Fast Interleaved Multislice T1 Mapping: Model-Based Reconstruction of Single-Shot Inversion-Recovery Radial FLASH

被引:6
作者
Wang, Xiaoqing [1 ,2 ]
Voit, Dirk [2 ]
Roeloffs, Volkert [2 ]
Uecker, Martin [1 ,3 ]
Frahm, Jens [2 ,3 ]
机构
[1] Univ Med Ctr, Dept Diagnost & Intervent Radiol, Gottingen, Germany
[2] Biomed NMR Forsch GmbH, Max Planck Inst Biophys Chem, D-37070 Gottingen, Germany
[3] DZHK German Ctr Cardiovasc Res, Partner Site, Gottingen, Germany
关键词
IMAGE-RECONSTRUCTION; RELAXATION-TIMES; GOLDEN RATIO; HUMAN BRAIN; T-1; MRI; QUANTIFICATION; RESOLUTION; ACCURATE;
D O I
10.1155/2018/2560964
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose. To develop a high-speed multislice T1 mapping method based on a single-shot inversion-recovery (IR) radial FLASH acquisition and a regularized model-based reconstruction. Methods. Multislice radial k-space data are continuously acquired after a single nonselective inversion pulse using a golden-angle sampling scheme in a spoke-interleaved manner with optimized flip angles. Parameter maps and coil sensitivities of each slice are estimated directly from highly undersampled radial k-space data using a model-based nonlinear inverse reconstruction in conjunction with joint sparsity constraints. The performance of the method has been validated using a numerical and experimental T1 phantom as well as demonstrated for studies of the human brain and liver at 3T. Results. The proposed method allows for 7 simultaneous T1 maps of the brain at 0.5 x 0.5 x 4 mm(3) resolution within a single IR experiment of 4 s duration. Phantom studies confirm similar accuracy and precision as obtained for a single-slice acquisition. For abdominal applications, the proposed method yields three simultaneous T1 maps at 1.25 x 1.25 x 6 mm(3) resolution within a 4 s breath hold. Conclusion. Rapid, robust, accurate, and precise multislice T1 mapping may be achieved by combining the advantages of a model-based nonlinear inverse reconstruction, radial sampling, parallel imaging, and compressed sensing.
引用
收藏
页数:8
相关论文
共 46 条
[21]  
Lugauer F, 2018, MAGN RESON MATER PHY, V31, P399, DOI 10.1007/s10334-017-0670-8
[22]   Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement [J].
Moon, James C. ;
Messroghli, Daniel R. ;
Kellman, Peter ;
Piechnik, Stefan K. ;
Robson, Matthew D. ;
Ugander, Martin ;
Gatehouse, Peter D. ;
Arai, Andrew E. ;
Friedrich, Matthias G. ;
Neubauer, Stefan ;
Schulz-Menger, Jeanette ;
Schelbert, Erik B. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2013, 15
[23]   Influence of RF Spoiling on the Stability and Accuracy of T1 Mapping Based on Spoiled FLASH With Varying Flip Angles [J].
Preibisch, C. ;
Deichmann, R. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (01) :125-135
[24]   Model-Based Reconstruction for T1 Mapping Using Single-Shot Inversion-Recovery Radial FLASH [J].
Roeloffs, Volkert ;
Wang, Xiaoqing ;
Sumpf, Tilman J. ;
Untenberger, Markus ;
Voit, Dirk ;
Frahm, Jens .
INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 2016, 26 (04) :254-263
[25]   Spoiling without additional gradients: Radial FLASH MRI with randomized radiofrequency phases [J].
Roeloffs, Volkert ;
Voit, Dirk ;
Frahm, Jens .
MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (05) :2094-2099
[26]   A new method for fast multislice T1 mapping [J].
Shah, NJ ;
Zaitsev, M ;
Steinhoff, S ;
Zilles, K .
NEUROIMAGE, 2001, 14 (05) :1175-1185
[27]   Model-Based Nonlinear Inverse Reconstruction for T2 Mapping Using Highly Undersampled Spin-Echo MRI [J].
Sumpf, Tilman J. ;
Uecker, Martin ;
Boretius, Susann ;
Frahm, Jens .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 34 (02) :420-428
[28]   Model-Based Reconstruction for Real-Time Phase-Contrast Flow MRI: Improved Spatiotemporal Accuracy [J].
Tan, Zhengguo ;
Roeloffs, Volkert ;
Voit, Dirk ;
Joseph, Arun A. ;
Untenberger, Markus ;
Merboldt, K. Dietmar ;
Frahm, Jens .
MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (03) :1082-1093
[29]   Model-Based Acceleration of Look-Locker T1 Mapping [J].
Tran-Gia, Johannes ;
Wech, Tobias ;
Bley, Thorsten ;
Koestler, Herbert .
PLOS ONE, 2015, 10 (04)
[30]   Optimized and combined T1 and B1 mapping technique for fast and accurate T1 quantification in contrast-enhanced abdominal MRI [J].
Treier, Reto ;
Steingoetter, Andreas ;
Fried, Michael ;
Schwizer, Werner ;
Boesiger, Peter .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (03) :568-576