Traceable laser power measurements of diode laser radiation in the near infrared

被引:3
作者
Brandt, Friedhelm [1 ]
Kueck, Stefan [1 ]
Gruetz, Andre [2 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] LIMO Lissotschenko Mikroopt GmbH, Dortmund, Germany
来源
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA | 2010年 / 25卷 / 01期
关键词
Beam Size; Uncertainty Budget; Device Under Test; Spectral Responsivity; Standard Detector;
D O I
10.1007/s12647-010-0006-x
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report a new calibration setup for laser radiometry at the Physikalisch-Technische Bundesanstalt, the German National Metrology Institute. Measurements of laser diode power of free beam diode lasers in the near infrared spectral range at a wavelength of 808 nm for powers up to 250 W and at wavelengths of 915 nm, 940 nm, and 980 nm for laser powers up to 25 W have been established. The calibration setup, the standard detector, the uncertainty budget and first calibration results will be presented and discussed. The standard uncertainty of the HLR302 standard detector is 0.2%. This uncertainty might be the main contribution to the overall uncertainty in customer calibrations, depending on the quality of the transfer detectors.
引用
收藏
页码:29 / 35