Comparison of LaBr3:Ce and NAI(Tl) scintillators for radio-isotope identification devices

被引:77
作者
Milbrath, B. D. [1 ]
Choate, B. J. [1 ]
Fast, J. E. [1 ]
Hensley, W. K. [1 ]
Kouzes, R. T. [1 ]
Schweppe, J. E. [1 ]
机构
[1] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
lanthanum bromide; sodium iodide; scintillator; RIID; lanthanum-138; isotope identification;
D O I
10.1016/j.nima.2006.12.003
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Lanthanum bromide (LaBr3:Ce) scintillators offer significantly better resolution (< 3 percent at 662 keV) relative to sodium iodide (NaI(Tl)) but contain internal radioactivity that contributes to spectral counts. LaBr3:Ce has recently become available commercially in sizes large enough for the hand-held radio-isotope identification device (RIID) market. To study its potential for RIIDs, a series of measurements were performed comparing a 1.5 x 1.5-in. LaBr3:Ce detector with an Exploranium GR-135 RIID, which contains a 1.5 x 2.2-in. NaI(Tl) detector. Measurements were taken for short time frames and included examples of naturally occurring radioactive material (NORM), typically found in cargo, and special nuclear materials. To facilitate direct comparison, spectra from the different detectors were analyzed with the same isotope identification software (ORTEC ScintiVision (TM)). In general, the LaBr3:Ce detector was able to find more peaks and find them faster than the NaI(Tl) detector. To the same level of significance, the LaBr3:Ce detector was usually two to three times faster. The notable exception was for K-40-containing NORM where interfering internal activity due to La-138 in the LaBr3:Ce detector exists and NaI(Tl) consistently outperformed LaBr3:Ce. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:774 / 784
页数:11
相关论文
共 26 条
[1]  
*AM NAT STAND I, 2003, N42342003 ANSI
[2]   Comparison of LaCl3:Ce and NaI(Tl) scintillators in γ-ray spectrometry [J].
Balcerzyk, M ;
Moszynski, M ;
Kapusta, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 537 (1-2) :50-56
[3]  
BATES D, 2004, PNNL14757
[4]   Gamma ray spectroscopy with a ⊘19 x19 mm3 LaBr3 : 0.5% Ce3+ scintillator [J].
Dorenbos, P ;
de Haas, JTM ;
van Eijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (03) :1289-1296
[5]  
Firestone R.B., WWW TABLE RADIOACTIV
[6]   Observations on the background spectra of four LaCl3(Ce) scintillation detectors [J].
Hartwell, JK ;
Gehrke, RJ .
APPLIED RADIATION AND ISOTOPES, 2005, 63 (02) :223-228
[7]  
ILTIS A, 2004, IEEE NSS C ROM IT OC
[8]   Lanthanum halide scintillators: Properties and applications [J].
Iltis, Alain ;
Mayhugh, M. R. ;
Menge, P. ;
Rozsa, C. M. ;
Selles, O. ;
Solovyev, V. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 563 (02) :359-363
[9]  
KERNAN WJ, 2004, IEEE NSS C ROM IT OC
[10]  
KOUZES RT, 2004, PUBLIC PROTECTION NU