Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

被引:55
作者
Lewis, David F. [1 ]
Chan, Maria F. [2 ]
机构
[1] RC Film Consulting LLC, 54 Benedict Rd, Monroe, CT 06468 USA
[2] Mem Sloan Kettering Canc Ctr, 136 Mt View Blvd, Basking Ridge, NJ 07920 USA
关键词
radiochromic film dosimetry; QA; IMRT; radiotherapy; lateral response; FLAT-BED SCANNERS; RADIOCHROMIC FILM; NONUNIFORMITY CORRECTION; DOSIMETRY; CALIBRATION;
D O I
10.1118/1.4939226
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners. Owing its lower sensitivity, EBT-XD film requires higher dose to reach the same response as EBT3, resulting in lesser impact of the lateral response artifact. For doses >10 Gy, the slopes of the EBT-XD red and green channel dose response curves are greater than the corresponding ones for EBT3. For these two reasons, the authors prefer EBT-XD for doses exceeding about 10 Gy. 2016 American Association of Physicists in Medicine.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 14 条
  • [1] Calibration of EBT2 film by the PDD method with scanner non-uniformity correction
    Chang, Liyun
    Chui, Chen-Shou
    Ding, Hueisch-Jy
    Hwang, Ing-Ming
    Ho, Sheng-Yow
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (18) : 5875 - 5887
  • [2] Incorporating a Lateral Scan Effect Correction in a EBT3 Calibration Protocol
    Crijns, W.
    Slagmolen, P.
    Maes, F.
    Haustermans, K.
    Van den Heuvel, F.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 4009 - 4009
  • [3] Clinical use of EBT model Gafchromic™ film in radiotherapy
    Fiandra, Christian
    Ricardi, Umberto
    Ragona, Riccardo
    Anglesio, Silvia
    Giglioli, Francesca Romana
    Calamia, Elisa
    Lucio, Francesco
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4314 - 4319
  • [4] Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry
    Lewis, David
    Chan, Maria F.
    [J]. MEDICAL PHYSICS, 2015, 42 (01) : 416 - 429
  • [5] Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC® film
    Lynch, Bart D.
    Kozelka, Jakub
    Ranade, Manisha K.
    Li, Jonathan G.
    Simon, William E.
    Dempsey, James F.
    [J]. MEDICAL PHYSICS, 2006, 33 (12) : 4551 - 4556
  • [6] Model selection for radiochromic film dosimetry
    Mendez, I.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (10) : 4089 - 4104
  • [7] Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure
    Menegotti, L.
    Delana, A.
    Martignano, A.
    [J]. MEDICAL PHYSICS, 2008, 35 (07) : 3078 - 3085
  • [8] Multichannel film dosimetry with nonuniformity correction
    Micke, Andre
    Lewis, David F.
    Yu, Xiang
    [J]. MEDICAL PHYSICS, 2011, 38 (05) : 2523 - 2534
  • [9] Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry
    Paelinck, L.
    De Neve, W.
    DeWagter, C.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (01) : 231 - 242
  • [10] A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry
    Poppinga, D.
    Schoenfeld, A. A.
    Doerner, K. J.
    Blanck, O.
    Harder, D.
    Poppe, B.
    [J]. MEDICAL PHYSICS, 2014, 41 (02) : 021707