What's floating on my plasma?

被引:8
作者
Cadamuro, Janne [1 ]
Wiedemann, Helmut [1 ]
Felder, Thomas K. [1 ]
Mrazek, Cornelia [1 ]
Kipman, Ulrike [2 ]
Hannes, Oberkofler [1 ]
Haschke-Becher, Elisabeth [1 ]
机构
[1] Paracelsus Med Univ, Dept Lab Med, Salzburg, Austria
[2] UT SPSS Stat, Hallein, Austria
关键词
case report; white particulate matter; lithium-heparin tubes; separator gel; WHITE PARTICULATE MATTER; COMPONENTS;
D O I
10.11613/BM.2017.047
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We report on a preanalytical issue we encountered during routine clinical chemistry analyses, potentially leading to deviated analysis results and believe that it might help other laboratories to overcome similar problems. In a heparin-gel tube we measured an implausible glucose value of 0.06 mmol/L. Re-measurement of the same sample resulted in a glucose value of 5.4 mmol/L. After excluding an analytical error, we inspected the sample closer and found a white material as well as fatty droplets floating on the surface of the plasma tube. Evaluation of these structures revealed that the white particulate matter (WPM) consisted of fibrinogen, platelets and leukocytes and the fatty droplets most probably originated from the separator gel. We concluded that these structures formed a temporary clot in the instruments pipetting needle thereby altering the sampling volume and subsequently the measured glucose value. The formation of WPM might be attributable to high speed centrifugation, high cholesterol levels, the gel formulation or a combination of several issues such as temperature, heparin concentration, pH and patient-specific factors. The gel droplets were most probably caused by an aberrant gel formulation in combination with an improper storage of the empty tubes on the wards prior to phlebotomy. After adding an additional instrument cleansing cycle and changing to another batch of heparin tubes the problems could be significantly reduced.
引用
收藏
页码:430 / 433
页数:4
相关论文
共 8 条
[1]   Roche IFCC methods for lactate dehydrogenase tested for duplicate errors with Greiner and Becton-Dickinson lithium-heparin and Greiner serum samples [J].
Dimeski, G ;
Badrick, T ;
Flatman, R ;
Ormiston, B .
CLINICAL CHEMISTRY, 2004, 50 (12) :2391-2392
[2]   Description and investigation of white particulate matter in additive solution-1 red blood cell units [J].
Hillyer, CD ;
Roback, JD ;
Hillyer, KL ;
Josephson, CD ;
Page, PL .
TRANSFUSION, 2004, 44 (07) :977-983
[3]   Rapid evaluation of risk of white particulate matter in blood components by a statewide survey of transfusion reactions [J].
Iwamoto, M ;
Curns, AT ;
Blake, PA ;
Jernigan, DB ;
Holman, RC ;
Lance-Parker, SE ;
Chamberland, ME ;
Kuehnert, MJ .
TRANSFUSION, 2004, 44 (07) :967-972
[4]   White particulate matter: report of the ad hoc industry review group [J].
McCullough, F ;
Dodd, R ;
Gilcher, R ;
Murphy, S ;
Sayers, M .
TRANSFUSION, 2004, 44 (07) :1112-1118
[5]   Factors affecting the formation of white particulate matter in red blood cell components [J].
Patel, HD ;
Byrne, KM ;
Horne, M ;
Leitman, SE ;
Stroncek, DE .
TRANSFUSION, 2005, 45 (07) :1127-1132
[6]   White particulate matter found in blood collection bags consist of platelets and leukocytes [J].
Rentas, FJ ;
Macdonald, VW ;
Rothwell, SW ;
McFaul, SJ ;
Asher, LV ;
Kennedy, AM ;
Hmel, PJ ;
Meledandri, CJ ;
Salata, JM ;
Harman, RW ;
Reid, TJ .
TRANSFUSION, 2004, 44 (07) :959-966
[7]  
Sachdev Suchet, 2011, Asian J Transfus Sci, V5, P175, DOI 10.4103/0973-6247.83247
[8]   White particulate matter: this time too much precaution [J].
Simon, TL .
TRANSFUSION, 2004, 44 (07) :956-958