): : 455 - 461
  • [27] N Latex FLC serum free light-chain assays in patients with renal impairment
    Jacobs, Joannes F. M.
    Hoedemakers, Rein M. J.
    Teunissen, Elisa
    te Velthuis, Henk
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (06) : 853 - 859
  • [28] Comparison of 2 Free Light Chain Assays: Performance of the Free Light Chain Ratio as a Risk Factor for MGUS Progression
    Wang, Qian
    Andress, Benjamin D.
    Pazdernik, Vanessa M. K.
    Larson, Dirk R.
    Coker, Jonathan D.
    Dasari, Surendra
    Rajkumar, Vincent
    Dispenzieri, Angela
    Murray, David L.
    Willrich, Maria Alice, V
    CLINICAL CHEMISTRY, 2024, 70 (10) : 1268 - 1278
  • [29] Comparison of two serum free light chain assays for the diagnosis of primary plasma cell malignant proliferative disease
    Yang, Yang
    Han, Xiaoyan
    Zheng, Gaofeng
    Cai, Zhen
    HEALTH SCIENCE REPORTS, 2019, 2 (04)
  • [30] Analytical issues of serum free light chain assays and the relative performance of polyclonal and monoclonal based reagents
    Carr-Smith, Hugh D.
    Jenner, Ellen L.
    Evans, Josie A. R.
    Harding, Stephen J.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (06) : 997 - 1003