Experimental analysis of standard, asymmetric and constraint liners for total hip replacement: Investigation of range of motion and stability against dislocation

被引:12
作者
Bader, R
Steinhauser, E
Scholz, R
Simnacher, M
Mittelmeier, W
机构
[1] Univ Rostock, Orthopad Klin & Poliklin, D-18057 Rostock, Germany
[2] Tech Univ Munich, Klin Orthopad & Sportorthopad, D-8000 Munich, Germany
[3] Univ Leipzig, Orthopad Klin & Poliklin, Leipzig, Germany
[4] PI Precis Implants AG, Aarau, Switzerland
来源
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE | 2004年 / 142卷 / 05期
关键词
total hip replacement; impingement; dislocation; cup design;
D O I
10.1055/s-2004-832318
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aim: Recurrent dislocation after total hip replacement is a severe complication, which requires specific treatment and implants. The purpose of the present study was to compare a constraint liner with an elevated rim and standard liner regarding their range of motion and dislocation stability. Method: With a test device, range of motion until impingement (ROMImp) and dislocation (ROMLux) were experimentally analyzed using the above-mentioned insert types of a commercial total hip system. On the basis of movement combinations associated with dislocation, the ROM was determined. Further measuring parameter was the resisting moment against subluxation of the femoral head. Results: The constraint liners showed clear restriction of the movements "internal rotation combined with 90degrees flexion and 0degrees adduction" and "external rotation with 10degrees extension and 15degrees adduction" of up to 20degrees compared to the neutral liner. ROMImp was only decreased by about 8degrees with the elevated-rim liner. The constraint liners revealed the highest resisting moments in subluxation, however, at adequate orientation in the acetabular cup the elevated-rim liners provided a higher ROMLux of UP to 12degrees. Both designs were superior to the neutral liner at retroversion and steep cup position regarding resisting moment and ROMLux. Conclusion: In case of insufficient soft tissue tension the use of constraint liners may increase the dislocation stability, however, in contrast to elevated-rim liners the impingement-free movement interval is clearly reduced. Thus, material damage and high shear stress in the bone interface can result. Therefore, constraint liners should only be used in exceptional cases.
引用
收藏
页码:577 / 585
页数:9
相关论文
共 23 条
[1]   Dislocationm after revision total hip arthroplasty - An analysis of risk factors and treatment options [J].
Alberton, GM ;
High, WA ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10) :1788-1792
[2]   Method for the evaluation of factors influencing the dislocation stability of total hip endoprotheses [J].
Bader, R ;
Scholz, R ;
Steinhauser, E ;
Busch, R ;
Mittlemeier, W .
BIOMEDIZINISCHE TECHNIK, 2004, 49 (05) :137-144
[3]   Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters [J].
Bader, R ;
Steinhauser, E ;
Gradinger, R ;
Willmann, G ;
Mittelmeier, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2002, 140 (03) :310-316
[4]   Impingement and rim wear associated with early osteolysis after a total hip replacement - A case report [J].
Barrack, RL ;
Schmalzried, TP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (07) :1218-1220
[5]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[6]   Recurrent hip arthroplasty dislocation - Good outcome after cup augmentation in 20 patients followed for 2 years [J].
Charlwood, AP ;
Thompson, NW ;
Thompson, NS ;
Beverland, DE ;
Nixon, JR .
ACTA ORTHOPAEDICA SCANDINAVICA, 2002, 73 (05) :502-505
[7]   The elevated-rim acetabular liner in total hip arthroplasty: Relationship to postoperative dislocation [J].
Cobb, TK ;
Morrey, BF ;
Ilstrup, DM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (01) :80-86
[8]  
GENOUD P, 2000, 12 ESB C DUBL, P137
[9]  
Goetz DD, 1998, CLIN ORTHOP RELAT R, P171
[10]   The effect of acetabular cup orientations on limiting hip rotation [J].
Kummer, FJ ;
Shah, S ;
Iyer, S ;
DiCesare, PE .
JOURNAL OF ARTHROPLASTY, 1999, 14 (04) :509-513