Automated ultrafast kilovoltage-megavoltage cone-beam CT for image guided radiotherapy of lung cancer: System description and real-time results

被引:3
作者
Blessing, Manuel [1 ]
Arns, Anna [1 ]
Wertz, Hansjoerg [1 ]
Stsepankou, Dzmitry [1 ]
Boda-Heggemann, Judit [1 ]
Hesser, Juergen [1 ]
Wenz, Frederik [1 ]
Lohr, Frank [1 ]
机构
[1] Heidelberg Univ, Univ Med Ctr Mannheim, Dept Radiat Oncol, Bldg 4,Level 0,Theodor Kutzer Ufer 1-3, D-68167 Mannheim, Germany
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2018年 / 28卷 / 02期
关键词
IGRT; CBCT; Kilovoltage-megavoltage imaging; Patient positioning; Data acquisition; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; TRACKING; MOTION; TUMOR; IRRADIATION; IMRT;
D O I
10.1016/j.zemedi.2018.01.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To establish a fully automated kV-MV CBCT imaging method on a clinical linear accelerator that allows image acquisition of thoracic targets for patient positioning within one breath-hold (similar to 15 s) under realistic clinical conditions. Methods and materials: Our previously developed FPGA-based hardware unit which allows synchronized kV-MV CBCT projection acquisition is connected to a clinical linear accelerator system via a multi-pin switch; i.e. either kV-MV imaging or conventional clinical mode can be selected. An application program was developed to control the relevant linac parameters automatically and to manage the MV detector readout as well as the gantry angle capture for each MV projection. The kV projections are acquired with the conventional CBCT system. GPU-accelerated filtered backprojection is performed separately for both data sets. After appropriate grayscale normalization both modalities are combined and the final kV-MV volume is re-imported in the CBCT system to enable image matching. To demonstrate adequate geometrical accuracy of the novel imaging system the Penta-Guide phantom QA procedure is performed. Furthermore, a human plastinate and different tumor shapes in a thorax phantom are scanned. Diameters of the known tumor shapes are measured in the kV-MV reconstruction. Results: An automated kV-MV CBCT workflow was successfully established in a clinical environment. The overall procedure, from starting the data acquisition until the reconstructed volume is available for registration, requires similar to 90 s including 17 s acquisition time for 100 degrees rotation. It is very simple and allows target positioning in the same way as for conventional CBCT. Registration accuracy of the QA phantom is within +/- 1 mm. The average deviation from the known tumor dimensions measured in the thorax phantom was 0.7 mm which corresponds to an improvement of 36% compared to our previous kV-MV imaging system. Conclusions: Due to automation the kV-MV CBCT workflow is speeded up by a factor of > 10 compared to the manual approach. Thus, the system allows a simple, fast and reliable imaging procedure and fulfills all requirements to be successfully introduced into the clinical workflow now, enabling single-breath-hold volume imaging.
引用
收藏
页码:110 / 120
页数:11
相关论文
共 40 条
[1]   Megavoltage cone beam computed tomography: Commissioning and evaluation of patient dose [J].
Abou-Elenein, Hassan S. ;
Attalla, Ehab M. ;
Ammar, H. ;
Eldesoky, Ismail ;
Farouk, Mohamed ;
Zaghloul, Mohamed S. .
JOURNAL OF MEDICAL PHYSICS, 2011, 36 (04) :205-212
[2]  
Arns A, 2017, PLOS ONE, V12, P1
[3]  
Arns A, 2016, STRAHLENTHER ONKOL, V192, P312, DOI 10.1007/s00066-016-0947-2
[4]   Feasibility of using single photon counting X-ray for lung tumor position estimation based on 4D-CT [J].
Aschenbrenner, Katharina P. ;
Guthier, Christian V. ;
Lyatskaya, Yulia ;
Boda-Heggemann, Judit ;
Wenz, Frederik ;
Hesser, Juergen W. .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (03) :243-254
[5]   Investigation of cone-beam CT image quality trade-off for image-guided radiation therapy [J].
Bian, Junguo ;
Sharp, Gregory C. ;
Park, Yang-Kyun ;
Ouyang, Jinsong ;
Bortfeld, Thomas ;
El Fakhri, Georges .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (09) :3317-3346
[6]  
Bibault J-E., 2012, Radiat Oncol, V7, P1
[7]   BREATH-HOLD TARGET LOCALIZATION WITH SIMULTANEOUS KILOVOLTAGE/MEGAVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY AND FAST RECONSTRUCTION [J].
Blessing, Manuel ;
Stsepankou, Dzmitry ;
Wertz, Hansjoerg ;
Arns, Anna ;
Lohr, Frank ;
Hesser, Juergen ;
Wenz, Frederik .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (04) :1219-1226
[8]   Deep Inspiration Breath Hold-Based Radiation Therapy: A Clinical Review [J].
Boda-Heggemann, Judit ;
Knopf, Antje-Christin ;
Simeonova-Chergou, Anna ;
Wertz, Hansjoerg ;
Stieler, Florian ;
Jahnke, Anika ;
Jahnke, Lennart ;
Fleckenstein, Jens ;
Vogel, Lena ;
Arns, Anna ;
Blessing, Manuel ;
Wenz, Frederik ;
Lohr, Frank .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 94 (03) :478-492
[9]   Clinical outcome of hypofractionated breath-hold image-guided SABR of primary lung tumors and lung metastases [J].
Boda-Heggemann, Judit ;
Frauenfeld, Anian ;
Weiss, Christel ;
Simeonova, Anna ;
Neumaier, Christian ;
Siebenlist, Kerstin ;
Attenberger, Ulrike ;
Heussel, Claus Peter ;
Schneider, Frank ;
Wenz, Frederik ;
Lohr, Frank .
RADIATION ONCOLOGY, 2014, 9
[10]   Multiple breath-hold CBCT for online image guided radiotherapy of lung tumors: Simulation with a dynamic phantom and first patient data [J].
Boda-Heggemann, Judit ;
Fleckenstein, Jens ;
Lohr, Frank ;
Wertz, Hansjoerg ;
Nachit, Mohammed ;
Blessing, Manuel ;
Stsepankou, Dzimitry ;
Loeb, Iris ;
Kuepper, Beate ;
Kavanagh, Anthony ;
Hansen, Vibeke N. ;
Brada, Michael ;
Wenz, Frederik ;
McNair, Helen .
RADIOTHERAPY AND ONCOLOGY, 2011, 98 (03) :309-316