Perioperative Noninvasive Blood Pressure Monitoring

被引:27
作者
Kuck, Kai [1 ]
Baker, Philip D. [2 ]
机构
[1] Univ Utah, Sch Med, Dept Anesthesiol, 30 N 1900 E Room 3c444, Salt Lake City, UT 84132 USA
[2] Diagnost Imaging Northwest, Portland, OR USA
关键词
OSCILLOMETRIC MAXIMUM;
D O I
10.1213/ANE.0000000000002619
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The most commonly monitored variable for perioperative hemodynamic management is blood pressure. Several indirect noninvasive blood pressure monitoring techniques have been developed over the last century, including intermittent techniques such as auscultation (Riva-Rocci and Korotkoff) and oscillometry (Marey) and continuous techniques. With the introduction of automated noninvasive blood pressure devices in the 1970s, the oscillometric technique quickly became and remains the standard for automated, intermittent blood pressure measurement. It tends to estimate more extreme high and low blood pressures closer to normal than what invasive measurements indicate. The accuracy of the oscillometric maximum amplitude algorithm for estimating mean arterial pressure is affected by multiple factors, including the cuff size and shape, the shape of the arterial compliance curve and arterial pressure pulse, and pulse pressure itself. Additionally, the technique typically assumes a consistent arterial compliance and arterial pressure pulse, thus changes in arterial compliance and arrhythmias that lead to variation in the pressure pulse can affect accuracy. Volume clamping, based on the Penaz principle, and arterial tonometry provide continuous tracking of the arterial pressure pulse. The ubiquitous use of blood pressure monitoring is in contrast with the lack of evidence for optimal perioperative blood pressure targets.
引用
收藏
页码:408 / 411
页数:4
相关论文
共 15 条
[1]  
Association for the Advancement of Medical Instrumentation, 2009, 8106022009 ANSIAAMII, P1
[2]   Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure [J].
Baker, PD ;
Westenskow, DR ;
Kuck, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) :271-278
[3]   Blood Pressure Monitoring for the Anesthesiologist: A Practical Review [J].
Bartels, Karsten ;
Esper, Stephen A. ;
Thiele, Robert H. .
ANESTHESIA AND ANALGESIA, 2016, 122 (06) :1866-1879
[4]   Incidence of intraoperative hypotension as a function of the chosen definition - Literature definitions applied to a retrospective cohort using automated data collection [J].
Bijker, Jilles B. ;
van Klei, Wilton A. ;
Kappen, Teus H. ;
van Wolfswinkel, Leo ;
Moons, Karel G. M. ;
Kalkman, Cor J. .
ANESTHESIOLOGY, 2007, 107 (02) :213-220
[5]  
CANNESSON M, 2011, CRIT CARE, V15
[6]   THEORY OF THE OSCILLOMETRIC MAXIMUM AND THE SYSTOLIC AND DIASTOLIC DETECTION RATIOS [J].
DRZEWIECKI, G ;
HOOD, R ;
APPLE, H .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) :88-96
[7]   Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry [J].
Dueck, Ron ;
Goedje, Oliver ;
Clopton, Paul .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2012, 26 (02) :75-83
[8]   Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops [J].
Fortin, J. ;
Marte, W. ;
Grullenberger, R. ;
Hacker, A. ;
Habenbacher, W. ;
Heller, A. ;
Wagner, C. H. ;
Wach, P. ;
Skrabal, F. .
COMPUTERS IN BIOLOGY AND MEDICINE, 2006, 36 (09) :941-957
[9]  
Geddes LA., 1970, Am Heart J, V80, P435
[10]   A Systematic Review and Meta-Analysis on the Use of Preemptive Hemodynamic Intervention to Improve Postoperative Outcomes in Moderate and High-Risk Surgical Patients [J].
Hamilton, Mark A. ;
Cecconi, Maurizio ;
Rhodes, Andrew .
ANESTHESIA AND ANALGESIA, 2011, 112 (06) :1392-1402