Investigation of Apex Locators and Related Morphological Factors

被引:42
作者
Ding, Jiangfeng [1 ,2 ]
Gutmann, James L. [3 ]
Fan, Bing [1 ,2 ]
Lu, Yao [1 ,2 ]
Chen, Hao [1 ,2 ]
机构
[1] Wuhan Univ, Sch Stomatol, Minist Educ, State Key Lab Breeding Base Basic Sci Stomatol Hu, Wuhan 430079, Peoples R China
[2] Wuhan Univ, Hosp Stomatol, Minist Educ, Key Lab Oral Biomed, Wuhan 430079, Peoples R China
[3] Texas A&M Univ Syst, Hlth Sci Ctr, Dept Endodont, Baylor Coll Dent, Dallas, TX USA
基金
中国国家自然科学基金;
关键词
Electronic apex locator; Root ZX; Raypex; 5; Elements Apex Locator; morphological factor; ELEMENTS DIAGNOSTIC UNIT; WORKING LENGTH DETERMINATION; ROOT-CANAL LENGTH; IN-VITRO; VIVO ACCURACY; ZX; SIZE; PRECISION;
D O I
10.1016/j.joen.2010.04.006
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The purpose of this study was to investigate the ability of three electronic apex locators (EALs) to detect the minor foramen and morphological influencing factors relative to working length determination. Methods: Three hundred fifty-six extracted teeth were decoronated, and the coronal portion of the canal was flared. The distance between the major foramen and the file tips (DMFF) was determined by different EALs. The relationship between the DMFFs determined by the EAL and the morphological features of the root apex was analyzed by linear regression analysis. Results: The average DMFFs were 0.261mm, 0.376 mm, and 0.383 mm for the Root ZX (J. Morita, Kyoto, Japan), Raypex 5 (VDW, Munich, Germany), and Elements Apex Locator (SybronEndo, Anaheim, CA), respectively. The file tips determined by EALs were much closer to the major foramen in teeth with a "lateral major foramen" (p < 0.001). The area and diameters of the minor foramen were significantly related to the variation of the DMFFs determined by EALs. Conclusion: When the "minor foramen" reading was given, the file tip connected to the Root ZX was much closer to the major foramen than the other two EALs. The minor foramen's morphology and the major foramen's location were both important influencing factors on the performance of EALs. (J Endod 2010;36:1399-1403)
引用
收藏
页码:1399 / 1403
页数:5
相关论文
共 26 条
[1]   Influence of pulp vitality on length determination by using the elements diagnostic unit and apex locator [J].
Akisue, Eduardo ;
Gavini, Giulio ;
Poli de Figueiredo, Jose Antonio .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2007, 104 (04) :E129-E132
[2]   Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX d-30 [J].
Bernardes, Ricardo Affonso ;
Duarte, Marco A. H. ;
Vasconcelos, Bruno C. ;
Moraes, Ivaldo G. ;
Bernardineli, Norberti ;
Garcia, Roberto B. ;
Baldi, Jarcio V. ;
Victorino, Fausto R. ;
Bramante, Clovis M. .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2007, 104 (04) :E91-E94
[3]   Influence of instrument size on the accuracy of different apex locators:: An in vitro study [J].
Briseno-Marroquin, Benjamin ;
Frajlich, Santiago ;
Goldberg, Fernando ;
Willershausen, Brita .
JOURNAL OF ENDODONTICS, 2008, 34 (06) :698-702
[4]   Morphometric study of the apical anatomy of C-shaped root canal systems in mandibular second molars [J].
Cheung, G. S. P. ;
Yang, J. ;
Fan, B. .
INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (04) :239-246
[5]   The effects of file size, sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals:: an in vitro study [J].
Ebrahim, A. K. ;
Yoshioka, T. ;
Kobayashi, C. ;
Suda, H. .
AUSTRALIAN DENTAL JOURNAL, 2006, 51 (02) :153-157
[6]   A simple mounting model for consistent determination of the accuracy and repeatability of apex locators [J].
ElAyouti, A ;
Löst, C .
INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (02) :108-112
[7]   Evaluation of the accuracy of three electronic apex locators using glass tubules [J].
Fan, W ;
Fan, B ;
Gutmann, JL ;
Bian, Z ;
Fan, MW .
INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (02) :127-135
[8]   ACCURACY OF THE ENDEX WITH VARIATIONS IN CANAL IRRIGANTS AND FORAMEN SIZE [J].
FOUAD, AF ;
RIVERA, EM ;
KRELL, KV .
JOURNAL OF ENDODONTICS, 1993, 19 (02) :63-67
[9]   Electronic apex locators [J].
Gordon, MPJ ;
Chandler, NP .
INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (07) :425-437
[10]   Influence of apical constriction diameter on root ZX apex locator precision [J].
Herrera, Manuela ;
Abalos, Camilo ;
Planas, Amparo Jimenez ;
Llamas, Rafael .
JOURNAL OF ENDODONTICS, 2007, 33 (08) :995-998