Evaluation and comparison of the kinetic performance of ultra-high performance liquid chromatography and high-performance liquid chromatography columns in hydrophilic interaction and reversed-phase liquid chromatography conditions

被引:32
|
作者
Song, Huiying [1 ]
Adams, Erwin [1 ]
Desmet, Gert [2 ]
Cabooter, Deirdre [1 ]
机构
[1] Katholieke Univ Leuven, Dept Pharmaceut & Pharmacol Sci, Leuven, Belgium
[2] Vrije Univ Brussel, Dept Chem Engn, B-1050 Brussels, Belgium
关键词
HILIC; RPLC; Kinetic plot; Porosity; Flow resistance; PARTICLE-SIZE DISTRIBUTION; SUB-2; MU-M; MASS-SPECTROMETRY; DEAD-VOLUME; SEPARATION; RETENTION; QUANTIFICATION; ADSORPTION; EPHEDRINES; MECHANISM;
D O I
10.1016/j.chroma.2014.10.002
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An intrinsic performance comparison is made of the reduction in analysis time that can be obtained when switching from HPLC to UHPLC column formats in HILIC and reversed-phase conditions. A detailed overview of the packing characteristics of both stationary phase types is given first. It is demonstrated that HILIC columns demonstrate higher external porosity values than their reversed-phase counterparts resulting in lower flow resistance values. Column total porosity values determined from the elution time of a small marker molecule are shown to depend strongly on the composition of the mobile phase. To omit errors that might arise from an over- or underestimation of the column void time, all plate height and kinetic plot data are therefore expressed as a function of the interstitial velocity. Although only a limited number of columns are evaluated in this study, it is shown that the column efficiency of the HILIC columns is lower than that of their reversed-phase counterparts, at least for the compounds evaluated here. Despite this lower efficiency, the kinetic performance of both stationary phase types is similar, due to the much lower viscosity of the mobile phases typically used in HILIC conditions. Finally, it is demonstrated that a similar, yet slightly larger reduction in analysis time can be obtained when switching from HPLC column formats to UHPLC formats in HILIC compared to reversed-phase conditions. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:83 / 91
页数:9
相关论文
共 50 条
  • [31] Determination of Enantiomer in Tofacitinib Citrate Using Reversed-Phase Chiral High-Performance Liquid Chromatography
    Wang, Xue
    Jin, Bo
    Wang, Zhijun
    Guo, Kaijing
    Zhang, Tingting
    Ma, Chen
    SEPARATIONS, 2024, 11 (03)
  • [32] Computer-assisted transfer of programmed elutions in reversed-phase high-performance liquid chromatography
    Garcia-Lavandeira, J.
    Martinez-Pontevedra, J. A.
    Lores, M.
    Cela, R.
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1128 (1-2) : 17 - 26
  • [33] Reversed-phase high-performance liquid chromatography of the stable electrophoretic fractions of soil humic acids
    Trubetskoi, O. A.
    Trubetskaya, O. E.
    EURASIAN SOIL SCIENCE, 2015, 48 (02) : 148 - 156
  • [34] DETERMINATION OF SULFUR IN PHARMACEUTICAL PREPARATIONS USING REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    KUROSAKA, K
    KUCHIKI, A
    NAKAGAWA, H
    CHEMICAL & PHARMACEUTICAL BULLETIN, 1991, 39 (08) : 2138 - 2139
  • [35] A rapid analytical method of major milk proteins by reversed-phase high-performance liquid chromatography
    Ma, Lu
    Yang, Yongxin
    Chen, Jingting
    Wang, Jiaqi
    Bu, Dengpan
    ANIMAL SCIENCE JOURNAL, 2017, 88 (10) : 1623 - 1628
  • [36] Analysis of Grape Proteins from Wines by Perfusion Reversed-Phase High-Performance Liquid Chromatography
    Rodriguez-Nogales, Jose M.
    Vila-Crespo, Josefina
    Fernandez-Fernandez, Encarnacion
    FOOD ANALYTICAL METHODS, 2013, 6 (04) : 1234 - 1243
  • [37] Peptide mapping by reversed-phase high-performance liquid chromatography employing silica rod monoliths
    Hennessy, TP
    Boysen, RI
    Huber, MI
    Unger, KK
    Hearn, MTW
    JOURNAL OF CHROMATOGRAPHY A, 2003, 1009 (1-2) : 15 - 28
  • [38] Preparation of monolithic silica columns for high-performance liquid chromatography
    Nunez, Oscar
    Nakanishi, Kazuki
    Tanaka, Nobuo
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1191 (1-2) : 231 - 252
  • [39] Determination of organic acids in alcoholic and nonalcoholic beverages by reversed-phase high-performance liquid chromatography
    Amelin, V. G.
    Podkolzin, I. V.
    Tretiakov, A. V.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2012, 67 (03) : 262 - 268
  • [40] Rapid determination of human globin chains using reversed-phase high-performance liquid chromatography
    Wan, Jun-Hui
    Tian, Pei-Ling
    Luo, Wei-Hao
    Wu, Bing-Yi
    Xiong, Fu
    Zhou, Wan-Jun
    Wei, Xiang-Cai
    Xu, Xiang-Min
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2012, 901 : 53 - 58