In Vivo Verification of Implantable Antennas Using Rats as Model Animals

被引:64
作者
Karacolak, Tutku [1 ]
Cooper, Robert [2 ]
Butler, James [2 ]
Fisher, Stephen [2 ]
Topsakal, Erdem [1 ]
机构
[1] Mississippi State Univ, Dept Elect & Comp Engn, Mississippi State, MS 39762 USA
[2] Mississippi State Univ, Coll Vet Med, Mississippi State, MS 39762 USA
来源
IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS | 2010年 / 9卷
关键词
Implantable antenna; in vivo; Industrial; Scientific; and Medical (ISM) band; Medical Implant Communications Service (MICS) band; DIELECTRIC-PROPERTIES; MICROSTRIP ANTENNA; TELEMETRY; DESIGN; SKIN; POTENTIOSTAT; TISSUE;
D O I
10.1109/LAWP.2010.2048693
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The main objective of this study is to test in vivo a small-size dual-band implantable antenna operating in Medical Implant Communications Service (MICS) (402-405 MHz) and Industrial, Scientific, and Medical (ISM) (2.4-2.48 GHz) bands and to investigate the effects of live tissue on the antenna performance. To do so, we have designed and fabricated three identical antennas that were surgically implanted into rats at Mississippi State University's (MSU) College of Veterinary Medicine. X-ray images are also taken to ensure the proper placement of the antenna within the body. Results regarding return loss measurements and the performance change of the antenna are presented. We have also investigated possible candidate materials that can be used for encasing of the antenna for biocompatibility.
引用
收藏
页码:334 / 337
页数:4
相关论文
共 13 条
[1]   Subminiature implantable potentiostat and modified commercial telemetry device for remote glucose monitoring [J].
Beach, RD ;
Von Küster, F ;
Moussy, F .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (06) :1239-1245
[2]   Towards a miniature implantable in vivo telemetry monitoring system dynamically configurable as a potentiostat or galvanostat for two- and three-electrode biosensors [J].
Beach, RD ;
Conlan, RW ;
Godwin, MC ;
Moussy, F .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2005, 54 (01) :61-72
[3]   Impedance matching and implementation of planar space-filling dipoles as intraocular implanted antennas in a retinal prosthesis [J].
Gosalia, K ;
Humayun, MS ;
Lazzi, G .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2005, 53 (08) :2365-2373
[4]   Dielectric properties of blood: an investigation of temperature dependence [J].
Jaspard, F ;
Nadi, M .
PHYSIOLOGICAL MEASUREMENT, 2002, 23 (03) :547-554
[5]   Design of a dual-band implantable antenna and development of skin mimicking gels for continuous glucose monitoring [J].
Karacolak, Tutku ;
Hood, Aaron Z. ;
Topsakal, Erdem .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2008, 56 (04) :1001-1008
[6]   Electrical Properties of Rat Skin and Design of Implantable Antennas for Medical Wireless Telemetry [J].
Karacolak, Tutku ;
Cooper, Robert ;
Topsakal, Erdem .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2009, 57 (09) :2806-2812
[7]   Implanted antennas inside a human body: Simulations, designs, and characterizations [J].
Kim, J ;
Rahmat-Samii, Y .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2004, 52 (08) :1934-1943
[8]  
KWAK S, 2005, P AS PAC MICR C DEC, V1
[9]   Ultrawideband temperature-dependent dielectric properties of animal liver tissue in the microwave frequency range [J].
Lazebnik, M ;
Converse, MC ;
Booske, JH ;
Hagness, SC .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (07) :1941-1955
[10]   Changes in the dielectric properties of rat tissue as a function of age at microwave frequencies [J].
Peyman, A ;
Rezazadeh, AA ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (06) :1617-1629