Slit design for efficient and accurate MTF measurement at megavoltage x-ray energies

被引:15
|
作者
Sawant, Arnit [1 ]
Antonuk, Larry [1 ]
El-Mohri, Youcef [1 ]
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48103 USA
关键词
modulation transfer function; MTF measurement; slit method; megavoltage x-ray imaging; portal imaging; PANEL IMAGER AMFPI; TO-NOISE ANALYSIS; FLAT-PANEL; QUANTUM EFFICIENCY; PERFORMANCE; DETECTORS; RADIOTHERAPY; PROTOTYPE; SYSTEM;
D O I
10.1118/1.2717405
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Empirical determination of the modulation transfer function (MTF) for analog and digital megavoltage x-ray imagers is a challenging task. The most common method used to determine MTF at megavoltage x-ray energies employs a long, narrow slit formed by two parallel, metal blocks in order to form a "slit beam." In this work, a detailed overview of some of the important considerations of slit design is presented. Based on these considerations, a novel, compact slit, using 19 cm thick tungsten blocks, was designed. The prototype slit was configured to attach to the accessory slot of the gantry of a linear accelerator, which greatly simplified the measurement process. Measurements were performed to determine the presampling MTF at 6 MV for an indirect detection active matrix flat panel imager prototype previously developed for megavoltage imaging applications. In addition, the effects of two important slit design parameters, material type and thickness, on the accuracy of MTF determination were investigated via a Monte Carlo-based theoretical study. Empirically determined MTFs obtained from the prototype slit closely match those from an earlier, less compact slit design based on 40 cm thick steel blocks. The results of the Monte Carlo-based theoretical studies indicate that the prototype slit achieves close-to-ideal performance in terms of accurately determining the MTF by virtue of practically 100% beam attenuation in regions other than the slit gap. Furthermore, the theoretical results suggest that it may be possible to achieve even further reductions in slit thickness without compromising measurement accuracy. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1535 / 1545
页数:11
相关论文
共 50 条
  • [21] Raytracing with MARX - X-ray observatory design, calibration, and support
    Davis, John E.
    Bautz, Mark W.
    Dewey, Daniel
    Heilmann, Ralf K.
    Houck, John C.
    Huenemoerder, David P.
    Marshall, Herman L.
    Nowak, Michael A.
    Schattenburg, Mark L.
    Schulz, Norbert S.
    Smith, Randall
    SPACE TELESCOPES AND INSTRUMENTATION 2012: ULTRAVIOLET TO GAMMA RAY, 2012, 8443
  • [22] X-ray telescope design and technology: what the future holds
    Willingale, Richard
    SPACE TELESCOPES AND INSTRUMENTATION 2010: ULTRAVIOLET TO GAMMA RAY, 2010, 7732
  • [23] Bayesian optimization to design a novel x-ray shaping device
    Whelan, Brendan
    Trovati, Stefania
    Wang, Jinghui
    Fahrig, Rebecca
    Maxim, Peter G.
    Hanuka, Adi
    Shumail, Muhammad
    Tantawi, Sami
    Merrick, Julian
    Perl, Joseph
    Keall, Paul
    Loo, Billy W., Jr.
    MEDICAL PHYSICS, 2022, 49 (12) : 7623 - 7637
  • [24] Accurate contrast determination for X-ray speckle visibility spectroscopy
    Sun, Yanwen
    Montana-Lopez, Jordi
    Fuoss, Paul
    Sutton, Mark
    Zhu, Diling
    JOURNAL OF SYNCHROTRON RADIATION, 2020, 27 : 999 - 1007
  • [25] Research on Slit Method for Measuring Focal Spot Size of X-ray Tube
    Li W.
    Wang L.
    Zheng J.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2024, 58 (02): : 498 - 505
  • [26] Evaluation of 3D-printed bolus for radiotherapy using megavoltage X-ray beams
    Zhang, Chunsu
    Lewin, Will
    Cullen, Ashley
    Thommen, Daniel
    Hill, Robin
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (03) : 414 - 421
  • [27] Evaluation of 3D-printed bolus for radiotherapy using megavoltage X-ray beams
    Chunsu Zhang
    Will Lewin
    Ashley Cullen
    Daniel Thommen
    Robin Hill
    Radiological Physics and Technology, 2023, 16 : 414 - 421
  • [28] A novel x-ray detector design with higher DQE and reduced aliasing: Theoretical analysis of x-ray reabsoprtion in detector converter material
    Nano, Tomi
    Escartin, Terenz
    Karim, Karim S.
    Cunningham, Ian A.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [29] Accurate determination of small spot X-ray tube spectra by means of direct measurement in a calibrated instrumental setup
    Rackwitz, Vanessa
    Warrikhoff, Alexander
    Hodoroaba, Vasile-Dan
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (03) : 458 - 463
  • [30] Effect of X-ray energies on induced photo-neutron doses
    Khaled, N. E.
    Ghanim, E. H.
    Shinashin, Kh
    El-Sersy, A. R.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2014, 169 (03): : 239 - 248