Motional Stark effect diagnostic on TEXTOR

被引:16
作者
Jakubowska, K
De Bock, M
Jaspers, R
von Hellermann, M
Shmaenok, L
机构
[1] EURATOM, FOM Inst Plasma Phys Rijnhuizen, NL-3430 BE Nieuwegein, Netherlands
[2] AF Ioffe Phys Tech Inst, St Petersburg 194021, Russia
[3] Phystex, NL-6290 AB Vaals, Netherlands
关键词
D O I
10.1063/1.1784535
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A motional Stark effect diagnostic at the tokamak TEXTOR has been constructed and brought recently into operation. In contrast to diagnostics used on other tokamaks, this diagnostic reveals the direction of the magnetic field from the intensity ratio of the pi and sigma components of the emitted Balmer-alpha and not from a polarization measurement of a single line. Moreover, the complete spectrum is measured which allows determining the radial position of the measurement, and in principle the radial electric field. (C) 2004 American Institute of Physics.
引用
收藏
页码:3475 / 3477
页数:3
相关论文
共 50 条
  • [41] A Spectrally Resolved Motional Stark Effect Diagnostic for the TJ-II Stellarator
    McCarthy, K. J.
    Panadero, N.
    Lopez-Fraguas, A.
    Hernandez, J.
    van Milligen, B.
    CONTRIBUTIONS TO PLASMA PHYSICS, 2015, 55 (06) : 459 - 469
  • [42] Four-dimensional calibration turntable of the motional Stark effect diagnostic on EAST
    Huang, X.
    Liu, D. M.
    Liu, C.
    Fu, J.
    Wan, B. N.
    Lyu, B.
    Wu, Z. W.
    Holcomb, C. T.
    Ko, J.
    Rowan, W. L.
    Huang, H.
    Miao, G. Z.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (10)
  • [43] Robotic calibration of the motional Stark effect diagnostic on Alcator C-Mod
    Mumgaard, Robert T.
    Scott, Steven D.
    Ko, Jinseok
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (05)
  • [44] Initial operation of a newly developed multichord motional Stark effect diagnostic in KSTAR
    Chung, J.
    Ko, J.
    Wi, H.
    Messmer, M.
    Schenkelaars, S.
    Scheffer, M.
    Jaspers, R. J. E.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11)
  • [45] Motional Stark effect diagnostics for KSTAR
    Chung, J.
    Ko, J.
    Howard, J.
    Michael, C.
    von Nessi, G.
    Thorman, A.
    De Bock, M. F. M.
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (08) : 1257 - 1260
  • [46] Motional Stark effect diagnostics for KSTAR
    J. Chung
    J. Ko
    J. Howard
    C. Michael
    G. von Nessi
    A. Thorman
    M. F. M. De Bock
    Journal of the Korean Physical Society, 2014, 65 : 1257 - 1260
  • [47] MOTIONAL STARK-EFFECT POLARIMETRY FOR A CURRENT PROFILE DIAGNOSTIC IN DIII-D
    WROBLEWSKI, D
    BURRELL, KH
    LAO, LL
    POLITZER, P
    WEST, WP
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) : 2988 - 2988
  • [48] MAGNETIC-FIELD PITCH ANGLE DIAGNOSTIC USING THE MOTIONAL STARK-EFFECT
    LEVINTON, FM
    GAMMEL, GM
    KAITA, R
    KUGEL, HW
    ROBERTS, DW
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) : 2914 - 2919
  • [49] A digital lock-in upgrade of the motional Stark effect diagnostic on DIII-D
    King, J. D.
    Makowski, M. A.
    Holcomb, C. T.
    Allen, S. L.
    Geer, R.
    Meyer, W. H.
    Hill, D. N.
    Pham, D.
    Morse, E. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (10)
  • [50] q profile measurements with the motional Stark effect diagnostic in the DIII-D tokamak
    Rice, BW
    FUSION ENGINEERING AND DESIGN, 1997, 34-35 : 135 - 142