Design and application of a flow cell for carbon-film based electrochemical enzyme biosensors

被引:28
作者
Barsan, Madalina M.
Klincar, Janja
Batic, Martin
Brett, Christopher M. A. [1 ]
机构
[1] Univ Coimbra, Dept Quim, P-3004535 Coimbra, Portugal
[2] Univ Ljubljana, Food Sci & Technol Dept, Biotech Fac, Ljubljana 1000, Slovenia
[3] Res Inst Viticulture & Enol, Minist Agr & Rural Dev, H-6001 Kecskemet, Hungary
关键词
flow cell; electrochemical enzyme biosensor; enzyme immobilisation; sol-gel; glucose determination; grape must;
D O I
10.1016/j.talanta.2006.08.032
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow cell has been designed for use with an electrochemical enzyme biosensor, based on low-cost carbon-film electrodes. Three types of mediators were used: cobalt and copper hexacyanoferrates and poly(neutral red) (PNR), covered with glucose oxidase (GOx) immobilised by cross-linking with glutaraldehyde in the presence of bovine serum albumin or inside a oxysilane sol-gel network. Mixtures of sol-gel precursors were made from 3-aminopropyl-triethoxysilane (APTOS) together with methyltrimethoxysilane (MTMOS), methyltriethoxysilane (MTEOS), tetraethyloxysilane (TEOS) or 3-glycidoxypropyl-trimethoxysilane (GOPMOS), and the best chosen for encapsulation. Optimisation in batch mode, using amperometric detection at fixed potential, showed the PNR-GOx modified carbon-film electrodes to be best for flow analysis for both glutaraldehyde and sol-gel enzyme immobilisation. Both types of enzyme electrode were tested under flow conditions and the reproducibility and stability of the biosensors were evaluated. The biosensors were used for fermentation monitoring of glucose in grape must and interference studies were also performed. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:1893 / 1900
页数:8
相关论文
共 39 条
  • [1] DYE-MODIFIED ELECTRODES FOR PHOTOGALVANIC CELLS
    BAULDREAY, JM
    ARCHER, MD
    [J]. ELECTROCHIMICA ACTA, 1983, 28 (11) : 1515 - 1522
  • [2] Electrochemical behaviour of poly(neutral red) on an ITO electrode
    Benito, D
    Garcia-Jareno, JJ
    Navarro-Laboulais, J
    Vicente, F
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1998, 446 (1-2): : 47 - 55
  • [3] Brett CMA, 2001, ELECTROANAL, V13, P765, DOI 10.1002/1521-4109(200105)13:8/9<765::AID-ELAN765>3.0.CO
  • [4] 2-3
  • [5] Use of the terms "recovery" and "apparent recovery" in analytical procedures - (IUPAC recommendations 2002)
    Burns, DT
    Danzer, K
    Townshend, A
    [J]. PURE AND APPLIED CHEMISTRY, 2002, 74 (11) : 2201 - 2205
  • [6] Biosensors for life quality -: Design, development and applications
    Castillo, J
    Gáspár, S
    Leth, S
    Niculescu, M
    Mortari, A
    Bontidean, I
    Soukharev, V
    Dorneanu, SA
    Ryabov, AD
    Csöregi, E
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2004, 102 (02): : 179 - 194
  • [7] Application of amperometric sol-gel biosensor to flow injection determination of glucose
    Couto, CMCM
    Araújo, AN
    Montenegro, MCBSM
    Rohwedder, J
    Raimundo, I
    Pasquini, C
    [J]. TALANTA, 2002, 56 (06) : 997 - 1003
  • [8] Carbon film electrodes for oxidase-based enzyme sensors in food analysis
    De Luca, S
    Florescu, M
    Ghica, ME
    Lupu, A
    Palleschi, G
    Brett, CMA
    Compagnone, D
    [J]. TALANTA, 2005, 68 (02) : 171 - 178
  • [9] Development of biosensors based on hexacyanoferrates
    de Mattos, IL
    Gorton, L
    Laurell, T
    Malinauskas, A
    Karyakin, AA
    [J]. TALANTA, 2000, 52 (05) : 791 - 799
  • [10] Characterization of carbon film electrodes for electroanalysis by electrochemical impedance
    Filipe, OMS