Radiation exposure with use of the mini-c-arm for routine orthopaedic imaging procedures

被引:61
作者
Badman, BL
Rill, L
Butkovich, B
Arreola, M
Griend, RAV
机构
[1] Univ Florida, Shands Teaching Hosp, Dept Orthoped, Gainesville, FL 32608 USA
[2] Univ Florida, Shands Teaching Hosp, Dept Radiol, Gainesville, FL 32608 USA
关键词
D O I
10.2106/JBJS.D.02162
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The use of mobile fluoroscopic devices during orthopaedic procedures is associated with substantial concern with regard to the radiation exposure to surgeons and support staff. The perceived increased risks associated with large c-arm devices have been well documented. However, no study to date has documented the relative radiation risk associated with the use of a mini-c-arm device. The purpose of the current study was to determine the amount of radiation received by the surgeon during the use of a mini-c-arm device and to compare this amount with documented measurements associated with the large c-arm device. Methods: With use of a radiation dosimeter, measurements were carried out with tissue-equivalent anthropomorphic phantoms to quantitatively determine exposure rates at various locations and distances from the mini-c-arm for two common upper and lower extremity procedures. Results: Regardless of position, distance, or relative duration of exposure, exposure rates resulting from the use of the mini-c-arm device were one to two orders of magnitude lower than those reported in the literature in association with the use of the large c-arm device. Conclusions: The mini-c-arm device should be utilized whenever feasible in order to eliminate many of the concerns associated with use of the large c-arm device, specifically those related to cumulative radiation hazards, positioning considerations, relative distance from the beam, and the need for protective shielding.
引用
收藏
页码:13 / 17
页数:5
相关论文
共 17 条
[1]  
[Anonymous], 1983, ITS TOO LATE SCI CAS
[2]  
BARRY TP, 1984, CLIN ORTHOP RELAT R, P160
[3]  
BUSHBERG JT, 2001, ESSENTIAL PHYS MED I, P52
[4]  
Bushong S. C., 1977, Physics Teacher, V15, P135, DOI 10.1119/1.2339571
[5]  
Cohen BL, 1971, CONCEPTS NUCL PHYS
[6]  
Dosch J C, 1983, Hefte Unfallheilkd, V161, P36
[7]  
GLACINO AA, 1980, J BONE JOINT SURG BR, V62, P227
[8]   RADIATION EXPOSURE TO THE SURGEON DURING CLOSED INTERLOCKING INTRAMEDULLARY NAILING [J].
LEVIN, PE ;
SCHOEN, RW ;
BROWNER, BD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (05) :761-766
[9]   Radiation exposure to the orthopaedic surgical team during fluoroscopy: "How far away is far enough?" [J].
Mehlman, CT ;
DiPasquale, TG .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (06) :392-398
[10]  
MILLER ME, 1983, J BONE JOINT SURG AM, V65, P1