Two- to five-year follow-up of 78 patients after treatment with the Flow Redirection Endoluminal Device

被引:13
作者
Luecking, Hannes [1 ]
Doerfler, Arnd [1 ]
Goelitz, Philipp [1 ]
Hoelter, Philip [1 ]
Engelhorn, Tobias [1 ]
Lang, Stefan [1 ]
机构
[1] Univ Erlangen Nurnberg, Dept Neuroradiol, Erlangen, Germany
关键词
Cerebral aneurysm; flow diversion; Flow Redirection Endoluminal Device; PIPELINE EMBOLIZATION DEVICE; CEREBRAL ANEURYSMS; INTRACRANIAL ANEURYSMS; COIL EMBOLIZATION; ENDOVASCULAR TREATMENT; DIVERTOR DEVICES; SAFETY; STENT; COMPLICATIONS; DIVERSION;
D O I
10.1177/1591019919878551
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and purpose Flow-diverter stents are well-established for the treatment of cerebral aneurysms. Flow Redirection Endoluminal Device differs from other flow-diverter stents by its dual-layer design and has proved equality to other devices in numerous short-term surveys. However, follow-up data covering substantially more than one year are still limited for this device. We present our long-term experience with Flow Redirection Endoluminal Device. Materials and methods Seventy-eight patients harboring distal internal carotid artery (91%) or vertebrobasilar (9%) cerebral aneurysms treated with Flow Redirection Endoluminal Device with or without adjunctive coiling met the inclusion criteria. All cases were evaluated for aneurysm occlusion (according to Modified Raymond Roy Classification, MRRC), for flow-diverter stents patency and configuration and for procedure- and device-related morbidity and mortality. Results Mean follow-up interval was 36.99.5 months (<30 months: n=18; 31-42 months: n=31;>42 months: n=24). Total and subtotal aneurysm occlusion after six months was assessed in 92.0% (MRRC1=77.3%, MRRC2=14.7%, MRRC3a=2.7%, MRRC3b=4.1%) and increased to 95.9% (MRRC1=90.5%, MRRC2=5.4%, MRRC3a=2.7%). There was one case of aneurysm growth requiring early re-treatment. Procedure-related morbidity was observed in three cases (3.8%; one transient hemiparesis, one suspected foreign-body reaction, and one micro-wire perforation). There was no procedure- or device-related mortality. In-stent stenosis due to intimal hyperplasia was observed in two cases and fish-mouthing in three cases. Conclusions Our long-term data covering two to five years after flow diversion confirm that Flow Redirection Endoluminal Device is a safe and effective device for the treatment of cerebral aneurysms with progressive high aneurysm occlusion rates; recurrence rates were very low. Overall device-related morbidity was low and was not observed later than six months after intervention.
引用
收藏
页码:38 / 44
页数:7
相关论文
共 25 条
[1]   A numerical framework for the mechanical analysis of dual-layer stents in intracranial aneurysm treatment [J].
Alherz, Ali I. ;
Tanweer, Omar ;
Flamini, Vittoria .
JOURNAL OF BIOMECHANICS, 2016, 49 (12) :2420-2427
[2]   Long-Term Clinical and Angiographic Outcomes Following Pipeline Embolization Device Treatment of Complex Internal Carotid Artery Aneurysms: Five-Year Results of the Pipeline for Uncoilable or Failed Aneurysms Trial [J].
Becske, Tibor ;
Brinjikji, Waleed ;
Potts, Matthew B. ;
Kallmes, David F. ;
Shapiro, Maksim ;
Moran, Christopher J. ;
Levy, Elad I. ;
McDougall, Cameron G. ;
Szikora, Istvan ;
Lanzino, Giuseppe ;
Woo, Henry H. ;
Lopes, Demetrius K. ;
Siddiqui, Adnan H. ;
Albuquerque, Felipe C. ;
Fiorella, David J. ;
Saatci, Isil ;
Cekirge, Saruhan H. ;
Berez, Aaron L. ;
Cher, Daniel J. ;
Berentei, Zsolt ;
Marosfoi, Miklos ;
Nelson, Peter K. .
NEUROSURGERY, 2017, 80 (01) :40-48
[3]   Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical Trial [J].
Becske, Tibor ;
Kallmes, David F. ;
Saatci, Isil ;
McDougall, Cameron G. ;
Szikora, Istvn ;
Lanzino, Giuseppe ;
Moran, Christopher J. ;
Woo, Henry H. ;
Lopes, Demetrius K. ;
Berez, Aaron L. ;
Cher, Daniel J. ;
Siddiqui, Adnan H. ;
Levy, Elad I. ;
Albuquerque, Felipe C. ;
Fiorella, David J. ;
Berentei, Zsolt ;
Marosfoi, Miklos ;
Cekirge, Saruhan H. ;
Nelson, Peter K. .
RADIOLOGY, 2013, 267 (03) :858-868
[4]   Postprocedural, midterm, and long-term results of cerebral aneurysms treated with flow-diverter devices: 7-year experience at a single center [J].
Briganti, Francesco ;
Leone, Giuseppe ;
Cirillo, Luigi ;
de Divitiis, Oreste ;
Solari, Domenico ;
Cappabianca, Paolo .
NEUROSURGICAL FOCUS, 2017, 42 (06)
[5]   Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review [J].
Briganti, Francesco ;
Leone, Giuseppe ;
Marseglia, Mariano ;
Mariniello, Giuseppe ;
Caranci, Ferdinando ;
Brunetti, Arturo ;
Maiuri, Francesco .
NEURORADIOLOGY JOURNAL, 2015, 28 (04) :365-375
[6]   Long-Term Follow-Up Results following Elective Treatment of Unruptured Intracranial Aneurysms with the Pipeline Embolization Device [J].
Chiu, A. H. Y. ;
Cheung, A. K. ;
Wenderoth, J. D. ;
De Villiers, L. ;
Rice, H. ;
Phatouros, C. C. ;
Singh, T. P. ;
Phillips, T. J. ;
McAuliffe, W. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (09) :1728-1734
[7]   Delayed complications after flow-diverter stenting: Reactive in-stent stenosis and creeping stents [J].
Cohen, Jose E. ;
Gomori, John Moshe ;
Moscovici, Samuel ;
Leker, Ronen R. ;
Itshayek, Eyal .
JOURNAL OF CLINICAL NEUROSCIENCE, 2014, 21 (07) :1116-1122
[8]   Enhancing Brain Lesions after Endovascular Treatment of Aneurysms [J].
Cruz, J. P. ;
Marotta, T. ;
O'Kelly, C. ;
Holtmannspoetter, M. ;
Saliou, G. ;
Willinsky, R. ;
Krings, T. ;
Agid, R. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (10) :1954-1958
[9]   Long-Term Effect of Flow Diversion on Large and Giant Aneurysms: MRI-DSA Clinical Correlation Study [J].
John, Seby ;
Bain, Mark D. ;
Hussain, M. Shazam ;
Bauer, Andrew M. ;
Toth, Gabor .
WORLD NEUROSURGERY, 2016, 93 :60-66
[10]   Effect of Structural Remodeling (Retraction and Recoil) of the Pipeline Embolization Device on Aneurysm Occlusion Rate [J].
Jou, L. -D. ;
Mitchell, B. D. ;
Shaltoni, H. M. ;
Mawad, M. E. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (09) :1772-1778