A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement

被引:21
作者
Bitter, Thom [1 ]
Khan, Imran [2 ]
Marriott, Tim [2 ]
Lovelady, Elaine [2 ]
Verdonschot, Nico [1 ,3 ]
Janssen, Dennis [1 ]
机构
[1] Radboudumc, Orthopaed Res Lab, Nijmegen, Netherlands
[2] Zimmer Biomet, Swindon, Wilts, England
[3] Univ Twente, Enschede, Netherlands
关键词
Fretting; finite element modelling; total hip arthroplasty; taper junction; wear testing; wear modelling; LARGE-DIAMETER METAL; CORROSION; INTERFACE; ARTHROPLASTY; PREDICTION;
D O I
10.1177/0954411917713774
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fretting corrosion at the taper interface of modular hip implants has been implicated as a possible cause of implant failure. This study was set up to gain more insight in the taper mechanics that lead to fretting corrosion. The objectives of this study therefore were (1) to select experimental loading conditions to reproduce clinically relevant fretting corrosion features observed in retrieved components, (2) to develop a finite element model consistent with the fretting experiments and (3) to apply more complicated loading conditions of activities of daily living to the finite element model to study the taper mechanics. The experiments showed similar wear patterns on the taper surface as observed in retrievals. The finite element wear score based on Archard's law did not correlate well with the amount of material loss measured in the experiments. However, similar patterns were observed between the simulated micromotions and the experimental wear measurements. Although the finite element model could not be validated, the loading conditions based on activities of daily living demonstrate the importance of assembly load on the wear potential. These findings suggest that finite element models that do not incorporate geometry updates to account for wear loss may not be appropriate to predict wear volumes of taper connections.
引用
收藏
页码:862 / 870
页数:9
相关论文
共 28 条
[1]  
Bergmann G., 2002, HIP98 LOADING HIP JO
[2]   Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings [J].
Bishop, Nicholas ;
Witt, Florian ;
Pourzal, Robin ;
Fischer, Alfons ;
Ruetschi, Marcel ;
Michel, Markus ;
Morlock, Michael .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (07) :1116-1122
[3]   Experimental Measurement of the Static Coefficient of Friction at the Ti-Ti Taper Connection in Total Hip Arthroplasty [J].
Bitter, T. ;
Khan, I. ;
Marriott, T. ;
Schreurs, B. W. ;
Verdonschot, N. ;
Janssen, D. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2016, 138 (03)
[4]  
BOBYN JD, 1994, CLIN ORTHOP RELAT R, P27
[5]   Pseudotumour Formation Due to Tribocorrosion at the Taper Interface of Large Diameter Metal on Polymer Modular Total Hip Replacements [J].
Cook, Richard B. ;
Bolland, Benjamin J. R. F. ;
Wharton, Julian A. ;
Tilley, Simon ;
Latham, Jeremy M. ;
Wood, Robert J. K. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (08) :1430-1436
[6]   Metal-on-metal bearings total hip arthroplasty: The cobalt and chromium ions release concern [J].
Delaunay, C. ;
Petit, I. ;
Learmonth, I. D. ;
Oger, P. ;
Vendittoli, P. A. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2010, 96 (08) :894-904
[7]   Total hip arthroplasty head-neck contact mechanics: A stochastic investigation of key parameters [J].
Donaldson, Finn E. ;
Coburn, James C. ;
Siegel, Karen Lohmann .
JOURNAL OF BIOMECHANICS, 2014, 47 (07) :1634-1641
[8]   Environmental effect on fretting of metallic materials for orthopaedic implants [J].
Duisabeau, L ;
Combrade, P ;
Forest, B .
WEAR, 2004, 256 (7-8) :805-816
[9]   Finite element analysis of the head-neck taper interface of modular hip prostheses [J].
Dyrkacz, R. M. R. ;
Brandt, J. M. ;
Morrison, J. B. ;
O' Brien, S. T. ;
Ojo, O. A. ;
Turgeon, T. R. ;
Wyss, U. P. .
TRIBOLOGY INTERNATIONAL, 2015, 91 :206-213
[10]   Stability and Trunnion Wear Potential in Large-diameter Metal-on-Metal Total Hips: A Finite Element Analysis [J].
Elkins, Jacob M. ;
Callaghan, John J. ;
Brown, Thomas D. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (02) :529-542