Determination of nebulizer droplet size distribution: A method based on impactor refrigeration

被引:47
作者
Berg, Elna [1 ]
Svensson, Jan Olof
Asking, Lars
机构
[1] AstraZeneca R&D, Analyt Dept, S-22187 Lund, Sweden
[2] LAB Pharma Ltd, Turku, Finland
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 2007年 / 20卷 / 02期
关键词
droplet size distribution; next generation pharmaceutical impactor (NGI); impaction; nebulizer;
D O I
10.1089/jam.2007.0556
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Size distributions of droplets generated by nebulizers are difficult to determine because of evaporation after aerosolization. We describe a method whereby a Next Generation Pharmaceutical Impactor (NGI; MSP Corporation, Shoreview, MN) is refrigerated at 5 degrees C before connecting it to the nebulizer in order to ensure an environment inside the NGI at close to 100% relative humidity (RH). This, in turn, reduces droplet evaporation between the nebulizer and impaction. The method development was performed with a Pari LC Plus jet nebulizer operated at 2.0 bar, with the NGI set at a flow rate of 15 L/min and, with salbutamol 5.0 mg/mL as the test solution. The droplet size distributions were expressed in terms of mass median aerodynamic diameter (MMAD) and geometric standard deviation (GSD). Variation in test conditions showed that the NGI should be cooled for at least 90 min, that nebulization should be started within 5 min after removal from the refrigerator, and that coating of collecting cups to prevent "'bouncing" is not necessary. Variation of ambient temperature and humidity had no relevant effect on results. MMAD and GSD results showed that refrigeration of the NGI resulted in droplet size distributions that are likely to reflect those originally delivered at the mouthpiece by the nebulizer. The method was shown to be robust, accurate with recovery of test solutions exceeding 99%, reproducible, and to be suitable for use with a wide range of commercially available nebulizers.
引用
收藏
页码:97 / 104
页数:8
相关论文
共 13 条
[1]  
Ferron G. A., 1977, Journal of Aerosol Science, V8, P251, DOI 10.1016/0021-8502(77)90045-3
[2]   Variations in predicted regional lung deposition of salbutamol sulphate between 19 nebulizer types [J].
Finlay, WH ;
Stapleton, KW ;
Zuberbuhler, P .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1998, 11 (02) :65-80
[3]   Undersizing of droplets from a vented nebulizer caused by aerosol heating during transit through an Anderson impactor [J].
Finlay, WH ;
Stapleton, KW .
JOURNAL OF AEROSOL SCIENCE, 1999, 30 (01) :105-109
[4]  
JAUERNIG J, 2002, RESP DRUG DELIVERY 8, V2, P537
[5]   Use of the next generation pharmaceutical impactor for particle size distribution measurements of live viral aerosol vaccines [J].
Leung, K ;
Louca, E ;
Gray, M ;
Tipples, G ;
Coates, AL .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2005, 18 (04) :414-426
[6]  
LEWIS RA, 1984, N ENGL REG ALLERGY P, V5, P23
[7]   Next generation pharmaceutical impactor: A new impactor for pharmaceutical inhaler testing. Part III. Extension of archival calibration to 15 L/min [J].
Marple, VA ;
Olson, BA ;
Santhanakrishnan, K ;
Roberts, DL ;
Mitchell, JP ;
Hudson-Curtis, BL .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2004, 17 (04) :335-343
[8]   OPERATING CHARACTERISTICS OF SOME COMPRESSED-AIR NEBULIZERS [J].
MERCER, TT ;
TILLERY, MI ;
CHOW, HY .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1968, 29 (01) :66-+
[9]   Cascade impactors for the size characterization of aerosols from medical inhalers: Their uses and limitations [J].
Mitchell, JP ;
Nagel, MW .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2003, 16 (04) :341-+
[10]   Validation of laser diffraction method as a substitute for cascade impaction in the European project for a nebulizer standard [J].
None, LV ;
Grimbert, D ;
Becquemin, MH ;
Boissinot, E ;
Le Pape, A ;
Lemarié, E ;
Diot, P .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2001, 14 (01) :107-114