Melodic algorithms for pulse oximetry to allow audible discrimination of abnormal systolic blood pressures

被引:1
作者
Chima, Ranjit S. [1 ]
Ortega, Rafael [2 ]
Connor, Christopher W. [2 ]
机构
[1] Texas Tech Univ, Hlth Sci Ctr, Lubbock, TX 79430 USA
[2] Boston Med Ctr, Dept Anesthesiol, Boston, MA 02118 USA
关键词
Blood pressure; Alarm; Melody; Pulse oximeter; Monitoring; ALARMS; INFORMATION;
D O I
10.1007/s10877-014-9558-6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
An anesthesiologist must remain vigilant of the patient's clinical status, incorporating many independent physiological measurements. Oxygen saturation and heart rate are represented by continuous audible tones generated by the pulse oximeter, a mandated monitoring device. Other important clinical parameters-notably blood pressure-lack any audible representation beyond arbitrarily-configured threshold alarms. Attempts to introduce further continuous audible tones have apparently foundered; the complexity and interaction of these tones have exceeded the ability of clinicians to interpret them. Instead, we manipulate the tonal and rhythmic structure of the accepted pulse oximeter tone pattern melodically. Three melodic algorithms were developed to apply tonal and rhythmic variations to the continuous pulse oximeter tone, dependent on the systolic blood pressure. The algorithms distort the original audible pattern minimally, to facilitate comprehension of both the underlying pattern and the applied variations. A panel of anesthesia practitioners (attending anesthesiologists, residents and nurse anesthetists) assessed these algorithms in characterizing perturbations in cardiopulmonary status. Twelve scenarios, incorporating combinations of oxygen desaturation, bradycardia, tachycardia, hypotension and hypertension, were tested. A rhythmic variation in which additional auditory information was conveyed only at halftime intervals, with every other "beat" of the pulse oximeter, was strongly favored. The respondents also strongly favored the use of musical chords over single tones. Given three algorithms of tones embedded in the pulse oximeter signal, anesthesiologists preferred a melodic tone to signal a significant change in blood pressure.
引用
收藏
页码:597 / 603
页数:7
相关论文
共 16 条
[1]   STATISTICS NOTES - DIAGNOSTIC-TESTS-1 - SENSITIVITY AND SPECIFICITY .3. [J].
ALTMAN, DG ;
BLAND, JM .
BRITISH MEDICAL JOURNAL, 1994, 308 (6943) :1552-1552
[2]   DIAGNOSTIC-TESTS-2 - PREDICTIVE VALUES .4. [J].
ALTMAN, DG ;
BLAND, JM .
BRITISH MEDICAL JOURNAL, 1994, 309 (6947) :102-102
[3]  
American Society of Anesthesiologists, 2011, STAND BAS AN MON
[4]  
Cage the Elephant (Musical group) and Copyright Collection (Library of Congress), 2009, CAG EL BACK WALLM VI
[5]  
Chandra D, 2006, CAN J ANAESTH, V53, P26, DOI 10.1007/BF03021524
[6]   Triggering of systolic arterial pressure alarms using statistics-based versus threshold alarms [J].
Connor, C. W. ;
Gohil, B. ;
Harrison, M. J. .
ANAESTHESIA, 2009, 64 (02) :131-135
[7]  
Connor CW, 2013, EXECUTABLE FILE WIND
[8]   Continuous auditory monitoring - how much information do we register? [J].
Craven, RM ;
McIndoe, AK .
BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (05) :747-749
[9]   Intra-operative monitoring - many alarms with minor impact [J].
de Man, F. R. ;
Greuters, S. ;
Boer, C. ;
Veerman, D. P. ;
Loer, S. A. .
ANAESTHESIA, 2013, 68 (08) :804-810
[10]   Statistics-based alarms from sequential physiological measurements [J].
Harrison, M. J. ;
Connor, C. W. .
ANAESTHESIA, 2007, 62 (10) :1015-1023