A novel, open access, elliptical cross-section magnet for paediatric MRI

被引:15
作者
Crozier, S
Forbes, LK
Doddrell, DM
机构
[1] Univ Queensland, Gehrmann Labs, Ctr Magnet Resonance, Brisbane, Qld 4072, Australia
[2] Univ Queensland, Dept Math, St Lucia, Qld 4072, Australia
关键词
D O I
10.1088/0957-0233/9/1/015
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Almost all clinical magnetic resonance imaging systems are based on circular cross-section magnets. Recent advances in elliptical cross-section RF probe and gradient coil hardware raise the question of the possibility of using elliptical cross-section magnet systems, This paper presents a methodology for calculating rapidly the magnetic fields generated by a multi-turn coil of elliptical cross-section and incorporates this in a stochastic optimization method for magnet design, An open magnet system of elliptical cross-section is designed that both reduces the claustrophobia for the patients and allows ready access by attending physicians, The magnet system is optimized for paediatric use, The coil geometry produced by the optimization method has several novel features.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 20 条
  • [1] BLIMSON WE, 1992, SOC MAGN RES MED 11, P272
  • [2] Brechna H., 1973, Superconducting Magnet Systems
  • [3] Optimization of transverse gradient coils with coaxial return paths by simulated annealing
    Buszko, ML
    Kempka, MF
    Szczesniak, E
    Wang, DC
    Andrew, ER
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 112 (03): : 207 - 213
  • [4] Corona A, 1987, ACM T MATH SOFTWARE, V13, P262
  • [5] THE DESIGN OF TRANSVERSE GRADIENT COILS OF RESTRICTED LENGTH BY SIMULATED ANNEALING
    CROZIER, S
    FORBES, LK
    DODDRELL, DM
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES A, 1994, 107 (01) : 126 - 128
  • [6] GRADIENT-COIL DESIGN BY SIMULATED ANNEALING
    CROZIER, S
    DODDRELL, DM
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES A, 1993, 103 (03) : 354 - 357
  • [7] ERNST RR, 1987, PRINCIPOLES NUCL MAG
  • [8] An analysis and optimization of elliptical RF probes used in magnetic resonance imaging
    Forbes, LK
    Crozier, S
    Doddrell, DM
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (09) : 1281 - 1290
  • [9] FORBES LK, 1997, IEEE T MAGN, V33, P1
  • [10] CONSIDERATIONS IN THE DESIGN OF MRI MAGNETS WITH REDUCED STRAY FIELDS
    HAWKSWORTH, DG
    MCDOUGALL, IL
    BIRD, JM
    BLACK, D
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 1987, 23 (02) : 1309 - 1314