A new screening method to diagnose coronary artery disease using multichannel magnetocardiogram and simple exercise

被引:37
作者
Kanzaki, H
Nakatani, S
Kandori, A
Tsukada, K
Miyatake, K
机构
[1] Natl Cardiovasc Ctr, Dept Cardiol, Suita, Osaka 5658565, Japan
[2] Hitachi Ltd, Cent Res Lab, Tokyo, Japan
关键词
magnetocardiography; coronary artery disease; myocardial ischemia; magnetic field; QRS complex;
D O I
10.1007/s00395-003-0392-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Magnetocardiography (MCG) is a non-contact mapping technique to record cardiac action currents. The Master's two-step electrocardiogram (ECG) test is a simple exercise method for screening coronary artery disease (CAD), but it is inadequate concerning the sensitivity. Our aim was to develop a new screening method using multichannel MCG instead of ECG. Methods Thirty subjects (aged 54 +/- 16 years, 27 males), 17 of whom had CAD confirmed by coronary angiography, underwent the Master's exercise ECG test. After the exercise, MCG signals were acquired every minute during recovery with a 64-channel MCG system (MC-6400, Hitachi Ltd). We integrated tangential components of the MCG signals within QRS (during 20, 40, 80, and 120 ms centering on R-wave peak) immediately after exercise (Iex) and 5 minutes after exercise (Irec). The exercise-induced change of currents [(Iex-Irec)/Irec] was determined and normalized for each channel, and the maximal change among 64 channels, maximal QRS integral change, was used as a diagnostic index for myocardial ischemia. Results The maximal QRS integral change during 40 ms was significantly higher in the CAD group than in the control group (0.81 +/- 0.51 vs. 0.36 +/- 0.19, p < 0.01). A sensitivity and specificity for predicting CAD by the change > 0.44 were 82 % and 85 %, respectively, yielding a diagnostic accuracy of 83 %. The conventional Master's ECG test identified the CAD patients with a diagnostic accuracy of 63 % (sensitivity 47 %, specificity 85 %). Conclusion The Master's two-step exercise test with a 64-channel MCG system showed the high diagnostic accuracy, despite of non-contact recording and simple exercise. The magnetic field in the depolarization process has the potential to detect the subtle myocardial ischemia induced by exercise.
引用
收藏
页码:124 / 132
页数:9
相关论文
共 31 条
[1]   THE QRS COMPLEX DURING TRANSIENT MYOCARDIAL ISCHEMIA - STUDIES IN PATIENTS WITH VARIANT ANGINA-PECTORIS AND IN A CANINE PREPARATION [J].
BARNHILL, JE ;
WIKSWO, JP ;
DAWSON, AK ;
GUNDERSEN, S ;
ROBERTSON, RMS ;
ROBERTSON, D ;
VIRMANI, R ;
SMITH, RF .
CIRCULATION, 1985, 71 (05) :901-911
[2]   PHARMACOLOGICAL STRESS IMAGING [J].
BELLER, GA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1991, 265 (05) :633-638
[3]   MAGNETOCARDIOGRAPHY AND EXERCISE TESTING [J].
BROCKMEIER, K ;
COMANI, S ;
ERNE, SN ;
DILUZIO, S ;
PASQUARELLI, A ;
ROMANI, GL .
JOURNAL OF ELECTROCARDIOLOGY, 1994, 27 (02) :137-142
[4]   Magnetocardiography and 32-lead potential mapping: Repolarization in normal subjects during pharmacologically induced stress [J].
Brockmeier, K ;
Schmitz, L ;
Chavez, JB ;
Burghoff, M ;
Koch, H ;
Zimmermann, R ;
Trahms, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (06) :615-626
[5]  
Cantor AA, 2000, CATHETER CARDIO INTE, V50, P177, DOI 10.1002/(SICI)1522-726X(200006)50:2<177::AID-CCD6>3.0.CO
[6]  
2-H
[7]   A computer heart model incorporating anisotropic propagation .4. Simulation of regional myocardial ischemia [J].
Dube, B ;
Gulrajani, RM ;
Lorange, M ;
LeBlanc, AR ;
Nasmith, J ;
Nadeau, RA .
JOURNAL OF ELECTROCARDIOLOGY, 1996, 29 (02) :91-103
[8]  
Gibbons RJ, 1997, J AM COLL CARDIOL, V30, P260
[9]   Spatial distribution of QT dispersion measured by magnetocardiography under stress in coronary artery disease [J].
Hailer, B ;
Van Leeuwen, P ;
Lange, S ;
Wehr, M .
JOURNAL OF ELECTROCARDIOLOGY, 1999, 32 (03) :207-216
[10]   Recording locations in multichannel magnetocardiography and body surface potential mapping sensitive for regional exercise-induced myocardial ischemia [J].
Hänninen, H ;
Takala, P ;
Mäkijärvi, M ;
Montonen, J ;
Korhonen, P ;
Oikarinen, L ;
Simelius, K ;
Nenonen, J ;
Katila, T ;
Toivonen, L .
BASIC RESEARCH IN CARDIOLOGY, 2001, 96 (04) :405-414