A Sensitive Photometric Procedure for Cobalt Determination in Water Employing a Compact Multicommuted Flow Analysis System

被引:4
|
作者
Magalhaes, Ticiane da Silva [1 ]
Reis, Boaventura F. [2 ]
机构
[1] Univ Sao Paulo, Inst Quim Sao Carlos, Sao Carlos, SP, Brazil
[2] Univ Sao Paulo, Ctr Energia Nucl Agr, Ave Centenario 303, BR-13416000 Piracicaba, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
Cobalt in water; light-emitting-diode-based photometry; LED; multicommuted flow analysis; green chemistry; portable setup; CLOUD POINT EXTRACTION; SPECTROPHOTOMETRIC DETERMINATION; ONLINE PRECONCENTRATION; CATALYTIC DETERMINATION; INJECTION ANALYSIS; SAMPLES; CHLORIDE; CADMIUM; RESIN; CELL;
D O I
10.1177/0003702817713749
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this work, a multicommuted flow analysis procedure is proposed for the spectrophotometric determination of cobalt in fresh water, employing an instrument setup of downsized dimension and improved cost-effectiveness. The method is based on the catalytic effect of Co(II) on the Tiron oxidation by hydrogen peroxide in alkaline medium, forming a complex that absorbs radiation at 425nm. The photometric detection was accomplished using a homemade light-emitting-diode (LED)-based photometer designed to use a flow cell with an optical path-length of 100mm to improve sensitivity. After selecting adequate values for the flow system variables, adherence to the Beer-Lambert-Bouguer law was observed for standard solution concentrations in the range of 0.13-1.5 mu gL(-1) Co(II). Other useful features including a relative standard deviation of 2.0% (n=11) for a sample with 0.49 mu gL(-1) Co(II), a detection limit of 0.06 mu gL(-1) Co(II) (n=20), an analytical frequency of 42 sample determinations per hour, and waste generation of 1.5mL per determination were achieved.
引用
收藏
页码:2154 / 2163
页数:10
相关论文
共 50 条