High-temperature high-pressure oscillating tube densimeter

被引:23
作者
Chang, RF
Moldover, MR
机构
[1] Thermophysics Division, Chem. Sci. and Technology Laboratory, Natl. Inst. of Std. and Technology, Gaithersburg
关键词
D O I
10.1063/1.1146580
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe an oscillating tube densimeter for use at temperatures up to at least 575 K and at pressures up to 20 MPa. After the densimeter was calibrated under vacuum and filled with water, it was used to measure the density of toluene from 298 to 575 K at 13.8 MPa. The results agree (0.1% rms deviation) with those obtained using other techniques. In the present densimeter, an alternating current is passed through the tube containing the sample to force the tube to oscillate in the field of a permanent magnet. This design avoids the use of electromagnets (with their attendant polymer or ceramic insulations) and does not require attachment of appendages to the oscillating tube. This densimeter oscillates in an overtone rather than in its fundamental mode, thereby achieving improved isolation from environmental noise and a shorter response time. The densimeter is small, weighing 370 g. If transformers are used to couple electrical signals to the oscillating tube, the densimeter itself may be constructed entirely out of metal. (C) 1996 American Institute of Physics.
引用
收藏
页码:251 / 256
页数:6
相关论文
共 13 条
[1]  
AKHUNDOV TS, 1970, IZV VYSSH UCHEBN ZAV, V13, P67
[2]   HIGH-PRECISION FLOW DENSIMETER FOR FLUIDS AT TEMPERATURES TO 700-K AND PRESSURES TO 40-MPA [J].
ALBERT, HJ ;
WOOD, RH .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (04) :589-593
[3]  
[Anonymous], 1984, NBS/NRC Steam Tables: Thermodynamic and Transport Properties and Computer Programs for Vapor and Liquid States of Water in SI Units
[4]   COMPRESSED LIQUID DENSITIES AND SATURATION DENSITIES OF CHLORODIFLUOROMETHANE (R22) [J].
DEFIBAUGH, DR ;
MORRISON, G .
JOURNAL OF CHEMICAL AND ENGINEERING DATA, 1992, 37 (01) :107-110
[6]  
Kashiwagi H., 1982, INTR J THERMOPHYSICS, V3, P201
[7]  
KRAGAS TK, UNPUB
[8]  
KRATKY O, 1969, Z ANGEW PHYSIK, V27, P273
[9]   A NEW VERSION OF VIBRATING-TUBE FLOW DENSITOMETER FOR MEASUREMENTS AT TEMPERATURES UP TO 730-K [J].
MAJER, V ;
CROVETTO, R ;
WOOD, RH .
JOURNAL OF CHEMICAL THERMODYNAMICS, 1991, 23 (04) :333-344
[10]  
Morse PM., 1948, Vibration and Sound