Estimating organ dose in computed tomography using tube current modulation: A Monte Carlo simulation

被引:3
|
作者
Hosseinzadeh, V [1 ]
Ghaffari, H. [2 ]
Rezaeyan, A. [2 ]
Deilami, S. [3 ]
机构
[1] Islamic Azad Univ, Dept Nucl Engn, Arsanjan Branch, Arsanjan, Iran
[2] Iran Univ Med Sci, Sch Med, Dept Med Phys, Tehran, Iran
[3] Islamic Azad Univ, Dept Phys, Arsanjan Branch, Arsanjan, Iran
来源
关键词
Computed tomography; tube current modulation; organ dose; thermoluminsent dosimeter; Monte Carlo simulation; CT EXAMINATIONS; CANCER-RISK; VALIDATION; REDUCTION; ADULT; SCAN;
D O I
10.18869/acadpub.ijrr.19.3.575
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The computed tomography (CT) scan delivers a relatively high radiation dose to the patient. One of the critical factors that affects the absorbed dose is the intensity of tube current. The aim of this study is to measure and compare the radiation dose of three radiation-sensitive organs in constant current mode and tube current modulation (TCM) modes. Materials and Methods: CT-scans from the chest and abdomen-pelvis regions of adults in three different current modes were obtained. The absorbed doses of thyroid, lungs, and ovaries were measured using the thermoluminescent dosimeter (TLD) chips embedded in the RANDO phantom. Furthermore, the confirmation of the organ doses was simulated using the Monte Carlo (MC) simulation. The measured doses were evaluated and confirmed by comparison with the simulated doses. Results: The relative differences between the measured and simulated doses for thyroid, lung, and ovary were -4.7%, -1.3%, and -11.7% for constant current mode, -2.2%, -11.2%, and -6.3% for longitudinal modulation mode, and 0.0%, -14.6%, and -9.9% for angular modulation mode, respectively. With longitudinal modulation mode, thyroid, lung, and ovary doses were reduced by 34.0%, 19.0%, and 19.0% for the measured doses and 32.0%, 26.0%, and 13.0% for the simulated doses, respectively. The longitudinal modulation mode resulted in a greater dose reduction compared to the angular modulation for both measured and simulated doses. Conclusion: Using TCM resulted in reducing does received by the organs in both measured and simulated doses. The TCM reduces organ dose, which is more evident in the longitudinal modulation.
引用
收藏
页码:575 / 581
页数:7
相关论文
共 50 条
  • [31] Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT
    Angel, Erin
    Yaghmai, Nazanin
    Jude, Cecilia Matilda
    DeMarco, John J.
    Cagnon, Christopher H.
    Goldin, Jonathan G.
    Primak, Andrew N.
    Stevens, Donna M.
    Cody, Dianna D.
    McCollough, Cynthia H.
    McNitt-Gray, Michael F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (03): : 497 - 511
  • [32] Influence of dosimeter depth on estimating patient's eye lens dose in head computed tomography: A measurement and Monte Carlo simulation study
    Hirosawa, Ayaka
    Matsubara, Kosuke
    Fukuda, Atsushi
    Morioka, Yusuke
    Kitagawa, Masayasu
    Ogawa, Yoshinori
    RADIATION MEASUREMENTS, 2023, 166
  • [33] Monte Carlo calculation of patient organ doses from computed tomography
    Oono T.
    Araki F.
    Tsuduki S.
    Kawasaki K.
    Radiological Physics and Technology, 2014, 7 (1) : 176 - 182
  • [34] Organ doses for reference pediatric and adolescent patients undergoing computed tomography estimated by Monte Carlo simulation
    Lee, Choonsik
    Kim, Kwang Pyo
    Long, Daniel J.
    Bolch, Wesley E.
    MEDICAL PHYSICS, 2012, 39 (04) : 2129 - 2146
  • [35] Monte Carlo Simulation of Coherent Scatter Computed Tomography (CSCT)
    Lewis, K.
    Donnelly, E.
    Pickens, D.
    Price, R.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [36] Evaluation of Organ Doses for Pediatric Computed Tomography Using a Newly Designed Radiophotoluminescence Glass Dosimeter and Comparison with a Monte Carlo Simulation-based Dose Calculator
    Chang, Weishan
    Koba, Yusuke
    HEALTH PHYSICS, 2021, 120 (03): : 288 - 295
  • [37] Evaluation of the effect of patient dose from cone beam computed tomography on prostate IMRT using Monte Carlo simulation
    Chow, James C. L.
    Leung, Michael K. K.
    Islam, Mohammad K.
    Norrlinger, Bernhard D.
    Jaffray, David A.
    MEDICAL PHYSICS, 2008, 35 (01) : 52 - 60
  • [38] Evaluation of radiation dose to organs during kilovoltage cone-beam computed tomography using Monte Carlo simulation
    Son, Kihong
    Cho, Seungryong
    Kim, Jin Sung
    Han, Youngyih
    Ju, Sang Gyu
    Choi, Doo Ho
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 295 - 302
  • [39] A PC program for estimating organ dose and effective dose values in computed tomography
    Kalender, WA
    Schmidt, B
    Zankl, M
    Schmidt, M
    EUROPEAN RADIOLOGY, 1999, 9 (03) : 555 - 562
  • [40] A PC program for estimating organ dose and effective dose values in computed tomography
    W. A. Kalender
    B. Schmidt
    M. Zankl
    M. Schmidt
    European Radiology, 1999, 9 : 555 - 562