Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software

被引:36
作者
Badawi, Mohamed S. [1 ,2 ]
Jovanovic, Slobodan I. [3 ,4 ]
Thabet, Abouzeid A. [5 ]
El-Khatib, Ahmed M. [1 ]
Dlabac, Aleksandar D. [3 ]
Salem, Bohaysa A. [6 ]
Gouda, Mona M. [1 ]
Mihaljevic, Nikola N. [3 ,7 ]
Almugren, Kholud S. [8 ]
Abbas, Mahmoud I. [1 ]
机构
[1] Univ Alexandria, Dept Phys, Fac Sci, Alexandria 21511, Egypt
[2] Beirut Arab Univ, Dept Phys, Fac Sci, Beirut, Lebanon
[3] Univ Montenegro, Ctr Nucl Competence & Knowledge Management, Podgorica, Montenegro
[4] Univ Montenegro, Fac Math & Nat Sci, Dept Phys, Podgorica, Montenegro
[5] Pharos Univ Alexandria, Fac Allied Med Sci, Dept Med Equipment Technol, Alexandria, Egypt
[6] Pharos Univ Alexandria, Fac Phys Therapy, Dept Basic Sci, Alexandria, Egypt
[7] Univ Montenegro, Maritime Fac, Dept Math, Kotor, Montenegro
[8] Princess Nourah Bint Abdulrahaman Univ, Dept Phys, Fac Sci, Riyadh 1154455532, Saudi Arabia
关键词
GAMMA-RAY DETECTORS; SCINTILLATION DETECTORS; EFFICIENCY CALIBRATION; POINT; SPECTROMETRY; GEOMETRIES; NAI; GE;
D O I
10.1063/1.4978214
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The 4 pi NaI(Tl) gamma-ray detectors are consisted of the well cavity with cylindrical cross section, and the enclosing geometry of measurements with large detection angle. This leads to exceptionally high efficiency level and a significant coincidence summing effect, much more than a single cylindrical or coaxial detector especially in very low activity measurements. In the present work, the detection effective solid angle in addition to both full-energy peak and total efficiencies of well-type detectors, were mainly calculated by the new numerical simulation method (NSM) and ANGLE4 software. To obtain the coincidence summing correction factors through the previously men-tioned methods, the simulation of the coincident emission of photons was modeled mathematically, based on the analytical equations and complex integrations over the radioactive volumetric sources including the self-attenuation factor. The measured full-energy peak efficiencies and correction factors were done by using Eu-152, where an exact adjustment is required for the detector efficiency curve, because neglecting the coincidence summing effect can make the results inconsistent with the whole. These phenomena, in general due to the efficiency calibration process and the coincidence summing corrections, appear jointly. The full-energy peak and the total efficiencies from the two methods typically agree with discrepancy 10%. The discrepancy between the simulation, ANGLE4 and measured full-energy peak after corrections for the coincidence summing effect was on the average, while not exceeding 14%. There-fore, this technique can be easily applied in establishing the efficiency calibration curves of well-type detectors. (C) 2017 Author(s).
引用
收藏
页数:19
相关论文
共 34 条
[11]   Fast procedures for coincidence-summing correction in γ-ray spectrometry [J].
De Felice, P ;
Angelini, P ;
Fazio, A ;
Biagini, R .
APPLIED RADIATION AND ISOTOPES, 2000, 52 (03) :745-752
[12]   COINCIDENCE SUMMING CORRECTIONS IN GE(LI)-SPECTROMETRY AT LOW SOURCE-TO-DETECTOR DISTANCES [J].
DEBERTIN, K ;
SCHOTZIG, U .
NUCLEAR INSTRUMENTS & METHODS, 1979, 158 (2-3) :471-477
[13]  
Debertin K., 1988, Gamma- and X-Ray spectrometry with semiconductor detectors
[14]   Well-type NaI(Tl) detector efficiency using analytical technique and ANGLE 4 software based on radioactive point sources located out the well cavity [J].
El-Khatib, Ahmed M. ;
Badawi, Mohamed S. ;
Thabet, Abouzeid A. ;
Jovanovic, Slobodan I. ;
Gouda, Mona M. ;
Mohamed, Moustafa M. ;
Dlabac, Aleksandar D. ;
Abd-Elzaher, Mohamed ;
Mihaljevic, Nikola N. ;
Abbas, Mahmoud I. .
CHINESE JOURNAL OF PHYSICS, 2016, 54 (03) :338-346
[15]   WELL GE AND SEMI-PLANAR GE (HP) DETECTORS FOR LOW-LEVEL GAMMA-SPECTROMETRY [J].
ELDAOUSHY, F ;
GARCIATENORIO, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 356 (2-3) :376-384
[16]   Standardization of 67Ga by 4πγ (NaI) and 4πβ-γ coincidence methods [J].
García-Toraño, E ;
Roteta, M .
APPLIED RADIATION AND ISOTOPES, 2004, 60 (2-4) :353-356
[17]  
Gilmore G., 1995, Practical gamma-ray spectrometry
[18]   Calculation of Coincidence Summing Correction Factors for an HPGe detector using GEANT4 [J].
Giubrone, G. ;
Ortiz, J. ;
Gallardo, S. ;
Martorell, S. ;
Bas, M. C. .
JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2016, 158 :114-118
[19]   Calibration of well-type NaI(Tl) detector using a point sources measured out the detector well at different axial distances [J].
Gouda, M. M. ;
Badawi, M. S. ;
El-Khatib, A. M. ;
Mohamed, M. M. ;
Thabet, A. A. ;
Abbas, M. I. .
JOURNAL OF INSTRUMENTATION, 2015, 10
[20]   Accounting for incomplete charge collection in Monte Carlo simulations of the efficiency of well-type Ge-detectors [J].
Hernández, F ;
El-Daoushy, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 498 (1-3) :340-351