The performance of BD FACSPresto™ for CD4 T-cell count, CD4% and hemoglobin concentration test in Ethiopia

被引:6
作者
Gebremicael, Gebremedhin [1 ]
Belay, Yohanes [1 ]
Girma, Fitsum [1 ]
Abreha, Yemane [1 ]
Gebreegziabxier, Atsbeha [1 ]
Tesfaye, Simret [1 ]
Messele, Zelalem [2 ]
Assefa, Yibeltal [1 ]
Bellete, Bahrie [2 ]
Kassa, Desta [1 ]
Vojnov, Lara [3 ]
机构
[1] Ethiopian Publ Hlth Inst, HIV & TB Dis Res Directorate, Addis Ababa, Ethiopia
[2] Clinton Hlth Access Initiat Inc, Addis Ababa, Ethiopia
[3] Clinton Hlth Access Initiat, Boston, MA USA
来源
PLOS ONE | 2017年 / 12卷 / 04期
关键词
LYMPHOCYTES;
D O I
10.1371/journal.pone.0176323
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Introduction In Ethiopia, CD4+ T-cell counting is still required for all patients at baseline before antiretroviral therapy (ART) and to determine eligibility and follow-up of opportunistic infection prophylaxis. However, access to CD4+ T-cell count in rural health facilities remains a major challenge in Ethiopia like other resource-limited settings. Methodology Both capillary and venous blood was drawn from each of 325 study participant recruited in Addis Ababa and surroundings. The CD4+ T-cell count, CD4%, and hemoglobin (Hgb) were tested at one of the four study health facilities using capillary blood and BD FACSPresto (TM) device. These tests were also done at the national HIV reference laboratory, using venous blood with BD FACSCalibur (TM), Sysmex XT-1800i (TM), and BD FACSPresto (TM). Results BD FACSPresto (TM) had an absolute mean bias of -13.3 cells/mu l (-2.99%) and 28.3 cells/mu l (6.4%) using venous and capillary blood, respectively, compared with BD FACSCalibur (TM). The absolute CD4 assay on the BD FACSPresto (TM) had a regression coefficient (R-2) of 0.87 and 0.92 using capillary blood and venous blood samples, respectively, compared with BD FACSCalibur (TM). The percentage similarity of the BD FACSPresto (TM) using capillary and venous blood was 105.2% and 99.3%, respectively. The sensitivity of the FACSPresto (TM) using threshold of 500 cells/mu l for ART eligibility using capillary and venous blood was 87.9 and 94.3%, while the specificity was 91.4 and 83.8%, respectively. Furthermore, the BD FACSPresto (TM) had an absolute mean bias of -0.2 dl/mu l (0.0%) (95% LOA: -1.7, 1.3) and -0.59 dl/mu l (0.1%) (95% LOA: -1.49, 0.31) for Hgb using capillary and venous blood compared with the Sysmex XT-1800i (TM), respectively. Conclusion Our results showed acceptable agreement between the BD FACSPresto (TM) and BD FACS-Calibur (TM) for CD4+ T-cell counting and CD4%; and between the BD FACSPresto (TM) and Sysmex XT-1800i (TM) for measuring Hgb concentration.
引用
收藏
页数:15
相关论文
共 23 条
[1]  
[Anonymous], 2010, Antiretroviral therapy for HIV infection in adults and adolescents
[2]  
[Anonymous], JIAS
[3]  
[Anonymous], 2011, Traditional Medicines: Global Situation, Issues and Challenges, P1
[4]   Rapid Scale-Up of Antiretroviral Treatment in Ethiopia: Successes and System-Wide Effects [J].
Assefa, Yibeltal ;
Jerene, Degu ;
Lulseged, Sileshi ;
Ooms, Gorik ;
Van Damme, Wim .
PLOS MEDICINE, 2009, 6 (04)
[5]  
(BD) BD, 2016, BD FACSPRESTO TM NEA
[6]  
(BD) BD, 2007, BD FACSCALIBUR INSTR
[7]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[8]   The BD FACSPresto Point of Care CD4 Test Accurately Enumerates CD4+T Cell Counts [J].
Bwana, Priska ;
Vojnov, Lara ;
Adhiambo, Maureen ;
Akinyi, Catherine ;
Mwende, Joy ;
Prescott, Marta ;
Mwau, Matilu .
PLOS ONE, 2015, 10 (12)
[9]  
FMOH, 2013, GUID IMPL POC CD4 TE
[10]   CHARACTERIZATION OF T-LYMPHOCYTE SUBSET ALTERATIONS BY FLOW-CYTOMETRY IN HIV DISEASE [J].
GIORGI, JV .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1993, 677 :126-137