Iterative reconstruction technique with reduced volume CT dose index: diagnostic accuracy in pediatric acute appendicitis

被引:21
|
作者
Didier, Ryne A. [1 ]
Vajtai, Petra L. [1 ,2 ]
Hopkins, Katharine L. [1 ,2 ]
机构
[1] Oregon Hlth & Sci Univ, Dept Diagnost Radiol, Portland, OR 97239 USA
[2] Oregon Hlth & Sci Univ, Dept Pediat, Portland, OR 97239 USA
关键词
Iterative reconstruction; Radiation dose reduction; Appendicitis; Computed tomography; Children; CHILDREN; SIZE;
D O I
10.1007/s00247-014-3109-7
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Iterative reconstruction technique has been proposed as a means of reducing patient radiation dose in pediatric CT. Yet, the effect of such reductions on diagnostic accuracy has not been thoroughly evaluated. This study compares accuracy of diagnosing pediatric acute appendicitis using contrast-enhanced abdominopelvic CT scans performed with traditional pediatric weight-based protocols and filtered back projection reconstruction vs. a filtered back projection/iterative reconstruction technique blend with reduced volume CT dose index (CTDIvol). Results of pediatric contrast-enhanced abdominopelvic CT scans done for pain and/or suspected appendicitis were reviewed in two groups: A, 192 scans performed with the hospital's established weight-based CT protocols and filtered back projection reconstruction; B, 194 scans performed with iterative reconstruction technique and reduced CTDIvol. Reduced CTDIvol was achieved primarily by reductions in effective tube current-time product (mAs(eff)) and tube peak kilovoltage (kVp). CT interpretation was correlated with clinical follow-up and/or surgical pathology. CTDIvol, size-specific dose estimates (SSDE) and performance characteristics of the two CT techniques were then compared. Between groups A and B, mean CTDIvol was reduced by 45%, and mean SSDE was reduced by 46%. Sensitivity, specificity and diagnostic accuracy were 96%, 97% and 96% in group A vs. 100%, 99% and 99% in group B. Accuracy in diagnosing pediatric acute appendicitis was maintained in contrast-enhanced abdominopelvic CT scans that incorporated iterative reconstruction technique, despite reductions in mean CTDIvol and SSDE by nearly half as compared to the hospital's traditional weight-based protocols.
引用
收藏
页码:181 / 187
页数:7
相关论文
共 50 条
  • [41] Diagnostic accuracy of focused appendiceal CT in clinically equivocal cases of acute appendicitis
    Wijetunga, R
    Tan, BS
    Rouse, JC
    Bigg-Wither, GW
    Doust, BD
    RADIOLOGY, 2001, 221 (03) : 747 - 753
  • [42] Accuracy and radiation dose reduction of a limited abdominopelvic CT in the diagnosis of acute appendicitis
    Michael T. Corwin
    Melanie Chang
    Ghaneh Fananapazir
    Anthony Seibert
    Ramit Lamba
    Abdominal Imaging, 2015, 40 : 1177 - 1182
  • [43] Accuracy and radiation dose reduction of a limited abdominopelvic CT in the diagnosis of acute appendicitis
    Corwin, Michael T.
    Chang, Melanie
    Fananapazir, Ghaneh
    Seibert, Anthony
    Lamba, Ramit
    ABDOMINAL IMAGING, 2015, 40 (05): : 1177 - 1182
  • [44] CT imaging of congenital lung lesions: effect of iterative reconstruction on diagnostic performance and radiation dose
    Jay E. Haggerty
    Ethan A. Smith
    Shaun M. Kunisaki
    Jonathan R. Dillman
    Pediatric Radiology, 2015, 45 : 989 - 997
  • [45] How far can the radiation dose be lowered in head CT with iterative reconstruction? Analysis of imaging quality and diagnostic accuracy
    Tung-Hsin Wu
    Sheng-Che Hung
    Jing-Yi Sun
    Chung-Jung Lin
    Chung-Hsien Lin
    Chen Fen Chiu
    Min-Jsuan Liu
    Michael Mu Huo Teng
    Wan-Yuo Guo
    Cheng-Yen Chang
    European Radiology, 2013, 23 : 2612 - 2621
  • [46] Characterization of adaptive statistical iterative reconstruction algorithm for dose reduction in CT: A pediatric oncology perspective
    Brady, S. L.
    Yee, B. S.
    Kaufman, R. A.
    MEDICAL PHYSICS, 2012, 39 (09) : 5520 - 5531
  • [47] Impact of iterative reconstruction on the diagnosis of acute pulmonary embolism (PE) on reduced-dose chest CT angiograms
    François Pontana
    Simon Henry
    Alain Duhamel
    Jean-Baptiste Faivre
    Nunzia Tacelli
    Julien Pagniez
    Jacques Remy
    Martine Remy-Jardin
    European Radiology, 2015, 25 : 1182 - 1189
  • [48] Submilisievert ultralow-dose CT colonography using iterative reconstruction technique: a feasibility study
    Lambert, Lukas
    Danes, Jan
    Jahoda, Jiri
    Masek, Martin
    Lisy, Jiri
    Ourednicek, Petr
    ACTA RADIOLOGICA, 2015, 56 (05) : 517 - 525
  • [49] Adaptive Statistical Iterative Reconstruction Technique for Pulmonary CT: Image Quality of the Cadaveric Lung on Standard- and Reduced-Dose CT
    Yanagawa, Masahiro
    Honda, Osamu
    Yoshida, Shigeyuki
    Kikuyama, Ayano
    Inoue, Atsuo
    Sumikawa, Hiromitsu
    Koyama, Mitsuhiro
    Tomiyama, Noriyuki
    ACADEMIC RADIOLOGY, 2010, 17 (10) : 1259 - 1266
  • [50] Optimizing CT technique to reduce radiation dose: effect of changes in kVp, iterative reconstruction, and noise index on dose and noise in a human cadaver
    Chang K.J.
    Collins S.
    Li B.
    Mayo-Smith W.W.
    Radiological Physics and Technology, 2017, 10 (2) : 180 - 188