Optical distortion correction in Optical Coherence Tomography for quantitative ocular anterior segment by three-dimensional imaging

被引:141
作者
Ortiz, Sergio [1 ,2 ]
Siedlecki, Damian [1 ,3 ]
Grulkowski, Ireneusz [4 ]
Remon, Laura [1 ]
Pascual, Daniel [1 ]
Wojtkowski, Maciej [4 ]
Marcos, Susana [1 ]
机构
[1] CSIC, Inst Opt Daza de Valdes, Madrid 28006, Spain
[2] CIDA, Lab Ensayos, Madrid 28033, Spain
[3] Wroclaw Univ Technol, Inst Phys, PL-50370 Wroclaw, Poland
[4] Nicolaus Copernicus Univ, Inst Phys, PL-87100 Torun, Poland
关键词
POSTERIOR CORNEAL SURFACE; INTRAOCULAR LENSES; CRYSTALLINE LENS; IN-VIVO; REFRACTIVE-INDEX; SWEPT SOURCE; FIBER LASER; PURKINJE; NOISE; DECENTRATION;
D O I
10.1364/OE.18.002782
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A method for three-dimensional 3-D optical distortion (refraction) correction on anterior segment Optical Coherence Tomography (OCT) images has been developed. The method consists of 3-D ray tracing through the different surfaces, following denoising, segmentation of the surfaces, Delaunay representation of the surfaces, and application of fan distortion correction. The correction has been applied theoretically to realistic computer eye models, and experimentally to OCT images of: an artificial eye with a Polymethyl Methacrylate (PMMA) cornea and an intraocular lens (IOL), an enucleated porcine eye, and a human eye in vivo obtained from two OCT laboratory set-ups (time domain and spectral). Data are analyzed in terms of surface radii of curvature and asphericity. Comparisons are established between the reference values for the surfaces (nominal values in the computer model; non-contact profilometric measurements for the artificial eye; Scheimpflug imaging for the real eyes in vivo and vitro). The results from the OCT data were analyzed following the conventional approach of dividing the optical path by the refractive index, after application of 2-D optical correction, and 3-D optical correction (in all cases after fan distortion correction). The application of 3-D optical distortion correction increased significantly both the accuracy of the radius of curvature estimates and particularly asphericity of the surfaces, with respect to conventional methods of OCT image analysis. We found that the discrepancies of the radii of curvature estimates from 3-D optical distortion corrected OCT images are less than 1% with respect to nominal values. Optical distortion correction in 3-D is critical for quantitative analysis of OCT anterior segment imaging, and allows accurate topography of the internal surfaces of the eye. (C) 2010 Optical Society of America
引用
收藏
页码:2782 / 2796
页数:15
相关论文
共 40 条
[1]   Compact, broad-bandwidth fiber laser for sub-2-μm axial resolution optical coherence tomography in the 1300-nm wavelength region [J].
Bizheva, K ;
Povazay, B ;
Hermann, B ;
Sattmann, H ;
Drexler, W ;
Mei, M ;
Holzwarth, R ;
Hoelzenbein, T ;
Wacheck, V ;
Pehamberger, H .
OPTICS LETTERS, 2003, 28 (09) :707-709
[2]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[3]   Tilt and decentration of intraocular lenses in vivo from Purkinje and Scheimpflug imaging - Validation study [J].
de Castro, Alberto ;
Rosales, Patricia ;
Marcos, Susana .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (03) :418-429
[4]   Change in shape of the aging human crystalline lens with accommodation [J].
Dubbelman, M ;
Van der Heijde, GL ;
Weeber, HA .
VISION RESEARCH, 2005, 45 (01) :117-132
[5]   Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography [J].
Dubbelman, M ;
Weeber, HA ;
van der Heijde, RGL ;
Völker-Dieben, HJ .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2002, 80 (04) :379-383
[6]   Ultra high-speed swept source OCT imaging of the anterior segment of human eye at 200 kHz with adjustable imaging range [J].
Gora, Michalina ;
Karnowski, Karol ;
Szkulmowski, Maciej ;
Kaluzny, Bartlomiej J. ;
Huber, Robert ;
Kowalczyk, Andrzej ;
Wojtkowski, Maciej .
OPTICS EXPRESS, 2009, 17 (17) :14880-14894
[7]   Anterior segment imaging with Spectral OCT system using a high-speed CMOS camera [J].
Grulkowski, Ireneusz ;
Gora, Michalina ;
Szkulmowski, Maciej ;
Gorczynska, Iwona ;
Szlag, Daniel ;
Marcos, Susana ;
Kowalczyk, Andrzej ;
Wojtkowski, Maciej .
OPTICS EXPRESS, 2009, 17 (06) :4842-4858
[8]   Role of beat noise in limiting the sensitivity of optical coherence tomography [J].
Haskell, Richard C. ;
Liao, David ;
Pivonka, Adam E. ;
Bell, Tera L. ;
Haberle, Brendan R. ;
Hoeling, Barbara M. ;
Petersen, Daniel C. .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2006, 23 (11) :2747-2755
[9]  
HITZENBERGER CK, 1992, INVEST OPHTH VIS SCI, V33, P98
[10]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181