In the paper GUM-Dw3 graphite-walled cylindrical cavity ionization chamber (of design and volume typical for the chambers for radiation therapy dosimetry) was presented as a realization of an ionometric standard for absorbed dose to water. A state-of-the-art uncertainty budget and results of comparison to GUMs standards has been presented. Good agreement of absorbed dose to water with reference standard, with relative deviation of 0.28%, and standard uncertainty (k=2) of 0.48% indicates good quality of the GUM-Dw3 as a primary standard. This paper describes the step-by-step procedure to calculate a perturbation factor, that is a crucial issue for ionometric measurement of absorbed dose to water. Capabilities of FLUKA code for this factor determination have been also investigated.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] A standard for absorbed dose rate to water in a 60Co field using a graphite calorimeter at the National Metrology Institute of Japan
    Morishita, Y.
    Kato, M.
    Takata, N.
    Kurosawa, T.
    Tanaka, T.
    Saito, N.
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (03) : 331 - 339
  • [22] CALCULATION OF DOSE CONVERSION FACTORS FOR CO-60 GAMMA-RADIATION IN WATER
    EISENLOH.HH
    ABEDINZA.R
    STRAHLENTHERAPIE, 1972, 143 (04) : 410 - &
  • [23] Ionization chamber dosimetry based on 60Co absorbed dose to water calibration for diagnostic kilovoltage x-ray beams
    Araki, Fujio
    Ohno, Takeshi
    Umeno, Suzuna
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (18):
  • [24] ABSORBED DOSE DETERMINATION WITH A WATER CALORIMETER IN COMPARISON WITH AN IONIZATION-CHAMBER
    KUBO, H
    PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (12): : 1391 - 1399
  • [25] The PTB water calorimeter for the absolute determination of absorbed dose to water in 60Co radiation
    Krauss, A
    METROLOGIA, 2006, 43 (03) : 259 - 272
  • [26] Influence of total absorbed dose of Co-60 γ-radiation on the properties of h-MoO3 thin films
    Sen, Sapan Kumar
    Manir, M. S.
    Dutta, Supria
    Ali, M. H.
    Khan, M. N. I.
    Matin, M. A.
    Hakim, M. A.
    THIN SOLID FILMS, 2020, 693
  • [27] CERENKOV RADIATION FROM A CO-60 SOURCE IN WATER
    RICH, JA
    SLOVACEK, RE
    STUDER, FJ
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1953, 43 (09) : 750 - 752
  • [28] CALORIMETRIC AND IONOMETRIC DOSIMETRY INTERCOMPARISONS IN CO-60 AND PROTON THERAPY BEAMS
    VERHEY, L
    LYMAN, J
    MCDONALD, JC
    MEDICAL PHYSICS, 1987, 14 (03) : 469 - 469
  • [29] DEPTH DOSE DATA FOR CO-60 GAMMA-RADIATION
    SHACKLETON, D
    LEROUX, PLM
    MEDCALF, G
    KORRUBEL, J
    HEALTH PHYSICS, 1980, 38 (01): : 100 - 101
  • [30] DEPTH DOSE IN SKIN FOR OBLIQUELY INCIDENT CO-60 RADIATION
    ORTON, CG
    SEIBERT, JB
    BRITISH JOURNAL OF RADIOLOGY, 1972, 45 (532): : 271 - +