Determination and validation of the initial beam parameters of Elekta Agility collimator head by Monte Carlo simulations

被引:4
作者
Chand, Bhagat [1 ,2 ]
Singh, Ranjit [3 ]
Kumar, Mukesh [1 ]
机构
[1] Lovely Profess Univ, Dept Phys, Phagwara 1444141, Punjab, India
[2] Dr Rajendra Prasad Govt Med Coll, Dept Radiotherapy, Kangra 176001, Himachal Prades, India
[3] Postgrad Inst Med Educ & Res, Dept Radiotherapy, Chandigarh 160012, India
关键词
Radiotherapy; BEAMnrc; Agility; DOSXYZnrc; Beam profiles; MEGAVOLTAGE PHOTON BEAMS; SENSITIVITY;
D O I
10.1007/s13246-022-01159-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The availability of geometrical, physical, and initial beam parameters for Monte Carlo (MC) simulations of the Elekta Agility collimator head has become very difficult due to the proprietary nature of this data. This study presents strategies to independently determine the geometrical and physical properties of the components and initial beam parameters of the Agility collimator head for full beam simulations and postulates a benchmarking process using the EGSnrc MC toolkit. Target material of W (90%) and Re (10%) of 0.09 cm thickness, flattening filter of 1.77 cm thick stainless steel placed on 0.5 cm Al disc, and primary and secondary collimators of Tungsten alloy have been found to best fit the Agility head. The initial beam energy of 6.0 MeV with a radial distribution given as full-width half maxima (FWHM) of 0.301 cm (crossline) x 0.201 cm (inline) for 6 MV beam with a mean angular spread of 1.34 degrees has been found best fitting the model. Variations of 0.29% and 0.59% have been noted in the measured and calculated values of TPR20,10 and D-10 respectively. More than 90% dose points for all simulations passed the 2D gamma criteria of 3% DD, 3 mm DTA. This MC model of the Agility head can be used for dose calculation and validation of advanced treatment techniques.
引用
收藏
页码:889 / 899
页数:11
相关论文
共 28 条
[1]   Representation and illustration of the initial parameters in GATE 8.1 monte carlo simulation of an Elekta Versa-HD linear accelerator [J].
Abolaban, Fouad A. ;
Taha, Eslam M. .
JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2020, 13 (01) :642-647
[2]   Monte Carlo linear accelerator simulation of megavoltage photon beams: Independent determination of initial beam parameters [J].
Almberg, Sigrun Saur ;
Frengen, Jomar ;
Kylling, Arve ;
Lindmo, Tore .
MEDICAL PHYSICS, 2012, 39 (01) :40-47
[3]   Comparison of radiotherapy techniques with flattening filter and flattening filter-free in lung radiotherapy according to the treatment volume size [J].
Arslan, Alaettin ;
Sengul, Burak .
SCIENTIFIC REPORTS, 2020, 10 (01)
[4]  
Bahreyni-Toosi Mohammad-Taghi, 2014, J Med Signals Sens, V4, P267
[5]   On Monte Carlo modeling of megavoltage photon beams: A revisited study on the sensitivity of beam parameters [J].
Chibani, Omar ;
Moftah, Belal ;
Ma, C. -M. Charlie .
MEDICAL PHYSICS, 2011, 38 (01) :188-201
[6]   Monte Carlo and analytic modeling of an Elekta Infinity linac with Agility MLC: Investigating the significance of accurate model parameters for small radiation fields [J].
Gholampourkashi, Sara ;
Cygler, Joanna E. ;
Belec, Jason ;
Vujicic, Miro ;
Heath, Emily .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01) :55-67
[7]   Simulation of a 6 MV Elekta Precise Linac photon beam using GATE/GEANT4 [J].
Grevillot, L. ;
Frisson, T. ;
Maneval, D. ;
Zahra, N. ;
Badel, J-N ;
Sarrut, D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (04) :903-918
[8]   Analysis of characteristics and validation of 6 MV photon beam produced by Elekta Synergy linear accelerator using EGSnrc Monte Carlo code [J].
Gundem, Esin ;
Dirican, Bahar .
RADIATION PHYSICS AND CHEMISTRY, 2021, 184
[9]   A Monte Carlo evaluation of dose distribution of commercial treatment planning systems in heterogeneous media [J].
Hasani, Mohsen ;
Mohammadi, Kheirollah ;
Ghorbani, Mahdi ;
Gholami, Soraya ;
Knaup, Courtney .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 :S127-S134
[10]  
Health E, 2015, RADIOTHER ONCOL, VS792, DOI [10.1016/S0167-8140(15)41454-9, DOI 10.1016/S0167-8140(15)41454-9]