Apical extent of rotary canal instrumentation with an apex-locating handpiece in vitro

被引:25
作者
Campbell, D
Friedman, S
Nguyen, HQ
Kaufman, A
Keila, S
机构
[1] Univ Toronto, Fac Dent, Dept Endodont, Toronto, ON M5G 1G6, Canada
[2] Tel Aviv Univ, Maurice & Gabriela Goldschleger Sch Dent Med, Sect Endodontol, IL-69978 Tel Aviv, Israel
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS | 1998年 / 85卷 / 03期
关键词
D O I
10.1016/S1079-2104(98)90016-6
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Problem. The Tri Auto ZX (J. Morita Co., Kyoto, Japan) is a cordless endodontic handpiece with a built-in apex locator that is programmed to reverse the direction of rotation when the file reaches a predetermined apical level or when torque becomes excessive. Objective. nle purpose of this investigation was to examine the apical extent of rotary canal instrumentation and the ability to maintain apical constriction with the Tri Auto ZX at different automated settings. Study design. In 60 extracted teeth, canals were measured to the apical constriction, first visually and then electronically with the Tri Auto ZX; then they were instrumented with nickel titanium rotary files. For the instrumentation, the automatic apical reverse mechanism of the handpiece was set to 1, 1.5, or 2; these settings correspond to different distances from the apical foramen. Instrumentation was carried out apically until rotation was reversed by the automatic apical reverse function; the instrumented length was then measured, and the canal was filled with gutta-percha and sealer. The integrity of the apical constriction was assessed by exposing the apical 4 mm of the canal and observing the dentin-cementum junction. Paired t-tests were used to compare the visually measured length, the electronically measured length, and the instrumented length for each tooth. Results. On average, the electronically measured length was 0.54 mm shorter than the visually measured length (p < 0.05). When the automatic apical reverse mechanism's setting was 1, the instrumented length was 0.1 mm shorter than the electronically measured length; when the setting was 1.5, the instrumented length was 0.36 mm shorter than the electronically measured length (p < 0.01). Results were inconsistent when the setting was 2. Conclusion. Instrumentation with the automatic apical reverse feature set at 1 consistently approximated the apical constriction; however, the constriction was frequently enlarged.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 14 条
[1]   THE POSITION AND TOPOGRAPHY OF THE APICAL CANAL CONSTRICTION AND APICAL FORAMEN [J].
DUMMER, PMH ;
MCGINN, JH ;
REES, DG .
INTERNATIONAL ENDODONTIC JOURNAL, 1984, 17 (04) :192-198
[2]  
INGLE JI, 1994, ENDODONTICS, P92
[3]  
Katz A, 1996, J DENT CHILD, V63, P414
[4]   A new engine-driven canal preparation system with electronic canal measuring capability [J].
Kobayashi, C ;
Yoshioka, T ;
Suda, H .
JOURNAL OF ENDODONTICS, 1997, 23 (12) :751-754
[5]   NEW ELECTRONIC CANAL MEASURING DEVICE BASED ON THE RATIO METHOD [J].
KOBAYASHI, C ;
SUDA, H .
JOURNAL OF ENDODONTICS, 1994, 20 (03) :111-114
[6]  
KUTTLER Y, 1955, J Am Dent Assoc, V50, P544
[7]   AN EVALUATION OF THE CROWN-DOWN PRESSURELESS TECHNIQUE [J].
MORGAN, LF ;
MONTGOMERY, S .
JOURNAL OF ENDODONTICS, 1984, 10 (10) :491-498
[8]   Electronic length measurement using small and large files in enlarged canals [J].
Nguyen, HQ ;
Kaufman, AY ;
Komorowski, RC ;
Friedman, S .
INTERNATIONAL ENDODONTIC JOURNAL, 1996, 29 (06) :359-364
[9]   Comparison of radiographic and electronic working lengths [J].
Pratten, DH ;
McDonald, NJ .
JOURNAL OF ENDODONTICS, 1996, 22 (04) :173-176
[10]  
Saito T, 1990, Dent Jpn (Tokyo), V27, P65