Comparison of the effects of heat and moisture exchangers and heated humidifiers on ventilation and gas exchange during non-invasive ventilation

被引:60
作者
Jaber, S
Chanques, G
Matecki, S
Ramonatxo, M
Souche, B
Perrigault, PF
Eledjam, JJ
机构
[1] CHU Montpellier, Hop St Eloi, Unite Reanimat & Transplantat, F-34295 Montpellier, France
[2] CHU Montpellier, Hop St Eloi, Dept Anesthesie Reanimat B, F-34295 Montpellier, France
[3] CHU Montpellier, Hop Arnaud Villeneuve, Serv Cent Physiol Clin, F-34295 Montpellier, France
关键词
heat and moisture exchanger; heated humidifier; non-invasive ventilation;
D O I
10.1007/s00134-002-1441-0
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare the short-term effects of a heat and moisture exchanger (HME) and a heated humidifier (HH) during non-invasive ventilation (NIV). Design: Prospective, clinical investigation. Setting: Intensive care unit of a university hospital. Patients: Twenty-four patients with acute respiratory failure (ARF). Intervention: Each patient was studied with a HME and a HH in a random order during two consecutive 20min periods of NIV Measurements and results: Respiratory rate (RR), expiratory tidal volume (VTe) and expiratory minute ventilation (VE) were measured during the last 5 min of each period and blood gases were measured. Mean pressure support and positive end-expiratory pressure levels were, respectively, 15 +/- 4 and 6 +/- 2 cmH(2)O. VE was significantly greater with HME than with HH (14.8 +/- 4.8 vs 13.2 +/- 4.3 l/min; p<0.001). This increase in VE was the result of a greater RR for HME than for HH (26.5 +/- 10.6 vs 24.1 +/- 9.8 breaths/min; p=0.002), whereas the VT for HME was similar to that for HH (674+/-156 vs 643 +/- 148 ml; p=0.09). Arterial partial pressure of carbon dioxide (PaCO2) was significantly higher with a HME than with a HH (43.4 +/- 8.9 vs 40.8 +/- 8.2 mmHg; p<0.005), without significantly changing oxygenation. Conclusion: During NIV the increased dead space of a HME can negatively affect ventilatory function and gas exchange. The effect of HME dead space may decrease efficiency of NIV in patients with ARF.
引用
收藏
页码:1590 / 1594
页数:5
相关论文
共 21 条
[1]   P0.1 IS A USEFUL PARAMETER IN SETTING THE LEVEL OF PRESSURE SUPPORT VENTILATION [J].
ALBERTI, A ;
GALLO, F ;
FONGARO, A ;
VALENTI, S ;
ROSSI, A .
INTENSIVE CARE MEDICINE, 1995, 21 (07) :547-553
[2]   Mechanism of relief of tachypnea during pressure support ventilation [J].
Berger, KI ;
Sorkin, IB ;
Norman, RG ;
Rapoport, DM ;
Goldring, RM .
CHEST, 1996, 109 (05) :1320-1327
[3]   NONINVASIVE VENTILATION - PRACTICAL ISSUES [J].
BROCHARD, L .
INTENSIVE CARE MEDICINE, 1993, 19 (08) :431-432
[4]   NONINVASIVE VENTILATION FOR ACUTE EXACERBATIONS OF CHRONIC OBSTRUCTIVE PULMONARY-DISEASE [J].
BROCHARD, L ;
MANCEBO, J ;
WYSOCKI, M ;
LOFASO, F ;
CONTI, G ;
RAUSS, A ;
SIMONNEAU, G ;
BENITO, S ;
GASPARETTO, A ;
LEMAIRE, F ;
ISABEY, D ;
HARF, A .
NEW ENGLAND JOURNAL OF MEDICINE, 1995, 333 (13) :817-822
[5]  
Campbell R S, 2000, Respir Care, V45, P306
[6]   VENTILATORY MUSCLE SUPPORT IN RESPIRATORY-FAILURE WITH NASAL POSITIVE PRESSURE VENTILATION [J].
CARREY, Z ;
GOTTFRIED, SB ;
LEVY, RD .
CHEST, 1990, 97 (01) :150-158
[7]   EFFECTS OF DRY ANESTHETIC GASES ON TRACHEOBRONCHIAL CILIATED EPITHELIUM [J].
CHALON, J ;
MALEBRANCHE, J ;
LOEW, DAY .
ANESTHESIOLOGY, 1972, 37 (03) :338-+
[8]   USE OF HEAT AND MOISTURE EXCHANGING (HME) FILTERS IN MECHANICALLY VENTILATED ICU PATIENTS - INFLUENCE ON AIRWAY FLOW-RESISTANCE [J].
CHIARANDA, M ;
VERONA, L ;
PINAMONTI, O ;
DOMINIONI, L ;
MINOJA, G ;
CONTI, G .
INTENSIVE CARE MEDICINE, 1993, 19 (08) :462-466
[10]   Airway occlusion pressure at 0.1 s (P0.1) after extubation: an early indicator of postextubation hypercapnic respiratory insufficiency [J].
Hilbert, G ;
Gruson, D ;
Portel, L ;
Vargas, F ;
Gbikpi-Benissan, G ;
Cardinaud, JP .
INTENSIVE CARE MEDICINE, 1998, 24 (12) :1277-1282