Amorphous silicon, semiconductor X-ray converter detectors for protein crystallography

被引:15
作者
Ross, S [1 ]
Zentai, G [1 ]
Shah, KS [1 ]
Alkire, RW [1 ]
Naday, I [1 ]
Westbrook, EM [1 ]
机构
[1] RADIAT MONITORING DEVICES INC,WATERTOWN,MA 02172
关键词
X-ray detector; amorphous silicon; protein crystallography; lead iodide;
D O I
10.1016/S0168-9002(97)00793-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Hydrogenated amorphous silicon (a-Si:H) is a semiconductor material which call be inexpensively deposited to create a large array detector or readout structure. Combining it with a suitable semiconductor X-ray sensitive converter would produce a large, sensitive detector for X-rays with energies of 6-20 keV such as used in measurements of diffraction patterns for protein crystallography. In these experiments, the diffraction patterns created when X-rays strike crystallized protein samples are used to infer the physical structure of the molecules. The requirements for the detector include the ability to record signal peaks to high diffraction order to obtain accurate mapping of the electron density of the protein molecule, plus rapid sampling of the diffraction pattern to minimize radiation dose to the exposed crystal, while maintaining high signal-to-noise ratios. In this paper we summarize our results to date measuring and modeling the suitability of various semiconductor thin films for direct X-ray detection, and of the noise and readout properties of an amorphous silicon matrix array, We provide sample diffraction patterns obtained from a protein crystal taken at the Argonne Advanced Photon Source X-ray synchrotron using a phosphor coated a-Si: H array.
引用
收藏
页码:38 / 50
页数:13
相关论文
共 50 条
  • [31] Development and evaluation of a portable amorphous silicon flat panel X-ray detector
    Watanabe, N
    Mochizuki, C
    Kameshima, T
    Yamazaki, T
    Court, L
    Hayashida, S
    Morishita, M
    Ohta, S
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 103 - 114
  • [32] AMORPHOUS X-RAY DIFFRACTION AT HIGH PRESSURE: POLYAMORPHIC SILICON AND AMYLOID FIBRILS
    McMillan, Paul F.
    Daisenberger, Dominik
    Cabrera, Raul Quesada
    Meersman, Filip
    HIGH-PRESSURE CRYSTALLOGRAPHY: FROM FUNDAMENTAL PHENOMENA TO TECHNOLOGICAL APPLICATIONS, 2010, : 469 - +
  • [33] The influence of X-ray irradiation on structural relaxation and crystallization of amorphous silicon films
    Muto, S
    Kobayashi, Y
    Yu, KM
    Walukiewicz, W
    Echer, CJ
    McCormick, S
    Abelson, JR
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 1998, 37 (11): : 5890 - 5893
  • [34] Performance characteristics of an amorphous silicon flat panel X-ray imaging detector
    Granfors, PR
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 480 - 490
  • [35] Contaminant inclusion into protein crystals analyzed by electrospray mass spectrometry and X-ray crystallography
    Hirschler, J
    Halgand, F
    Forest, E
    Fontecilla-Camps, JC
    PROTEIN SCIENCE, 1998, 7 (01) : 185 - 192
  • [36] Silicon drift detectors for high count rate X-ray spectroscopy at room temperature
    Lechner, P
    Fiorini, C
    Hartmann, R
    Kemmer, J
    Krause, N
    Leutenegger, P
    Longoni, A
    Soltau, H
    Stötter, D
    Stötter, R
    Strüder, L
    Weber, U
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 458 (1-2) : 281 - 287
  • [37] STJ X-ray detectors with titanium sublayer
    Kozin, MG
    Romashkina, IL
    Sergeev, SA
    Nefedov, LV
    Andrianov, VA
    Naumkin, VN
    Koshelets, VP
    Filippenko, LV
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 520 (1-3) : 250 - 253
  • [38] Perovskite-Based X-ray Detectors
    Lin, Chen-Fu
    Huang, Kuo-Wei
    Chen, Yen-Ting
    Hsueh, Sung-Lin
    Li, Ming-Hsien
    Chen, Peter
    NANOMATERIALS, 2023, 13 (13)
  • [39] Recent developments in X-ray imaging detectors
    Moy, JP
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 442 (1-3) : 26 - 37
  • [40] X-ray fluorescence imaging with pixel detectors
    Tichy, V.
    Holy, T.
    Jakubek, J.
    Linhart, V.
    Pospisil, S.
    Vykydal, Z.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 591 (01) : 67 - 70