Influence of in vivo hemoglobin carbamylation on HbA1c measurements by various methods

被引:44
作者
Chachou, A
Randoux, C
Millart, H
Chanard, J
Gillery, P [1 ]
机构
[1] CHU Reims, Hop Robert Debre, Cent Biochim Lab, F-51092 Reims, France
[2] CHU Reims, Maison Blanche Hosp, Dept Nephrol, Reims, France
[3] CHE Reims, Lab Pharmacol & Toxicol, Maison Blanche Hosp, Reims, France
关键词
HbA(1c); carbamylated hemoglobin; renal failure; non-enzymatic glycation; high-performance liquid chromatography; immunoassay;
D O I
10.1515/CCLM.2000.046
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Increased carbamylated hemoglobin formed in erythrocytes during uremia may interfere with HbA(1c) assays, but few studies compared directly both parameters. We measured carbamylated hemoglobin by HPLC in 45 non-diabetic uremic patients (16 with acute and two with chronic renal failure, 27 with transplant recipients) as 57.8 +/- 22.3 mu g carbamylvaline /g Hb (mean +/- standard deviation) vs. 31.6 +/- 5.1 in 15 controls (+83%, p < 0.001). In these samples, HbA(1c) was evaluated by three ion-exchange HPLC methods, 1: Diamat (BioRad), 2: A1c2.2 (Tosoh) and 3: HA8140 (Menarini), and one immunoassay method (Tinaquant II Roche). Whichever the method, mean HbA(1c) values obtained increased in patients with high (> 60 mu g carbamylvaline/g Hb) vs. low (< 45) carbamylated hemoglobin values (+0.08 to 0.25% of total Hb), but differences were not significant. Minor peaks on the chromatograms were however increased in parallel to carbamylated hemoglobin. HbA(1c) values over 6% were found in 4, 1, 2 and 0 samples, with HPLC 1,2, 3 and immunoassay, respectively. Fructosamine values were not significantly altered. Our results show that Hb adducts, whether due to carbamylation or to other chemical reactions, interfere to a variable extent with different HbA(1c) assay methods, and confirm that HbA(1c) values should be interpreted with caution in uremic patients.
引用
收藏
页码:321 / 326
页数:6
相关论文
共 31 条
  • [1] Carbamylated hemoglobin and carbamylated plasma protein in hemodialyzed patients
    Balion, CM
    Draisey, TF
    Thibert, RJ
    [J]. KIDNEY INTERNATIONAL, 1998, 53 (02) : 488 - 495
  • [2] Heterogeneity of hemoglobin A(1d): Assessment and partial characterization of two new minor hemoglobins, A(1d3a) and A(1d3b), increased in uremic and diabetic patients, respectively
    Bisse, E
    HuamanGuillen, P
    Horth, P
    BusseGrawitz, A
    Lizama, M
    KramerGuth, A
    Haehnel, W
    Wieland, H
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1996, 687 (02): : 349 - 356
  • [3] Uremia and HbA1c measured by high-performance liquid chromotography
    Chevenne, D
    Fonfrède, M
    Ducrocq, R
    Chauffert, M
    Trivin, F
    [J]. DIABETES CARE, 1998, 21 (03) : 463 - 464
  • [4] DIFFERENTIATION OF ACUTE FROM CHRONIC RENAL IMPAIRMENT BY DETECTION OF CARBAMYLATED HEMOGLOBIN
    DAVENPORT, A
    JONES, SR
    GOEL, S
    ASTLEY, JP
    HARTOG, M
    [J]. LANCET, 1993, 341 (8861) : 1614 - 1617
  • [5] Eckfeldt JH, 1997, CLIN CHEM, V43, P1811
  • [6] ENGBAEK F, 1989, CLIN CHEM, V35, P93
  • [7] Chromatographic and electrophoretic methods for modified hemoglobins
    Frantzen, F
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1997, 699 (1-2): : 269 - 286
  • [8] CARBAMYLATED HEMOGLOBIN IN PRERENAL AZOTEMIA
    FRAZAO, JM
    BARTH, RH
    BERLYNE, GM
    [J]. NEPHRON, 1995, 71 (02): : 153 - 155
  • [9] GILLERY P, 1995, CLIN CHEM, V41, P1644
  • [10] Evaluation of GHb assays in France by national quality control surveys
    Gillery, P
    Dumont, G
    Vassault, A
    [J]. DIABETES CARE, 1998, 21 (02) : 265 - 270