Slit design for efficient and accurate MTF measurement at megavoltage x-ray energies

被引:15
|
作者
Sawant, Arnit [1 ]
Antonuk, Larry [1 ]
El-Mohri, Youcef [1 ]
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48103 USA
关键词
modulation transfer function; MTF measurement; slit method; megavoltage x-ray imaging; portal imaging; PANEL IMAGER AMFPI; TO-NOISE ANALYSIS; FLAT-PANEL; QUANTUM EFFICIENCY; PERFORMANCE; DETECTORS; RADIOTHERAPY; PROTOTYPE; SYSTEM;
D O I
10.1118/1.2717405
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Empirical determination of the modulation transfer function (MTF) for analog and digital megavoltage x-ray imagers is a challenging task. The most common method used to determine MTF at megavoltage x-ray energies employs a long, narrow slit formed by two parallel, metal blocks in order to form a "slit beam." In this work, a detailed overview of some of the important considerations of slit design is presented. Based on these considerations, a novel, compact slit, using 19 cm thick tungsten blocks, was designed. The prototype slit was configured to attach to the accessory slot of the gantry of a linear accelerator, which greatly simplified the measurement process. Measurements were performed to determine the presampling MTF at 6 MV for an indirect detection active matrix flat panel imager prototype previously developed for megavoltage imaging applications. In addition, the effects of two important slit design parameters, material type and thickness, on the accuracy of MTF determination were investigated via a Monte Carlo-based theoretical study. Empirically determined MTFs obtained from the prototype slit closely match those from an earlier, less compact slit design based on 40 cm thick steel blocks. The results of the Monte Carlo-based theoretical studies indicate that the prototype slit achieves close-to-ideal performance in terms of accurately determining the MTF by virtue of practically 100% beam attenuation in regions other than the slit gap. Furthermore, the theoretical results suggest that it may be possible to achieve even further reductions in slit thickness without compromising measurement accuracy. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1535 / 1545
页数:11
相关论文
共 50 条
  • [1] Technical Note: Accuracy of MTF measurements with an edge phantom at megavoltage x-ray energies
    Loot, Katharina
    Block, Andreas
    MEDICAL PHYSICS, 2019, 46 (12) : 5685 - 5689
  • [2] Comparison of Noise Power Spectrum Methodologies in Measurements by Using Megavoltage X-ray Energies
    Min, Jung-Whan
    Son, Jin-Hyun
    Jeong, Hoi-Woun
    Kim, Jung-Min
    Seoung, Youl-Hun
    Son, Soon-Yong
    Kim, Ho-Kyung
    Kim, Sang-Young
    Lee, Do-Wan
    Jung, Jae-Yong
    Suh, Tae-Suk
    Choe, Bo-Young
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 60 (01) : 129 - 136
  • [3] MTF and DQE enhancement using an apodized-aperture x-ray detector design
    Nano, Tomi F.
    Escartin, Terenz
    Ismailova, Elina
    Karim, Karim S.
    Lindstrom, Jan
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL PHYSICS, 2017, 44 (09) : 4525 - 4535
  • [4] Techniques to Improve the Accuracy of Presampling MTF Measurement in Digital X-ray Imaging Based on Constrained Spline Regression
    Zhou, Zhongxing
    Zhu, Qingzhen
    Zhao, Huijuan
    Zhang, Lixin
    Ma, Wenjuan
    Gao, Feng
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (04) : 1339 - 1349
  • [5] Rotating slit X-ray backscatter imaging
    Kacperski, Krzysztof
    Klimaszewski, Jan
    Kosinski, Tymoteusz
    Wronka, Slawomir
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2024, 56 (10) : 4247 - 4253
  • [6] An inherent anti-scatter detector for megavoltage x-ray imaging
    Teymurazyan, A.
    Pang, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05) : 1479 - 1493
  • [7] A simple system for measuring the transmittance curve of x-ray filters at low x-ray energies
    Cao, Jiewei
    Chen, Tianxiang
    He, Huilin
    Xu, Yupeng
    Cui, Xingzhu
    Gao, Na
    Xu, Xiongwei
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2022, 93 (12)
  • [8] Single-Pulse Measurement Electronics for Accurate Dosimetry in X-ray Radiation Therapy
    Pettinato, Sara
    Olivieri, Riccardo
    Salvatori, Stefano
    2021 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (IEEE MEMEA 2021), 2021,
  • [9] Enhanced slit methodology for characterization of x-ray focal properties
    Wang, Dazheng
    Chu, Mingzhang
    Li, Han
    X-RAY SPECTROMETRY, 2024,
  • [10] Difference in the relative response of the alanine dosimeter to megavoltage x-ray and electron beams
    Anton, M.
    Kapsch, R-P
    Krauss, A.
    von Voigts-Rhetz, P.
    Zink, K.
    McEwen, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (10) : 3259 - 3282